Abstract

We have developed a novel approach for using near infrared (NIR) spectroscopy; this is to create and apply calibration models “on-the-fly”. We define a process to create, validate, implement and approve a disposable library for discriminating between active and placebo pharmaceutical dosage forms packaged for clinical studies. This approach is based on two concepts; one germane to the particular needs and requirements of the pharmaceutical industry and the second based on an observation of the proposed experimental parameters. The first is that validation, in the general sense, implies control over a method. The second is that each clinical trial is a unique event, consisting of a finite or closed population of samples. The essence of the idea behind our process is to use disposable libraries, that is, libraries of NIR spectra created for one-time use and then archived. The justification for this approach is that the population of samples used both to create the library and subsequently to be identified using that library consists of a finite set of samples. This paper describes the rationale of this approach, discusses in detail the requirements for suitable protocols for using this method of verifying the dosages used in clinical trials, and describes the application of this approach on two separate clinical trials. Many other examples exist in the authors' laboratory. We also provide formatted templates that can be used to document and report NIR analytical results for discriminating between drug product and placebo dosage forms packaged in patient kits for clinical studies. The use of these templates, carefully designed to ensure regulatory requirements for validation are met, makes the implementation of the technique both rapid and compliant with regulatory requirements.

© 2001 NIR Publications

PDF Article
More Like This
Optical system for tablet variety discrimination using visible/near-infrared spectroscopy

Yongni Shao, Yong He, and Xingyue Hu
Appl. Opt. 46(34) 8379-8384 (2007)

Spatial complexity method for tracking brain development and degeneration using functional near-infrared spectroscopy

Zhenhu Liang, Yuxi Wang, Hao Tian, Yue Gu, Takeshi Arimitsu, Takao Takahashi, Yasuyo Minagawa, Haijing Niu, and Yunjie Tong
Biomed. Opt. Express 13(3) 1718-1736 (2022)

Multispectral method for skin imaging: development and validation

Tianchen Shi and Charles A. DiMarzio
Appl. Opt. 46(36) 8619-8626 (2007)

Cited By

You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access Optica Member Subscription

Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All rights reserved, including rights for text and data mining and training of artificial technologies or similar technologies.