Abstract
We present a single-shot detection method of terahertz correlated second harmonic generation in plasma-based sources by directly mixing an optical probe into femtosecond laser-induced plasma filaments in air. The single-shot second harmonic trace is obtained by measuring a second harmonic generation on a conventional CCD with a spatiotemporally distorted probe beam. The system shows a spectrometer resolution of 22 fs/pixel on the CCD and a true resolution on the order of the probe pulse duration. With considerable THz peak electric field strength, this formalism can open the door to single-shot THz detection without bandwidth limitations.
© 2024 Optica Publishing Group
Full Article | PDF ArticleMore Like This
Kareem J. Garriga Francis, Mervin Lim Pac Chong, E Yiwen, and X.-C. Zhang
Opt. Lett. 47(23) 6297-6300 (2022)
Shin Nakamura, Takashi Fujii, Masahiro Sato, Akiko Kumada, and Yuji Oishi
Opt. Lett. 49(3) 726-729 (2024)
Amit Beer, Dror Hershkovitz, and Sharly Fleischer
Opt. Lett. 44(21) 5190-5193 (2019)