Coupled optical microcavities: an enhanced refractometric sensing configuration

Open Access Open Access

Abstract

We theoretically investigate the application of coupled optical microcavities as refractive index sensors. Coupled microcavities support a very sharp asymmetrical Fano resonance, which gives rise to faster changes in output transmission than the changes from a single cavity. With the output transmission at a fixed wavelength that varies much faster than it does in a single-cavity resonance, the result is enhanced sensitivity of the device to the changes in refractive index. In addition, it is observed that both thermal and optical Kerr effects can be utilized to improve the sensitivity.

©2008 Optical Society of America

1. Introduction

Over the past few years, biological and chemical sensors based on microresonators have attracted much attention because they provide the significant advantages of high sensitivity, real-time response, minimal sample requirement, compatibility with microfluidics, and eliminating the requirement for fluorescent labeling [1–13]. In the resonator-based sensors, the light photons interact with the target molecules multiple times, which builds up the changes of the signal induced by the interactions. Specifically, the mode shift or transmission change at a fixed wavelength in response to variation of the effective optical path length is utilized as the sensing signal. Using this high-sensitivity mechanism, a whispering-gallery-type cavity with a quality factor in excess of 100 million has demonstrated single molecule resolution [14].

Resonance of a single cavity usually has a Lorenzian lineshape, which is symmetrical with respect to the resonant wavelength. Recently, a coupled-cavity configuration, consisting of a Fabry-Perot and ring cavities, has been proposed to give an asymmetric lineshape. This structure provides a larger transmission slope as compared to a single resonator, and enables improved sensitivity in sensing [15–17]. Here we propose an alternatively coupled microcavity geometry, which can generate a sharp asymmetrical line shape and offer one order of magnitude increase in transmission slope and, thereby, increase the sensitivity of the device.

2. Theoretical model

We consider a system in which two microresonators are coupled through a waveguide. One resonant mode is present in each of the cavities, as shown in Fig. 1a. In the case of slowly varying field amplitudes, the modes of this system can be described by the coupled harmonic oscillator model (which is well-known coupled-mode theory in the semi-classical limit), and the motion of the j th cavity mode (j=1, 2) is described by [18]

dcjdt=(iΔjκT,j2)cjκE,j12(ain(j)+bin(j)).

where cj(t) is the bosonic annihilation operator of the j th cavity mode with resonant frequenc y ωj; Δj=ω-ωj denotes the detuning of the input carrier from the resonant frequency; κT,j represents the total cavity decay rate with κT,j=κI,j+2κE,j, such that κI,j is the intrinsic ca vity decay rate and κE,j is the external cavity decay rate; a (j) in(b (j) in) and a (j) out(b (j) out) describe the input and output fields in the left and right ports, respectively.

 figure: Fig. 1.

Fig. 1. (a) Schematic diagram of two coupled cavities. (b) The typical reflection (R), transmission (T), and loss (cavity absorption) spectra. (c) The enlarged reflection and loss spectra of the coupled-cavity system. Used parameters: κE=20κI and θ=-π/8.

Download Full Size | PDF

We are interested in the steady-state regime of the system. The internal cavity mode operator can be expressed as

cj(ω)=κE,j12[ain(j)(ω)+bin(j)(ω)]iΔjκT,j2.

Considering the coupling between the resonators and the waveguide that gives rise to standard cavity input-output formalism X (j) out(ω)=Y (j) in(ω)+κ 1/2 E,jcj(ω), we obtain the output fields

Xout(j)(ω)=κE,jXin(j)(ω)+(iΔj+κE,jκT,j2)Yin(j)(ω)iΔjκT,j2,

where X,Y=a,b and XY. Eq. (3) can be written in a matrix form

(bin(j)(ω)bout(j)(ω))=Tj(ain(j)(ω)aout(j)(ω)),

with the transfer matrix

Tj=1iΔj+κE,jκT,j2(κE,jiΔjκT,j2iΔjκT,j2+2κI,jκE,j).

This is a general description for propagation of light through a single-cavity, wave-guide system. In the case of two coupled cavities, the whole transfer matrix can be written as [19]

(bin(2)(ω)bout(2)(ω))=T(ain(1)(ω)aout(1)(ω)),

with T=T 2 T 0 T 1. T 0 denotes the transfer matrix for propagation through the waveguide,

T0=(0eiθeiθ0).

where θ is the phase shift acquired by the wave-guide mode as it travels for a distance L, with θ=kL where k is the propagation constant. Assuming that there is no wave entering the right port, i.e., b (2) in(ω)=0, we can derive the field transmission and reflection rates, defined by T(ω)≡|b (2) out/(ω)/a (1) in(ω)|2 and R(ω)≡|a (1) out/a (1) in|2, respectively, with

T(ω)=eiθΠj=1,2(iΔj+κE,jκT,j2)e2iθΠj=1,2(iΔjκT,j2)κE,1κE,22,
R(ω)=e2iθ(iΔ2κT,22)κE,1+(iΔ1κT,12+2κE,1)κE,2e2iθΠj=1,2(iΔjκT,j2)κE,1κE,22.

For simplicity but without loss of the underlying physics, we assume the microcavities are identical, i.e., ω 1=ω 2, κ E,1(2)=κ E, κ I,1(2)=κI, κ T,1(2)=2Γ; and, κIκE (over-coupling regime). The over coupling is especially used to reduce the coupling instability (from perturbation of surrounding air and attractive gradient force of cavity mode field) because we can place the waveguide in contact with the microresonator in this case. With simple algebraic calculations, the maximum and minimum reflection rates are found at 0 and κE tan θ, respectively. It should be noted that these positions will be slightly modified if there are significant intrinsic losses and cavity-cavity detuning. In general, the point of the maximum reflection tends to drift from the minimum point, and it reduces the slope of the reflection.

To demonstrate the unique property of this coupled system, Fig. 1(b) shows the typical reflection spectra calculated from Eqs. (8) and (9) when the two cavities are resonant (ω 1=ω 2). Compared with the response of a single cavity, which behaves as a symmetrical Lorenzian lineshape, the coupled cavities possess a unique spectrum appearing as an asymmetric lineshape. This asymmetry originates from the conventional Fano effect [20–22], in which coherent, destructive interference occurs between two coupled cavities. For instance, the whole reflection is a sum of two optical paths: direct reflection from the first cavity and reflection from the second cavity. This configuration of coupled cavities has been proposed for storing photons [23–24] and implementing quantum information [25].

 figure: Fig. 2.

Fig. 2. (a) Slopes of the reflection spectra (dR/) in Fig. 1(b). (b) The slope of the reflection spectrum for various working conditions θ and ω-ω 0. Here κE=20κI.

Download Full Size | PDF

3. Refractometric sensing based on asymmetric transmission lineshape

We propose that this unique asymmetric line shape with enhanced resonance slope can be used for ultra-sensitive, label-free, microcavity-based biosensing. On one hand, both the reflection and transmission exhibit asymmetric line shapes. However, since the reflection signal possesses lower noise, it was chosen as the sensing signal in our example. On the other hand, as shown in Fig. 1a, our theoretical mode is built on the standing-mode assumption. This is valid for microcavity in photonic crystal, but here we focus on whispering gallery modes (WGMs) in realistic microcavities since high-Q standing WGMs can form in the presence of strong modal coupling [26]. The resonant wavelength λ for a WGM in ring-type microcavity (with angular momentum number l) is

λ2πa[ηns+(1η)nc]l,

where a=30 µm is the radius of the microcavities; η denotes the fraction of the light energy traveling in the surrounding medium; ns and nc represent the refractive indexes (RI) of the surrounding medium and the cavity, respectively. Therefore, we have

dλdns(2πal)η.

For a given detection wavelength, the derivation of the reflection rate R is thus obtained as

dRdns=dRdλdλdns=4π2caηlλ2dRdω,

where c is the light velocity in vacuum.

The reflection derivation, dR/dns defines the detection sensitivity of the coupled-cavity structure. As shown in Fig. 2, it provides more than an order of magnitude enhancement in sensitivity (directly proportional to the reflection slope) compared with a single-cavity configuration. With λ~640 nm, l~400, η=0.05,QI,j=ωj/2κI,j~108 [26], and from Fig. 2b, the detection sensitivity approaches 107 when the probe wavelength and θ are optimized, which means that the reflection rate will deviate by 0.1 even when ns changes by only 10-8. It is noted that the reflection slope is sensitive to the phase shift θ. The destructive interference of Fano resonance is strongly dependent on the distance L. In Fig. 2b, θ of the red area with the highest slope covers about π/30, corresponding to the precision in distance of ~20 nm. This precision in distance is not difficult to reach by designing the initial mask for microcavity preparation, since L is dominantly dependent on the center positions of the microcavities.

In addition, such a measurement system could be constructed such that only one of the cavities is used as the sensing head in the test solution, while the other is utilized under normal conditions as a feedback cavity. In this case, both the cavities will almost respond synchronously to the environmental variations (such as thermal fluctuations) on the device because they are located in a micro-scale space. As a result, the reflection spectrum of the coupled system has a shift, but no distortion of the lineshape occurs. When target biomolecules interact with the first cavity, the reflection spectrum will undergo distortion in addition to the shift. Therefore, we can remove the effect of the environmental variations if we can define this kind of spectrum aberrance as the sensing signal, which is our future work.

 figure: Fig. 3.

Fig. 3. Typical reflection spectra (a) and their slopes (b) when the two cavities have a detuning ω 21. Inset in (b) shows the maximum slope as a function of ω 21. Here κE=20κI, θ=-π/8.

Download Full Size | PDF

4. Discussion

Note that the analysis described above is based on ideal conditions in which two identical microcavities are coupled at a constant temperature with negligible, non-linear effect. It is important, however, to address these following issues. First, when the two resonators have a detuning of ω 21=ω 2-ω 1, we find that the detection sensitivity does not degrade dramatically, as shown in Fig. 3. On the other hand, a single cavity mode is in principle tunable with wet chemical etching [27–28], thermally [29] or electrically [30] tuning. Therefore, various techniques could be used to tune the resonance frequency of one of the cavities so that the initial detuning between the two microcavities could be optimized.

Second, the thermal effects (thermo-optic effect and thermal expansion) may play a significant role in the detection. The absorbed power (other than scattering loss) of the coupled structure will heat the cavities and lead to a shift of the coupled mode. We will briefly discuss this effect since it may affect the detection sensitivity and restrict the detection limit. When biomolecules interact with the cavity modes, the reflection spectrum of the coupled cavities undergoes a shift (usually a red shift), δλ=(/dns)δns. If the device works in the right side of the reflection spectrum, there will be a decrease of the reflection rate (point A in Fig. 1(c)) at λ=λ 0 with δR=(dR/dns)δns. However, the rate of absorption of the probe power (point B) by the coupled system increases due to the red shift of the spectrum, as shown in Fig. 1(c). As long as the device material has a positive thermo-optic coefficient (dn/dT), another red shift of δλ′ is added to the reflection peak due to the heating effect. Therefore, the total reflection rate changes δR+δR′ where δR′=(dR/)δλ′, which is larger than the change expected without thermal effect. It shows that the detection sensitivity can be improved by introducing the probe-induced thermal effect if the device is chosen to work in the right side of the spectrum. It must be remembered that there are always some instabilities in the probe power and surrounding temperature. Those instabilities will significantly degrade the advantages of sharp asymmetric line shapes, thus increasing the detection limit [31]. The detailed analysis should consider the dynamical thermal behavior and thermal self-stability. This will be addressed in the future work.

Third, the optical Kerr effect cannot be ignored in a high-Q microcavity due to large intra-cavity field intensity [32]. For example, when Q factors are as high as 108 and the diameter of the cavity is around 60 µm, only 1 mW of coupled power can produce an enormous circulating intra-cavity field intensity (I) on the order of 1 GW/cm2, The RI of the cavity material can be described by nc=n 0+n 2 I, where n 0 and n 2 are linear and non-linear RI, respectively. Similar to the probe-induced thermal effect, as long as the device is working in the right side of the spectrum, the probe-induced Kerr effect can also be used to improve the sensitivity. When the coupled mode has a red shift induced by biomolecules, the intracavity field intensity is increased, which will introduce another red shift due to the Kerr effect. Thus, the optical Kerr effect helps increase the cavity mode shift caused by a small change in the RI of the surrounding medium under appropriate conditions. In other words, the detection sensitivity is enhanced due to the probe-induced Kerr effect.

5. Summary

In conclusion, we theoretically analyzed the applicability of a system, which comprises of two coupled microcavities, as a RI sensor with enhanced sensitivity. Due to the sharp asymmetrical Fano resonance, the output transmission variation (at a fixed wavelength) from the coupled system is much larger than that from a single cavity resonance, thus resulting in an order of magnitude enhancement in detection sensitivity. The effects of thermal and optical Kerr effects on the performance of the system of coupled cavities are also discussed. It is found that both these effects can be used to improve the sensitivity by choosing the appropriate working points.

References and links

1. S. Blair and Y. Chen, “Resonant-enhanced evanescent-wave fluorescence biosensing with cylindrical optical cavities,” Appl. Opt. 40, 570–582 (2001). [CrossRef]  

2. R. W. Boyd and J. E. Heebner, “Sensitive disk resonator photonic biosensor,” Appl. Opt. 40, 5742–5747 (2001). [CrossRef]  

3. F. Vollmer, D. Braun, A. Libchaber, M. Khoshsima, I. Teraoka, and S. Arnold, “Protein detection by optical shift of a resonant microcavity,” Appl. Phys. Lett. 80, 4057–4059 (2002). [CrossRef]  

4. J. L. Nadeau, V. S. Iltchenko, D. Kossakovski, G. H. Bearman, and L. Maleki, “High-Q whispering-gallery mode sensor in liquids,” in Proc. SPIE 4629, 172–180 (2002). [CrossRef]  

5. E. Krioukov, J. Greve, and C. Otto, “Performance of integrated optical microcavities for refractive index and fluorescence sensing,” Sens. Actuators B 90, 58–67 (2003). [CrossRef]  

6. W. Fang, D. B. Buchholz, R. C. Bailey, J. F. Hupp, R. P. H. Chang, and H. Cao, “Detection of chemical species using ultraviolet microdisk lasers,” Appl. Phys. Lett. 85, 3666–3668 (2004). [CrossRef]  

7. M. Noto, M. Khoshsima, D. Keng, I. Teraoka, V. Kolchenko, and S. Arnold, “Molecular weight dependence of a whispering gallery mode biosensor,” Appl. Phys. Lett. 87, 223901 (2005). [CrossRef]  

8. I. M. White, H. Oveys, and X. Fan, “liquid-core optical ring-resonator sensors,” Opt. Lett. 31, 1319–1321 (2006). [CrossRef]   [PubMed]  

9. J. Yang and L. J. Guo, “Optical sensors based on active microcavities,” IEEE J. Sel. Top. Quantum Electron. 12, 143–147 (2006). [CrossRef]  

10. A. M. Armani and K J. Vahala, “Heavy water detection using ultra-high-Q microcavities,” Opt. Lett. 31, 1896–1898 (2006). [CrossRef]   [PubMed]  

11. P. Zijlstra, K. L. van der Molen, and A. P. Mosk, “Spatial refractive index sensor using whispering gallery modes in an optically trapped microsphere,” Appl. Phys. Lett. 90, 161101 (2007). [CrossRef]  

12. A. Francois and M. Himmelhaus, “Optical biosensor based on whispering gallery mode excitations in clusters of microparticles,” Appl. Phys. Lett. 92, 141107 (2008). [CrossRef]  

13. Jacob T. Robinson, Long Chen, and Michal Lipson, “On-chip gas detection in silicon optical microcavities”, Opt. Express 16, 4296–4301 (2008). [CrossRef]   [PubMed]  

14. A. M. Armani, R. P. Kulkarni, S. E. Fraser, R. C. Flagan, and K. J. Vahala, “Label-free, single-molecule detection with optical microcavities,” Science 317, 783–787 (2007). [CrossRef]   [PubMed]  

15. S. Fan, “Sharp asymmetric lineshapes in side-coupled waveguide—cavity systems,” Appl. Phys. Lett. 80, 908–910 (2002). [CrossRef]  

16. C.-Y. Chao and L. J. Guo, “Biochemical sensors based on polymer microrings with sharp asymmetrical resonance,” Appl. Phys. Lett. 83, 1527–1529 (2003). [CrossRef]  

17. W. M. N. Passaro and F. D. Leonardis, “Modeling and design of a novel high-Sensitivity electric field siliconon-insulator sensor based on a whispering-gallery-mode resonator,” IEEE J. Sel. Top. Quantum Electron. 12, 124–133 (2006). [CrossRef]  

18. C. W. Gardiner and P. Zoller, Quantum Noise, Springer, 2004.

19. J. K. S. Poon, J. Scheuer, S. Mookherjea, G. T. Paloczi, Y. Huang, and A. Yariv, “Matrix analysis of microring coupled-resonator optical waveguides,” Opt. Express 12, 91–103 (2004). [CrossRef]  

20. U. Fano, “Effects of configuration interaction on intensities and phase shifts,” Phys. Rev. 124, 1866–1878 (1961). [CrossRef]  

21. W. Liang, L. Yang, J. K. S. Poon, Y. Huang, K. J. Vahala, and A. Yariv, “Transmission characteristics of a Fabry-Perot etalon-microtoroid resonator coupled system,” Opt. Lett. 31, 510–512 (2006). [CrossRef]   [PubMed]  

22. L. Zhou and A. W. Poon, “Fano resonance-based electrically reconfigurable add-drop filters in silicon microring resonator-coupled Mach-Zehnder interferometers,” Opt. Lett. 32, 781–783 (2007). [CrossRef]   [PubMed]  

23. Q. Xu, S. Sandhu, M. L. Povinelli, J. Shakya, S. Fan, and M. Lipson, “Experimental Realization of an On-Chip All-Optical Analogue to Electromagnetically Induced Transparency,” Phys. Rev. Lett. 96, 123901 (2006). [CrossRef]   [PubMed]  

24. L. Maleki, A. B. Matsko, A. A. Savchenkov, and V. S. Ilchenko, “Tunable delay line with interacting whispering gallery mode resonators,” Opt. Lett. 29, 626–628 (2004). [CrossRef]   [PubMed]  

25. Y.-F. Xiao, J. Gao, X.-B. Zou, J. F. McMillan, X. Yang, Y.-L. Chen, Z.-F. Han, G.-C. Guo, and C. W. Wong, “Coupled quantum electrodynamics in photonic crystal nanocavities,” http://arxiv.org/abs/0707.2632.

26. T. J. Kippenberg, S. M. Spillane, and K. J. Vahala, “Modal coupling in traveling-wave resonators,” Opt. Lett. 27, 1669–1671 (2002). [CrossRef]  

27. P. E. Barclay, B. Lev, K. Srinivasan, O. Painter, and H. Mabuchi, “Integration of fiber-coupled high-Q SiNx microdisks with atom chips,” Appl. Phys. Lett. 89, 131108–131110 (2006). [CrossRef]  

28. I. M. White, N. M. Hanumegowda, H. Oveys, and X. Fan, “Tuning whispering gallery modes in optical microspheres with chemical etching,” Opt. Express 13, 10754–10759 (2005). [CrossRef]   [PubMed]  

29. M. L. Gorodetsky and I. S. Grudinin, “Fundamental thermal fluctuations in microspheres,” J. Opt. Soc. Am. B 21, 697–705 (2004). [CrossRef]  

30. K. Djordjev, S. Choi, and P. D. Dapkus, “Microdisk tunable resonant filters and switches,” IEEE Photon. Technol. Lett. 14, 828–830 (2002). [CrossRef]  

31. I. M. White and X. Fan, “On the performance quantification of resonant refractive index sensors,” Opt. Express 16, 1020–1028 (2008). [CrossRef]   [PubMed]  

32. T. J. Kippenberg, S. M. Spillane, and K. J. Vahala, “Kerr-nonlinearity optical parametric oscillation in an ultrahigh-Q toroid microcavity,” Phys. Rev. Lett. 93, 083904 (2004). [CrossRef]   [PubMed]  

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (3)

Fig. 1.
Fig. 1. (a) Schematic diagram of two coupled cavities. (b) The typical reflection (R), transmission (T), and loss (cavity absorption) spectra. (c) The enlarged reflection and loss spectra of the coupled-cavity system. Used parameters: κE =20κI and θ=-π/8.
Fig. 2.
Fig. 2. (a) Slopes of the reflection spectra (dR/) in Fig. 1(b). (b) The slope of the reflection spectrum for various working conditions θ and ω-ω 0. Here κE =20κI .
Fig. 3.
Fig. 3. Typical reflection spectra (a) and their slopes (b) when the two cavities have a detuning ω 21. Inset in (b) shows the maximum slope as a function of ω 21. Here κE =20κI , θ=-π/8.

Equations (12)

Equations on this page are rendered with MathJax. Learn more.

dc j dt = ( i Δ j κ T , j 2 ) c j κ E , j 1 2 ( a in ( j ) + b in ( j ) ) .
c j ( ω ) = κ E , j 1 2 [ a in ( j ) ( ω ) + b in ( j ) ( ω ) ] i Δ j κ T , j 2 .
X out ( j ) ( ω ) = κ E , j X in ( j ) ( ω ) + ( i Δ j + κ E , j κ T , j 2 ) Y in ( j ) ( ω ) i Δ j κ T , j 2 ,
( b in ( j ) ( ω ) b out ( j ) ( ω ) ) = T j ( a in ( j ) ( ω ) a out ( j ) ( ω ) ) ,
T j = 1 i Δ j + κ E , j κ T , j 2 ( κ E , j i Δ j κ T , j 2 i Δ j κ T , j 2 + 2 κ I , j κ E , j ) .
( b in ( 2 ) ( ω ) b out ( 2 ) ( ω ) ) = T ( a in ( 1 ) ( ω ) a out ( 1 ) ( ω ) ) ,
T 0 = ( 0 e i θ e i θ 0 ) .
T ( ω ) = e i θ Π j = 1 , 2 ( i Δ j + κ E , j κ T , j 2 ) e 2 i θ Π j = 1 , 2 ( i Δ j κ T , j 2 ) κ E , 1 κ E , 2 2 ,
R ( ω ) = e 2 i θ ( i Δ 2 κ T , 2 2 ) κ E , 1 + ( i Δ 1 κ T , 1 2 + 2 κ E , 1 ) κ E , 2 e 2 i θ Π j = 1 , 2 ( i Δ j κ T , j 2 ) κ E , 1 κ E , 2 2 .
λ 2 π a [ η n s + ( 1 η ) n c ] l ,
d λ d n s ( 2 π a l ) η .
dR dn s = dR d λ d λ dn s = 4 π 2 c a η l λ 2 dR d ω ,