Expand this Topic clickable element to expand a topic
Skip to content
Optica Publishing Group

Integrated optical frequency-resolved Mach-Zehnder interferometers for label-free affinity sensing

Open Access Open Access

Abstract

Integrated Optical Frequency-Resolved Mach-Zehnder Interferometry (IO FR-MZI) is introduced as an alternative, cost-efficient operation principle for integrated optical label-free affinity sensors that can combine high sensitivity with high versatility in terms of potential applications and experimental configurations. A detailed theoretical analysis of the method is presented followed by a semi-analytical approximation and numerical calculations in order to quantify the sensitivity and limits of detection of the FR-MZI over Single Wavelength MZI. The obtained results substantiate that IO FR-MZI- based sensors constitute a generic technological platform of high sensitivity that can be implemented into a plethora of detection schemes. For an optimized optical design well below 1mm in length the limit of detection can be as low as 0.025Å in terms of adlayer effective thickness allowing for truly miniaturized integrated optical sensors fabricated with high yield with standard microfabrication techniques.

©2010 Optical Society of America

1. Introduction

Over the past few years the ever increasing demand of the healthcare and pharmaceutical sectors for more powerful and cost-efficient diagnostic tools has been the main driving force behind the development of optical biosensing devices. Various types of label-free biochemical sensors have been developed which exhibit high-sensitivity and have the potential to be transformed to miniaturized devices for portable applications, such as carbon nanotube field effect transistors (FETs) [1] or silicon nanowire FETs [2]. However, optical label-free sensors [3] seem to be superior mostly because they are non-invasive and immune to external electromagnetic fields. In addition, they do not require reference electrodes, while they offer galvanic isolation of the optical transducer from the electronics. Most importantly, optical detection in many cases allows for real-time multi-analyte monitoring (multiplexing).

Optical label-free biochemical sensors make use of the changes in the refractive index (i.e. the optical path) occurring close to the reaction surface, or more accurately, changes that take place within the penetration depth of the evanescent field, due to a biochemical reaction. In affinity sensors the changes are induced by the binding between a ligand and a receptor immobilized on the sensor surface (e.g. an immuno-complex between antigen and antibody, or the hybridization of complementary DNA strands). The changes in the refractive index are manifested as a change in the photon flux monitored at the output of the sensor. These label-free sensors, widely known as evanescent-field sensors are mainly represented by the reflectometric interference spectroscopy (RIfS) [4], optical porous silicon biosensors [5], surface plasmon resonance (SPR) [6], nanowire waveguides [7], and planar photonic biosensors such as resonators [8,9] gratings [10,11], photonic crystals [12,13] and waveguide-based Integrated Optical Chips (IOCs) [e.g 1420.]. Compared to the other planar photonic biosensors, the IOC-based label-free biosensors are simpler and more versatile in design and have been proven to be capable of the ultimate degree of integration with the source and detector devices on a single-chip (as in [17] and [18]). More specifically, in the case of dielectric-waveguide IOCs, the Mach–Zehnder Interferometer (MZI) –apart from the incessant interest it attracts for telecommunication and all-optical gate applications (as demonstrated in [21] and [22]) - is one of the most promising biosensing devices due to its high sensitivity and the ability of optoelectronic integration in lab-on-a-chip microsystems [2327].

In this paper, Frequency-Resolved Mach-Zehnder Interferometry (FR-MZI) is presented as a novel operation principle for affinity sensors that can form the basis of versatile and ultra-sensitive label-free detection schemes. First, the basic principle and the main characteristics of the suggested method are presented and compared with the conventional Single-Wavelength MZI (SW-MZI). Then, a semi-analytical approximation combined with numerical results is presented in order to assess the limits of detection followed by a discussion about the usefulness of integrated FR-MZI affinity sensors as a generic cost-efficient technological platform.

2. FR-MZI principle of operation and advantages over SW-MZI

A SW-MZI is an interferometric device that can detect changes of the effective refractive index of a medium. The light from a coherent light source propagates though an optical waveguide via total internal reflection and then is divided into two branches, the sensing arm and the reference arm, by means of a Y-junction. The two beams are recombined again at a second Y-junction, Fig. 1a . Changing the refractive index over the sensing arm produces a change in the phase velocity between the propagating beams, which is translated to an intensity modulation at the output of the MZI mathematically expressed as:

Iout=Iin2[1+cos(Δφ(δα)]
where Iin, and Iou t are the input and output power intensities respectively, and Δφ(α) is the phase difference between the twο propagating beams produced by any parameter α over the sensing arm that can affect its index of refraction. For a typical SW-MZI the phase difference is generally expressed as:
Δϕ=2πLλ[Ns(λ,α)Nr(λ)]
where λ is the operating wavelength, L the sensing arm length and Ns, Nr the effective refractive indices of the sensing and reference arms respectively. The output intensity of the interferometric device is periodic with respect to the phase changes induced in the sensing arm, while Δφ depends on the product of the effective refractive index difference and the length of the sensing area. Therefore long sensing areas ensure high sensitivity even for very small changes of Ns rendering SW-MZI to a very powerful technique that can detect minute changes of the refractive index. For that reason, a biochemical interaction taking place on the sensing arm can be measured as output signal intensity change. Nonetheless, in addition to the advantages of SW-MZI devices, there are certain drawbacks when monochromatic light is used and in particular:

 figure: Fig. 1

Fig. 1 Schematic representation of (a) a standard planar waveguide MZI configuration, and (b) the cross-section of the simulated ridge-waveguide MZI

Download Full Size | PDF

  • (a) Ambiguity: because of the sinusoidal dependence of the output signal on the phase difference, it is not possible to deduce the direction of the phase changes when the phase difference between the two arms of the MZI is an integer multiple of π (Eq. (1)).
  • (b) Signal fading: the sensitivity depends on the initial phase difference between the interferometric arms. As the output intensity follows a cosine function, the device sensitivity dIout/dNs(δα) depends on the original operating point. If the interferometer is tuned close to one of the extrema of the transmission curve, small phase changes will generate low intensity variations; this renders each SW-MZI design applicable only to a limited number of sensing schemes, and consequently any affinity measurement requires its own specifically designed SW-MZI.
  • (c) Size versus Sensitivity: Despite the proven capability of SW-MZIs to successfully detect ∆Ns of the order of 10−6, it can be easily deduced from Eq. (2) that this can only be achieved by increasing considerably the sensing arm length therefore increasing the overall size of the IOC sensor. As an illustrative example, in order to detect a ΔNs~10−6 using a HeNe laser source a sensing arm of several mm in length is necessary [e.g 2327.].
  • (d) Input Sources: SW-MZI requires the use of laser sources with a direct impact on the size and cost of the experimental setup. Even if solid-state laser diodes are used (in order to decrease the size but not the cost), one stumbles again to the problem of having the SW-MZI redesigned for each application, since for one specific input wavelength operation near the quadrature points must be avoided. This puts a severe limitation of using a single SW-MZI interchangeably for various experiments, and the only two options available (either redesigning the interferometer or changing laser sources) are both costly and time-consuming.

In this work, Frequency-Resolved Mach-Zehnder Interferometry (FR-MZI) is introduced as an alternative, versatile and more cost-efficient detection method compared to conventional SW-MZI. In FR-MZI the “traditional” monochromatic laser input source is substituted by a white-light source. Nowadays, standard silicon microfabrication technology can easily produce planar ridge-waveguides that allow single-mode waveguiding of all wavelengths within the UV-NIR range with low transmission losses. In addition, careful selection of the waveguide thickness in conjunction with the nature of the cladding layers ensures rejection of the TM modes simply by exploiting the mode’s power leakage into the silicon substrate [28]. Therefore, on-chip polarization selection can be achieved and monomodal behavior of the waveguides can be ensured. On the detector’s side, the FR-MZI requires the use of a spectrophotometer. In a more advanced scheme, one can also envision a monolithically integrated approach with the MZI fully-integrated with the LED and photodiode on a single chip [17, 18].

The key point of the substitution of the input laser light source with a white-light source though is not simply to reduce the cost and size of the experimental setup, but to address the above mentioned drawbacks of the conventional SW-MZI device while rendering FR-MZI to an even more powerful detection method:

  • (a) Ambiguity: the problem of ambiguity is circumvented since the phase change for every wavelength is different (Eq. (2), therefore it is easier to deduce from a full spectrum changes induced by a biochemical event.
  • (b) Signal Fading: In SW-MZI very careful design of the sensor has to be performed prior to any application in order to avoid operation near the quadrature points. On the contrary, when a broad-band source is employed, there can always exist wavelengths that will be far off from the extrema of the corresponding transmission curves. Hence, any biochemical event or the presence of a different substance will procure an output signal change, in the form of a modified output spectrum independent of the specifics of the biochemical event.
  • (c) Size versus Sensitivity: In addition, the most attractive characteristic of FR-MZI is the following: at the local minima of the spectrum the condition for destructive interference can be written as

    2πLΔN(λ,α)λm=(2m+1)πLΔN(λ,α)λm=m+12,m=0,1,2,

where λm are the wavelengths of the local minima, ΔΝ(λ,α)=Ns(λ,α)-Nr(λ) is the effective index of refraction difference between the sensing and reference arms, while α represents the parameter that changes during a biochemical detection and can be for example the concentration of the buffer solution or the thickness of an adsorbed adlayer or any combination thereof. When the biochemical event takes place a change of δα occurs resulting in a change of the refractive index over the sensing arm δn<<n. Because of δn the local minima shift at a new position λ’=λm+δλ and the condition of Eq. (3) becomes
LΔN(λ,α+δa)λm=m+12
By Taylor expansion, Eq. (4) becomes
δλλm=δαΔΝ(λm,α)αΔΝ(λm,α)λmΔΝ(λ,α)λ|λ=λm
or more explicitly

δλλm=δαΝs(λm,α)α[Νs(λm,α)λmΝs(λ,α)λ|λ=λm][Νr(λm)λmNr(λ)λ|λ=λm]

The first notable feature of the FR-MZI arising from Eq. (6) is that the changes in the output signal do not depend on the length of the sensing arm -as is the case for SW-MZI. Long sensing arms in a FR-MZI only ensure that there can be several local minima within the spectral region of interest which can be analyzed during the signal analysis rendering the statistical analysis even more accurate (from Eq. (3) the longer the arm the larger the number m of λm that can satisfy the destructive interference condition).

In addition, it is to be noted that the spectral shifts can be determined with accuracies far better than the wavelength resolution of the spectrometer employed. This is enabled by fitting entire regions of the experimental curve and especially those regions where the wavenength mathematical dependence can be expressed in closed form. For example, one can fit the data near local extrema (minima or maxima) where the signal dependence on the wavelength is of the form b(λ-λm)2. Here λm is the observable to be tracked in time. As it will be shown in the next section, the spectral dependence of the output light has mostly a sinusoidal form and the parabolic fit is quite accurate around the minima/maxima and for a signal change of Im/4, where Im is the signal maximum. Then, the coefficient b in the parabolic law becomes b=Im/(M∆λ)2, where M are the number of points inside the parabolic region and ∆λ is the spectrometer resolution or wavelength step. Since any pair of values symmetrically located in the vicinity of the extremum can be used to obtain an independent estimate of λm, the set of M data points around λm, results in M/2 estimates. The arithmetic average is closer to the ideal λm by a factor of (M/2)1/2 compared to ∆λ/2. In case of a typical spectrometer with a resolution of ∆λ=0.5nm, the accuracy of a calculated spectral shift δλ can be as low as Δλ2M2 or else ≤0.1Å for M=100. This is the case, of course, if the noise sources do not significantly change the symmetric pairing of the data points on either side of the extremum. For this to be the case, the noise sources in total should be less than b(∆λ)2=Im/M2, which is the smallest signal change next to the extremum. The noise sources include the shot noise of the signal, and the quantization error of the analog to digital converter (ADC) of the spectrometer. The latter is Ms/2n where Ms is the maximum signal that can be read by the spectrometer and n is the bit resolution of the ADC. If 2n> M2, then the quantization error does not interfere. As far the shot noise is concerned, in the case we keep track of a maximum, the shot noise is (Im)1/2, provided the signal is expressed as the number of photons collected at the spectrometer channels per meaurement. Then, the condition Im/M2>(Im)1/2 becomes Im>M4 and applies for the performance to be limited by the spectrometer resolution only. In the case of a minimum, the signal at the minimum and, consequently, the shot noise is zero since ideally the sensing and reference arms are equal in magnitude and 180° out of phase. Near, but not at, the minimum the signal is finite and by applying the same upper limit to the shot noise at the minimum nearest points, where the signal is b(∆λ)2, one arrives at Im>M2, which is a much more relaxed condition compared to the quadratic one at the maximum. It is advantageous, therefore, to employ a local minimum since the conditions on the shot noise are far less stringent. At the same time, and due to destructive interference, the position of the minimum is more or less independent of the spectral content of the light that is coupled into the waveguide. Hence, independently of the length of the FR-MZI sensing arm and the input source employed, ultra-high resolution measurements can be achieved.

A second very important result of Eq. (6) is that the sensitivity of the FR-MZI for specific biochemical detection can be increased by decreasing the denominator. The terms of the denominator depend on the differences of the effective refractive indices of the two arms and their respective derivatives. Minimizing the denominator can be achieved by proper selection of the core and cladding layer materials and their thicknesses upon which the values of the effective indices depend.

These two important points will be analyzed in more detail and explicitly demonstrated in the following sections.

3. Semi-analytical approximation of the FR-MZI spectral output

The general detection principle behind all integrated optical (IO) affinity sensors is the interaction of the evanescent field of the waveguided modes with the sample medium. Local changes Δn(z) of the refractive index of the sample n(z) –where z is the distance from the waveguide- result in effective refractive index change δN. This quantity δN is translated into readily measurable output signals in different types of IO sensors, such as changes in the output intensity as in the case of SW-MZI or spectral shifts δλ as in the case of FR-MZI (see previous section). N for planar waveguides depends on the polarization of the guided modes the wavelength λ, the thickness of the waveguiding film t, as well as the refractive indices of the substrate nsub, the waveguiding film nwg, and the covering medium ncov.

Equation (6) includes the effective refractive indices of the sensing and reference arms and their derivatives with respect to the wavelength as well as the so-called in literature “optical sensitivity” Nα. The magnitude of the produced shift δλ due to a change of the refractive index δΝ for each wavelength and whether it can be measured constitutes the figure of merit for the FR-MZI. In the following semi-analytical approximation, it will be shown that for an IO FR-MZI no stringent selection rules for the geometrical characteristics of the interferometer are necessary, and that it is possible to create a generic parametric design for a low-cost highly-sensitive miniaturized affinity sensor that can be applied interchangeably to various applications.

The detailed analysis of each term of Eq. (6) is as follows:

In general, the effective refractive index changes δΝ in an affinity sensor can be induced mainly by two different effects:

  • (1) changes δncov of the refractive index ncov of a homogeneous covering medium (usually a an aqueous solution)
  • (2) the formation of an adlayer on the surface of the waveguide of thickness d and refractive index nfilm.

If both effects are simultaneously present the resulting δΝ can be expressed as:

δΝ=(Nncov)δncov+(Νd)δd

Two methods have been used in literature to obtain analytical expressions for the “optical sensitivities”. The first method [29] employs the exact mode-guiding condition for a planar two-layer waveguide consisting of the waveguiding film of thickness t and refractive index n w g, and an adsorbed or bound adlayer of thickness d and refractive index nfilm sandwiched between a substrate and a cover medium of refractive indices nsub and ncov, respectively, and then considers the limit of a very thin layer d<<t. The second method [30] applies the perturbation theory to a planar waveguide with any transverse refractive index distribution. Both methods yield the same results, albeit the second one provides more physical insight. The obtained results for Nncovand Ndare summarized in the formulas below as shown in [30]:

Nncov=ncovNnwg2N2nwg2ncov2δcovteff[2(Nncov)21]β
Nd=nwg2N2Nteffnfilm2ncov2nwg2ncov2[(Nncov)2+(Nnfilm)21(Nncov)2+(Nnwg)21]β
with β=0 for TE modes and β=1 for TM modes, where
teff=t+i=sub,covδi
is the effective thickness of the waveguide, and
δi=λ2π1Ν2ni2[1(Nnwg)2+(Nni)21]β
is the penetration depth of the evanescent field of the guided modes into medium i (i=sub for the substrate and i=cov for the cover).

In order to substantiate the high sensitivity and versatility of an IO FR-MZI, the following immunoassay experiment was emulated based on typical literature data. The IO FR-MZI consisted of a silicon nitride (Si3N4) ridge planar waveguide with a 2.5nm-thick and 2μm-wide ridge sandwiched between silicon dioxide (SiO2) cladding layers (Fig. 1b), while the length of the sensing arm was intentionally set to 300μm, a value well below than the mm-long sensing arms of typical SW-MZIs. Such a design can be easily produced with standard microfabrication techniques [23, 25, 28] and allows for single-mode TE00 operation within the VIS-NIR part of the spectrum (450-1200nm). For the simulation purposes, the sensing arm was assumed to be exposed to water and the Cauchy parameters for the calculation of the water’s refractive index were extracted by the experimental data of [31]. The simulations were based on the three-dimensional beam-propagation method (3D-BPM) and have been carried out with commercial software (BeamPROP®, RSoft Design Group, USA). The thickness of the waveguide was used as a free parameter ranging from 70 to 300nm, and the simulations were performed for a wavelength range between 450 and 1200nm.

A standard immunoassay experiment has in general 4 distinct steps: (1) introduction of a buffer solution (usually Phosphate Buffer Saline or PBS), (2) coating of the sensor surface with a monolayer of probe molecules (an antigen/antibody), (3) introduction of a blocking solution to prevent unspecific binding, and (4) introduction of the target molecules (the corresponding antibody/antigen) within the blocking solution that will attach themselves to the coating adlayer augmenting its thickness. Each step alters the refractive index causing a spectral shift for every wavelength and changing the output spectrum. Considering step (1) as the baseline, the introduction of the coating solution alters the refractive index over the sensing arm of the FR-MZI through the formation of the binding adlayer. Usually, the concentration of the coating solution is in the order of a few μg/ml and does not affect the refractive index of the cover medium in a measurable way compared to the change due to the adlayer formation that dominates as a term in Eq. (7) (in [32] it is argued that for aqueous protein solutions the linear dependence of the refractive index on concentration c for a large range of concentrations is dn/dc=0.18ml/gr, therefore in the case of the c~1μg/ml δn is less than 10−7 and then δΝ is less than 10−8). On the contrary, during the blocking step, the introduction of the dense blocking solution (~1-10mg/ml) causes changes in ncov comparable to the changes induced by the increase of the effective thickness of the coating adlayer because of the filling of empty spots with molecules from the blocking medium. Finally, during the last step of the immunoassay a new change in ncov and consequently of the final output spectrum is caused by the antigen-antibody binding and the augmentation of the adlayer. For this gedanken affinity experiment, Ns was calculated for the 450 - 1200nm spectrum for all four steps through Eqs. (8) and (9) using as the initial value of Ns the results of the 3D-BPM simulations for water. In order to calculate Ns in the presence of PBS, a difference of δncov=0.00167 compared to water was assumed based on [32]. For the second step, a uniform coating adlayer of 4nm effective thickness with a refractive index of nfilm=1.45 was further assumed [32, 33]. The introduction of the blocking solution was assumed to alter nc by δncov=0.00019 with respect to the buffer solution, and to increase the coating adlayer effective thickness by δd=1nm. Finally, the binding event was assumed to increase the effective thickness of the adlayer by δd=0.5nm. The effective refractive indices Ns calculated through Eqs. (7)-(9) were used in Eqs. (1) and (2) to calculate the anticipated spectra of two IO FR-MZI with Si3N4 thicknesses of 100 and 150nm (Fig. 2 ).

 figure: Fig. 2

Fig. 2 Output spectra of simulated IO FR-MZIs with a (a) 100nm-thick and (b) 150nm-thick ridge waveguide for a hypothetical immunoassay experiment. Binding of 0.5nm of target molecules on a uniform 4nm coating adlayer with nfilm=1.45 was assumed. Four distinct steps of the immunoassay are presented: introduction of the PBS buffer solution (circles), the coating of the sensing arm with a 4nm-thick adlayer of nf ilm=1.45 (trangles), introduction of the blocking solution and augmentation of the coating adlayer by 1nm (squares), and binding of the target molecules that increases the effective thickness of the adlayer by 0.5nm (diamonds). Insets: magnifications of the outlined regions.

Download Full Size | PDF

In Fig. 2a spectra, a sinusoidal behavior is observed for all immunoassay steps except for the 600-700nm wavelength region approximately where this behavior collapses. In this region, the interference pattern disappears and large variations of the output intensity take place. For wavelengths shorter than 600nm the monitored interference patterns are red shifted with the augmentation of the adlayer effective thickness, while for wavelengths longer than 700nm blue shift is observed. For this reason this spectral region was named transition region. This transition region is a characteristic of FR-MZI detection principle, and its position depends on the waveguide thickness, see for example Fig. 2b. The usefulness of the transition region and the exploitation of the various shifts (blue or red) will be analyzed in the next section.

4. Sensitivity and limit of detection of FR-MZI

From Fig. 2, it can be easily seen that regardless of the waveguide thickness, an effective thickness augmentation by 0.5nm is readily detectable with a FR-MZI since it results in shifts of the order of 10-20nm between the blocking and binding spectra (insets Fig. 2). From the parabolic fit at local minima, it can be derived that the FR-MZI can detect changes of the effective thickness of the binding layer down to 0.025Å (which is in essence translated to being able to detect single-molecule layers or solutions of very small molecules in ultra low concentrations). The limitations of the method are mostly imposed by the resolution Δλ of the spectrometer.

From Eqs. (8) and (9) and the discussion on the perturbational theory of [30], it is established that an important factor determining the sensitivity of a conventional SW evanescent-field sensor is the waveguide thickness. Both “sensitivities” are inversely proportional to the effective thickness of the waveguide, indicating that thinner waveguides can provide higher sensitivity. The physical reason is that thinner waveguides strongly confine the transverse field distribution of the guided waves resulting in high field strength of the “probing” evanescent field. Equations (8) and (9) can be used to calculate directly the effective refractive indices of the sensing arm due to any change δα of the analyte as well as to calculate analytically the numerator of Eq. (6).

The ridge waveguide used in the gedanken immunoassay of the previous section can be approximated to first order by a slab waveguide; that way N can be analytically expressed through the following transcendental equations for TE00 and TM00 modes respectively [34]:

2πλcosθ=tan1sin2θ(ncovnwg)2cos2θ+tan1sin2θ(ncovnsub)2cos2θ,forTE00
2πλcosθ=tan1(nwgncov)2sin2θ1(ncovnwg)2cos2θ+tan1(nwgnsub)2sin2θ1(nsubnwg)2cos2θ,forTM00
with
N=nwgsinθ
and
λ=λοnwg
where λο is the free-space wavelength.

By taking the derivatives of Eqs. (12) and (13) with respect to λο and employing Eq. (14), one ends up with analytical expressions for the derivatives dNi/dλ for TE00 and TM00 (where the subscript “o” has been omitted for simplicity):

Nλ=2πtΝλ2(nwg2N2)2πtλ+1N2ncov2+1N2nsub2,forTE00
Nλ=2πtΝλ2(nwg2N2)2πtλ+i=sub,covni2nwg2(N2ni2)12(ni2N2+nwg2N2ni2nwg2),forTM00
By using the values of Nr obtained by the simulations, and the values of Ns calculated for the emulated immunoassay, all terms of Eq. (6) can be calculated numerically through the expressions (8), (9), (16) and (17). That way, the anticipated spectral shifts δλ for any FR-MZI due to any change δα of the analyte can be caclulated. The numerical results for δλ due to the binding of the target molecules and the augmentation of the adlayer by δd=0.1nm are shown in Fig. 3 in conjunction with the differences of the anticipated spectra. It can be seen that there is good agreement between the shifts obtained by the analytical expressions and the calculation of the anticipated spectra, especially in terms of predicting the transition region. For any waveguide thickness, the shifts are in the order of 5 nm for the largest part of the VIS-NIR range with pronounced changes around the transition region. The waveguide thickness mostly affects the output signal by shifting the transition region towards the IR for thicker waveguides. Hence, one can employ any FR-MZI irrespective of its thickness for an affinity sensing measurement without transforming and adapting its design. This can be further illustrated in Fig. 4 , where for the random choice of a FR-MZI with t=100nm, the binding of 0.1nm of any protein with nfilm ranging from 1.38 to 1.56 [33] can be measured. In other words, the FR-MZI can operate equally satisfying for any possible affinity experiment with any input white-light source without having to recur to its redesign or the selection of a different laser line as would have been the case of a standard SW-MZI.

 figure: Fig. 3

Fig. 3 Spectral shifts (circles) as obtained through the analytical expressions of Eqs. (6) and (16) in conjunction with the anticipated spectral difference (squares) of the semi-analytical approximation for the binding of 0.1nm of adlayer on a (a) 100nm-thick and (b) 150nm-thick (open symbols) IO FR-MZI with 300μm-long sensing arms.

Download Full Size | PDF

 figure: Fig. 4

Fig. 4 Anticipated spectral shifts after the binding of 0.1nm of protein with nfilm=1.38-1.56 for a 100nm-thick FR-MZI for the emulated immunoassay experiment

Download Full Size | PDF

The existence of the transition region can be now explained in physical terms with the aid of Eq. (6) and (16)-(17). For a conventional SW-MZI the sensitivity can be increased by decreasing the thickness of the waveguide; for a FR-MZI the sensitivity (i.e. the spectral shift) can be increased by decreasing the denominator in Eq. (6). The denominator is the difference of the effective refractive indices of the sensing and reference arms and their first derivatives with respect to the wavelength and the physical quantity α. For relatively short wavelengths the effective refractive index of both arms is dominated by the index of refraction of the core material, hence ΔΝ → 0. On the contrary, the absolute values of the derivatives are large and the term Δ(∂N/∂λ) dominates over ΔΝ. Since ∂Ns/∂λ < ∂Nr/∂λ ∀λ, δλ is positive, therefore the spectrum for short wavelengths is red shifted. For longer wavelengths (i.e. by moving closer to the cut-off wavelength) the absolute value of the derivatives becomes increasingly smaller, hence the term Δ(∂N/∂λ) → 0 and ΔΝ<0 dominates and consequently the spectrum is blue shifted. For intermediate wavelengths the two terms of the denominator become comparable and the denominator is close to zero. Mathematically, this leads to infinite shifts; practically, this is the transition region where the red shift turns to blue shift and intense variations in the transmission efficiency occur as illustrated in Figs. 3 and 4.

As a result, when designing an IO FR-MZI the increase of the sensitivity is quite straightforward and is achieved through operation close to the transition region. It was established that irrespectively of the waveguide thickness and for most affinity sensing schemes where aqueous solutions are used (nfilm~1.33-1.35) and the refractive index of the adlayers is between 1.38 and 1.56, the spectral shifts for any FR-MZI are within the same order of magnitude and that limit of detection is an effective adlayer thickness of 0.025Å. Therefore for operation close to the transition region one needs to select the core and cladding layer materials and their thicknesses in such a way that the transition region falls close to the maximum of the emission spectrum and of the response curve of the available input source and detector, respectively. This is illustrated in Fig. 5 where the anticipated shifts (per nm of bound target molecule) for waveguides of t ranging from 60 to 300nm are shown for four different wavelengths (500nm, 700nm, 850nm and 1000nm).

 figure: Fig. 5

Fig. 5 Spectral shifts per nm of bound molecule as a function of the FR-MZI thickness for four wavelengths (λ=500, 700, 850 and 1000nm)

Download Full Size | PDF

In reverse, in case one has a specific finalized design of an IO FR-MZI, the choice of the appropriate input source becomes straightforward. It can be deduced from Eq. (6) (and taking into account Eqs. (8) and (9)) that the transition region for a waveguide of thickness t is close to

λt2πt
which is a convenient rule of thumb that can be used to select the operation spectral range without the need to recur to extensive simulations (see also Fig. 5). Moreover, the minimum shift δλ that can be obtained at wavelengths far from the transition region can be approximated with

δλλδαNsαΔN>0 (redshift) for λ<<λt 

since, for N→nwg, λNiλ2(Ninwg) (i=s or r), and

δλλδαNsαΔN<0(blueshift)forλ>>λt
since for λ→λcutoff, Δ(Nλ)0

Using the approximations of Eqs. (19) and (20) and by employing the effective refractive indices of the two MZI arms for any IO FR-MZI design in question (assuming that the sensing arm is exposed to water for quicker calculations), one can readily set the limits of detection for a planned affinity experiment. From the approximations of Eqs. (19) and (20) it can also be seen that for the entire VIS-NIR range the minimum adlayer thickness that can be detected is given

δλ4πδαNs2ncov2
For a given detector resolution and related signal analysis (as discussed in section 2), one can beforehand set the limit of detection of any planned affinity experiment for a given IO FR-MZI design.

5. Conclusions

IO Frequency-Resolved MZI overcomes all limitations of conventional Single-Wavelength MZI associated with phase ambiguity and signal fading. Moreover, the need for long SW-MZI for improved sensitivity and low limit of detection is resolved by employing spectral analysis. The IO FR-MZI provides high sensitivities (in terms of detectable spectral shifts per adlayer thickness) in any part of the VIS-NIR spectrum regardless of the design and the specifics of the affinity experiment. This means that any broad-band source within the VIS-NIR region can be used and that the geometry of the interferometer or the targeted application does not affect the selection of the input source. Finally, the performance and the limit of detection of any FR-MZI for any sensing scheme can be predicted by simple design rules as expressed by the set of Eqs. (19)-(21). Inversely, for an optimized optical design well below 1mm in length the limit of detection can be as low as 0.025Å in terms of adlayer effective thickness -in contrast with the standard SW-MZI- allowing for truly miniaturized integrated optical sensors fabricated with high yield with standard microfabrication techniques.

Acknowledgments

This work was supported by the EU-funded Project “PYTHIA” (FP7-ICT2-224030). The authors would like to thank Dr. S.E. Kakabakos and Dr. P.S. Petrou for helpful discussions.

References and Links

1. A. Star, E. Tu, J. Niemann, J. C. Gabriel, C. S. Joiner, and C. Valcke, “Label-free detection of DNA hybridization using carbon nanotube network field-effect transistors,” Proc. Natl. Acad. Sci. U.S.A. 103(4), 921–926 (2006). [CrossRef]   [PubMed]  

2. W. U. Wang, C. Chen, K. H. Lin, Y. Fang, and C. M. Lieber, “Label-free detection of small-molecule-protein interactions by using nanowire nanosensors,” Proc. Natl. Acad. Sci. U.S.A. 102(9), 3208–3212 (2005). [CrossRef]   [PubMed]  

3. X. Fan, I. M. White, S. I. Shopova, H. Zhu, J. D. Suter, and Y. Sun, “Sensitive optical biosensors for unlabeled targets: a review,” Anal. Chim. Acta 620(1-2), 8–26 (2008). [CrossRef]   [PubMed]  

4. J. Lu, C. M. Strohsahl, B. L. Miller, and L. J. Rothberg, “Reflective Interferometric Detection of Label-Free Oligonucleotides,” Anal. Chem. 76(15), 4416–4420 (2004). [CrossRef]   [PubMed]  

5. G. D. Francia, V. L. Ferrara, S. Manzo, and S. Chiavarini, “Towards a label-free optical porous silicon DNA sensor,” Biosens. Bioelectron. 21(4), 661–665 (2005). [CrossRef]   [PubMed]  

6. T. T. Goodrich, H. J. Lee, and R. M. Corn, “Direct detection of genomic DNA by enzymatically amplified SPR imaging measurements of RNA microarrays,” J. Am. Chem. Soc. 126(13), 4086–4087 (2004). [CrossRef]   [PubMed]  

7. J. Lou, L. Tong, and Z. Ye, “Modeling of silica nanowires for optical sensing,” Opt. Express 13(6), 2135–2140 (2005). [CrossRef]   [PubMed]  

8. A. M. Armani, R. P. Kulkarni, S. E. Fraser, R. C. Flagan, and K. J. Vahala, “Label-free, single-molecule detection with optical microcavities,” Science 317(5839), 783–787 (2007). [CrossRef]   [PubMed]  

9. A. Ksendzov and Y. Lin, “Integrated optics ring-resonator sensors for protein detection,” Opt. Lett. 30(24), 3344–3346 (2005). [CrossRef]  

10. J. Vörös, J. J. Ramsden, G. Csúcs, I. Szendro, S. M. De Paul, M. Textor, and N. D. Spencer, “Optical grating coupler biosensors,” Biomaterials 23(17), 3699–3710 (2002). [CrossRef]   [PubMed]  

11. M. Wiki and R. E. Kunz, “Wavelength-interrogated optical sensor for biochemical applications,” Opt. Lett. 25(7), 463–465 (2000). [CrossRef]  

12. E. Chow, A. Grot, L. W. M. Mirkarimi, M. Sigalas, and G. Girolami, “Ultracompact biochemical sensor built with two-dimensional photonic crystal microcavity,” Opt. Lett. 29(10), 1093–1095 (2004). [CrossRef]   [PubMed]  

13. P. Domachuck, H. Nguyen, H. C. Eggleton, B. J. Straub, and M. Gu, “Microfluidic tunable photonic band-gap device,” Appl. Phys. Lett. 84(11), 1838–1840 (2004). [CrossRef]  

14. M. I. Alayo, M. N. P. Carreρo, D. Criado, and I. Pereyra, “Optical and structural characterization of PECVD-silicon oxynitride films for waveguide device applications,” Proc. SPIE 5730, 250 (2005). [CrossRef]  

15. K. Schmitt, B. Schirmer, C. Hoffmann, A. Brandenburg, and P. Meyrueis, “Interferometric biosensor based on planar optical waveguide sensor chips for label-free detection of surface bound bioreactions,” Biosens. Bioelectron. 22(11), 2591–2597 (2007). [CrossRef]  

16. N. Skivesen, R. Horvath, S. Thinggaard, N. B. Larsen, and H. C. Pedersen, “Deep-probe metal-clad waveguide biosensors,” Biosens. Bioelectron. 22(7), 1282–1288 (2007). [CrossRef]  

17. B. Maisenhölder, H. P. Zappe, M. Moser, P. Riel, R. E. Kunz, and J. Edlinger, “Monolithically integrated optical interferometry for refractometry,” Electron. Lett. 33(11), 986 (1997). [CrossRef]  

18. K. Misiakos, S. E. Kakabakos, P. S. Petrou, and H. H. Ruf, “A monolithic silicon optoelectronic transducer as a real-time affinity biosensor,” Anal. Chem. 76(5), 1366–1373 (2004). [CrossRef]   [PubMed]  

19. W. Lukosz, “Integrated optical chemical and direct biochemical sensors,” Sen. Act. B 29(1-3), 37–50 (1995). [CrossRef]  

20. K. Zinoviev, L. G. Carrascosa, J. S. del Rio, B. Sepulveda, C. Dominguez, and L. M. Lechuga, “Silicon Photonic Biochips for Lab-on-a-Chip Applications,” Adv. Opt. Technol. 2008, 383927 (2008).

21. J. Song, Q. Fang, S. H. Tao, M. B. Yu, G. Q. Lo, and D. L. Kwong, “Passive ring-assisted Mach-Zehnder interleaver on silicon-on-insulator,” Opt. Express 16(12), 8359–8365 (2008). [CrossRef]   [PubMed]  

22. Y.-D. Wu, T.-T. Shih, and M. H. Chen, “New all-optical logic gates based on the local nonlinear Mach-Zehnder interferometer,” Opt. Express 16(1), 248–257 (2008). [CrossRef]   [PubMed]  

23. F. Prieto, B. Sep lveda, A. Calle, A. Llobera, C. Dom nguez, A. Abad, A. Montoya, and L. M. Lechuga, “An integrated optical interferometric nanodevice based on silicon technology for biosensor applications,” Nanotechnology 14(8), 907–912 (2003). [CrossRef]  

24. F. Prieto, B. Sepϊlveda, A. Calle, A. Llobera, C. Domνnguez, and L. M. Lechuga, “Integrated mach-Zehnder interferometer based on ARROW structures for biosensor applications,” Sens. Act. B 92(1-2), 151–158 (2003). [CrossRef]  

25. R. G. Heideman and P. V. Lambeck, “Remote opto-chemical sensing with extreme sensitivity: design, fabrication and performance of a pigtailed integrated optical phase-modulated Mach–Zehnder interferometer system,” Sens. Act. B 61(1-3), 100–127 (1999). [CrossRef]  

26. B. Y. Shew, C. H. Kuo, Y. C. Huang, and Y. H. Tsai, “UV-LIGA interferometer biosensor based on the SU-8 optical waveguide,” Sens. Act. A 120(2), 383–389 (2005). [CrossRef]  

27. B. Drapp, J. Piehler, A. Brecht, G. Gauglitz, B. J. Luff, J. S. Wilkinson, and J. Ingenhoff, “Integrated optical Mach-Zehnder interfrometers as simazine imuunoprobes,” Sens Act B 38–39, 277–282 (1997). [CrossRef]  

28. F. Morichetti, A. Melloni, M. Martinelli, R. G. Heideman, A. Leinse, D. H. Geuzebroek, and A. Borreman, “Box-Shaped Dielectric Waveguides: A New Concept in Integrated Optics?” J. Light. Technol. 25(9), 2579–2589 (2007). [CrossRef]  

29. W. Lukosz and K. Tiefenthaler, “Sensitivity of integrated optical grating and prism couplers as (bio)-chemical sensors,” Sens. Act 15(3), 273–284 (1988). [CrossRef]  

30. W. Lukosz, “Principles and sensitivities of integrated optical and surface plasmon sensors for direct affinity and immunosensing,” Biosens. Bioelectron. 6(3), 215–225 (1991). [CrossRef]  

31. P. D. T. Huibers, “Models for the wavelength dependence of the index of refraction of water,” Appl. Opt. 36(16), 3785–3787 (1997). [CrossRef]   [PubMed]  

32. Handbook of Chemistry and Physics, R. D. Weast, ed. (CRC Presse, Fla, 1978–79).

33. J. Voros, “The Density and Refractive Index of Adsorbing Protein Layers,” Biophys. J. 87(1), 553–561 (2004). [CrossRef]   [PubMed]  

34. H. Zappe, Introduction to Semiconductor Integrated Optics (Artech House), Chap. 7.

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (5)

Fig. 1
Fig. 1 Schematic representation of (a) a standard planar waveguide MZI configuration, and (b) the cross-section of the simulated ridge-waveguide MZI
Fig. 2
Fig. 2 Output spectra of simulated IO FR-MZIs with a (a) 100nm-thick and (b) 150nm-thick ridge waveguide for a hypothetical immunoassay experiment. Binding of 0.5nm of target molecules on a uniform 4nm coating adlayer with nfilm=1.45 was assumed. Four distinct steps of the immunoassay are presented: introduction of the PBS buffer solution (circles), the coating of the sensing arm with a 4nm-thick adlayer of nf ilm=1.45 (trangles), introduction of the blocking solution and augmentation of the coating adlayer by 1nm (squares), and binding of the target molecules that increases the effective thickness of the adlayer by 0.5nm (diamonds). Insets: magnifications of the outlined regions.
Fig. 3
Fig. 3 Spectral shifts (circles) as obtained through the analytical expressions of Eqs. (6) and (16) in conjunction with the anticipated spectral difference (squares) of the semi-analytical approximation for the binding of 0.1nm of adlayer on a (a) 100nm-thick and (b) 150nm-thick (open symbols) IO FR-MZI with 300μm-long sensing arms.
Fig. 4
Fig. 4 Anticipated spectral shifts after the binding of 0.1nm of protein with nfilm=1.38-1.56 for a 100nm-thick FR-MZI for the emulated immunoassay experiment
Fig. 5
Fig. 5 Spectral shifts per nm of bound molecule as a function of the FR-MZI thickness for four wavelengths (λ=500, 700, 850 and 1000nm)

Equations (21)

Equations on this page are rendered with MathJax. Learn more.

I o u t = I i n 2 [ 1 + cos ( Δ φ ( δ α ) ]
Δ ϕ = 2 π L λ [ N s ( λ , α ) N r ( λ ) ]
2 π L Δ N ( λ , α ) λ m = ( 2 m + 1 ) π L Δ N ( λ , α ) λ m = m + 1 2 , m = 0 , 1 , 2 ,
L Δ N ( λ , α + δ a ) λ m = m + 1 2
δ λ λ m = δ α Δ Ν ( λ m , α ) α Δ Ν ( λ m , α ) λ m Δ Ν ( λ , α ) λ | λ = λ m
δ λ λ m = δ α Ν s ( λ m , α ) α [ Ν s ( λ m , α ) λ m Ν s ( λ , α ) λ | λ = λ m ] [ Ν r ( λ m ) λ m N r ( λ ) λ | λ = λ m ]
δ Ν = ( N n cov ) δ n cov + ( Ν d ) δ d
N n cov = n cov N n w g 2 N 2 n w g 2 n cov 2 δ cov t e f f [ 2 ( N n cov ) 2 1 ] β
N d = n w g 2 N 2 N t e f f n f i l m 2 n cov 2 n w g 2 n cov 2 [ ( N n cov ) 2 + ( N n f i l m ) 2 1 ( N n cov ) 2 + ( N n w g ) 2 1 ] β
t e f f = t + i = s u b , cov δ i
δ i = λ 2 π 1 Ν 2 n i 2 [ 1 ( N n w g ) 2 + ( N n i ) 2 1 ] β
2 π λ cos θ = tan 1 sin 2 θ ( n cov n w g ) 2 cos 2 θ + tan 1 sin 2 θ ( n cov n s u b ) 2 cos 2 θ , f o r T E 00
2 π λ cos θ = tan 1 ( n w g n cov ) 2 sin 2 θ 1 ( n cov n w g ) 2 cos 2 θ + tan 1 ( n w g n s u b ) 2 sin 2 θ 1 ( n s u b n w g ) 2 cos 2 θ , f o r T M 00
N = n w g sin θ
λ = λ ο n w g
N λ = 2 π t Ν λ 2 ( n w g 2 N 2 ) 2 π t λ + 1 N 2 n cov 2 + 1 N 2 n s u b 2 , f o r T E 00
N λ = 2 π t Ν λ 2 ( n w g 2 N 2 ) 2 π t λ + i = s u b , cov n i 2 n w g 2 ( N 2 n i 2 ) 1 2 ( n i 2 N 2 + n w g 2 N 2 n i 2 n w g 2 ) , f o r T M 00
λ t 2 π t
δ λ λ δ α N s α Δ N > 0   ( redshift )  for  λ < < λ t  
δ λ λ δ α N s α Δ N < 0 ( blueshift ) for λ > > λ t
δ λ 4 π δ α N s 2 n cov 2
Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All rights reserved, including rights for text and data mining and training of artificial technologies or similar technologies.