Abstract
The spectral radiation characteristic of a non-luminous flame is analyzed. The apparatus and the calibration procedure based on infrared emission spectrometry for measurements of the flame are introduced. The influence of background radiation and stray light on the measurement results could be reduced and suppressed by the design of thermolator and digital lock-in technique. A blackbody cavity was used as reference emission source to calibrate the system that completed absolute measurement. The spectral measurement range is 1---20 ?m. The least measuring distance and the lowest power detected at the entrance pupil are 550 mm and 10^(-9) W/cm2, respectively. The experimental results show that the measure error is less than 10%.
© 2005 Chinese Optics Letters
PDF Article
More Like This
Cited By
You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.
Contact your librarian or system administrator