Abstract
Fluorescent proteins are excited by light that is polarized parallel to the dipole axis of the chromophore. In two-photon microscopy, polarized light is used for excitation. Here we reveal surprisingly strong polarization sensitivity in a class of genetically encoded, GPCR-based neurotransmitter sensors. In tubular structures such as dendrites, this effect led to a complete loss of membrane signal in dendrites running parallel to the polarization direction of the excitation beam. To reduce the sensitivity to dendritic orientation, we designed an optical device that generates interleaved pulse trains of orthogonal polarization. The passive device, which we inserted in the beam path of an existing two-photon microscope, removed the strong direction bias from fluorescence and second-harmonic (SHG) images. We conclude that for optical measurements of transmitter concentration with GPCR-based sensors, orthogonally polarized excitation is essential.
© 2022 Optica Publishing Group under the terms of the Optica Open Access Publishing Agreement
1. Introduction
Excitation of fluorescence depends on the orientation of the fluorophore with respect to the polarization of the excitation light. The emitted photons are also polarized, an effect known as fluorescence anisotropy. This effect can be exploited to measure the speed of rotation of fluorescent proteins in solution [1,2], for example to detect interactions during molecular signaling. Commercial two-photon microscopes use linearly polarized light for excitation. In the case of two-photon light-sheet microscopy, where fluorescence is detected orthogonally to the excitation sheet, it has been noted that better results are achieved when the direction of polarization is optimized [3]. In classical point-scanning two-photon microscopes, however, the direction of polarized excitation is ignored, as fluorescent or fluorescently-labeled molecules in biological preparations are in most cases randomly oriented.
To detect the release of neurotransmitters and neuromodulators in intact brain tissue, numerous genetically encoded sensors have been developed. A very successful strategy is to splice circularly permuted GFP (cpGFP) into G protein-coupled neurotransmitter receptors (GPCRs), which resulted in several highly specific sensors with a large dynamic range [4–8]. GPCRs have seven transmembrane domains, and the cpGFP is inserted into the third intracellular loop, tethering both sides of the GFP barrel to transmembrane helices of the receptor. Upon ligand binding, movements of the helices are transferred to the cpGFP, changing its fluorescence. The relative change in fluorescence (Δf/f0) depends on neurotransmitter concentration in a logistic fashion. However, depending on the length and rigidity of the linker sections, tethering cpGFP to transmembrane domains may orient the barrel parallel to the membrane.
Here we show that two-photon imaging of GPCR-based sensors is surprisingly sensitive to the orientation of the imaged dendrite relative to the polarization of the excitation beam. As a consequence, entire sections of reporter neurons can be non-fluorescent, which makes it impossible to draw conclusions about the spatial distribution of neurotransmitter release from the relative change in fluorescence. To solve this problem, we developed a passive optical device that produces two collinear output beams with orthogonal polarization (‘X-Pol’). When inserted into the beam path of a two-photon microscope, the X-Pol device removes the orientation bias when imaging membrane proteins. As the polarization sensitivity of genetically encoded sensors cannot be known a priori, we suggest using orthogonal polarized pulse trains as a new standard for two-photon microscopy.
2. Materials and methods
2.1 Organotypic slice cultures
Hippocampal slice cultures from Wistar rats of either sex were prepared at postnatal day 4–6. Rats were anesthetized with 80% CO2 20% O2 and decapitated. Hippocampi were dissected in cold slice culture dissection medium containing (in mM): 248 sucrose, 26 NaHCO3, 10 glucose, 4 KCl, 5 MgCl2, 1 CaCl2, 2 kynurenic acid, 0.001% phenol red (saturated with 95% O2, 5% CO2, pH 7.4). Tissue was cut into 400 µM thick sections on a tissue chopper (McIlwain) and cultured on membranes (Millipore PICMORG50) at 37° C in 5% CO2. The slice culture medium contained (for 500 ml): 394 ml Minimal Essential Medium, 100 ml heat inactivated donor horse serum, 1 mM L-glutamine, 0.01 mg ml−1 insulin, 1.45 ml 5M NaCl, 2 mM MgSO4, 1.44 mM CaCl2, 0.00125% ascorbic acid, 13 mM D-glucose. Wistar rats were housed and bred at the University Medical Center Hamburg-Eppendorf. All procedures were performed in compliance with German law and according to the guidelines of Directive 2010/63/EU. Protocols were approved by the Behörde für Gesundheit und Verbraucherschutz of the City of Hamburg.
2.2 Single cell electroporation
At DIV 13-17, CA1 neurons in rat organotypic hippocampal slice culture were transfected by single-cell electroporation [9]. Thin-walled borosilicate pipettes (∼10 MΩ) were filled with plasmid DNA encoding the GFP-based sensors along with a cytoplasmic red fluorescent protein tdimer2 diluted in intracellular solution to 50 ng µl−1 and 10 ng µl−1, respectively. Pipettes were positioned against neurons under visual control (IR-DIC) and DNA was ejected using an Axoporator 800A (Molecular Devices) with 50 hyperpolarizing pulses (−12 V, 0.5 ms) at 50 Hz. Imaging experiments were conducted 2-4 days after electroporation at room temperature (21-23°) in artificial cerebrospinal fluid (ACSF) containing (in mM) 135 NaCl, 2.5 KCl, 2 CaCl2, 1 MgCl2, 10 Na-HEPES, 12.5 D-glucose, 1.25 NaH2PO4 (pH 7.4).
2.3 sDarken sensor
We developed an improved serotonin darkening 5-HT1A receptor-based sensor named sDarken (darkening 5-HT1A receptor-based sensor). The sensor design is based on dLight1 [4]. Due to its high sensitivity, the native human 5-HT1A receptors was used as the sensing scaffold. We replaced the third intracellular loop of 5-HT1A with a circular mutated form of GFP (cpGFP) from GCaMP6 [10] flanked by mutated linker sequences. Linker length is identical to dLight1.2. sDarken is bright in the unbound form and decreases its fluorescence upon serotonin binding.
2.4 Bone sections
Femoral bone was obtained during autopsy in accordance with the local ethics regulations (PV 3486) [10]. A femoral cross section was cut from the mid-diaphysis using a diamond belt saw (EXAKT Advanced Technologies) and fixed in 3.7% formaldehyde. The sample was dehydrated in an increasing alcohol series and infiltrated with a plastic embedding medium based on glycolmethacrylate (Technovit 7200). Using an automatic grinding system (EXAKT Advanced Technologies), the sample was ground to a thickness of 100 µm. An osteon with an uninterrupted outer border was selected for the analysis.
2.5 Two-photon microscopy
The custom-built two-photon imaging setup was based on an Olympus BX51WI microscope with LUMPlan W-IR2 60× 1.0 NA objective. A galvanometric scanner (6215H, Cambridge Technology) was controlled by the open-source software package ScanImage [11]. Two pulsed Ti:Sapph lasers (MaiTai DeepSee, Spectra Physics) tuned to 930 nm and controlled by electro-optic modulators (350-80, Conoptics) were coupled at orthogonal polarization to excite GFP and tdimer2. In some experiments, we activated one or the other to generate images at orthogonal polarization (Fig. 2). For all other experiments, the polarization of a single Ti:Sapph laser was either rotated by 90° using a half-wave plate (690–1080 nm achromatic; Thorlabs) or split into orthogonally polarized pulse trains using the X-Pol device. Emitted photons were collected through the objective and oil-immersion condenser (1.4 NA, Olympus) with two pairs of photomultiplier tubes (H7422P-40, Hamamatsu) [12]. 560 DXCR dichroic mirrors and 525/50 and 607/70 emission filters (Chroma) were used to separate green and red fluorescence. Excitation light was blocked by short-pass filters (ET700SP-2P, Chroma).
2.6 X-Pol device
The prototype device used in this study is based on the 30 mm cage system (Linos, Thorlabs) and can be rotated around the optical axis. It consists of the following parts (Table 1):
2.7 Alignment procedure
The device was inserted with a 45° tilt before the beam expander and centered on the excitation laser beam using the entrance pinhole (iris diaphragm) and an infrared scope. Axial orientation is not very critical as the device basically acts as a parallel plate. Some angular deviation of the exit beam was noted due to the limited parallelism of the beam splitting cubes (+/- 3 arcmin). This deviation was compensated by adjusting the following beam steering mirrors (periscope). The mirror prism inside the device was coarsely adjusted so that a single output spot was produced on an infrared viewer card held at a distance of about 1 m. Imaging a fluorescent structure in focusing mode, the mirror prism was fine adjusted to achieve maximum image brightness, corresponding to perfect overlap of the PSFs generated by the orthogonally polarized beams. This is best achieved when the person adjusting the mirror prism is able to see the live image on the computer screen. Lastly, we adjusted the pulse pre-compensation (MaiTai DeepSee) to account for the added GVD of 20 mm N-BK7 glass (1400 fs2). Conveniently, rotating the device in its holders to horizontal (0°) or vertical (90°) orientation switches off its beam splitting function, resulting in output polarization (and repetition rate) that is identical to the input beam. In working position (near 45°), some power (∼12%) is lost at the second, beam-combining cube. This is due to imperfect polarization of the reflected beam in the first, beam-splitting cube. For safety, the exit of this light must be blocked, either with a black metal plate or by a Si-photodiode to measure laser power. Due to this power loss in the reflected sub-beam, the ideal 50/50 splitting position is ∼55° rotation, not 45°. The position can be optimized by temporarily blocking the reflected sub-beam inside the device, which should drop the output power to exactly 50%.
2.8 Image analysis
Two-photon image stacks were de-interleaved (red/green channel) and z-projected (maximum intensity projection, MIP). Displayed images are MIPs; magenta and green levels were adjusted separately and linearly (Gamma = 1). Intensity profiles along the x-axis of a rectangular region of interest placed across the apical dendrite were generated with Fiji/ImageJ 1.53c and plotted with GraphPad Prism 8.4.3.
3. Results
We transfected hippocampal CA1 neurons in organotypic slice culture [13] by single-cell electroporation [9] to express the following membrane-bound cpGFP-based sensors together with the red cytoplasmic label tdimer2 [14]: The GPCR-based dopamine sensor dLight1.2[4], the norepinephrine sensor GRABNE1h [7], a newly developed serotonin sensor (sDarken) based on dLight1, and the glutamate sensor iGluSnFR [15]. For comparison, we also expressed GPI-anchored GFP [16] together with tdimer2 in some cells. Coupling two identical Ti:Sapph lasers tuned to 930 nm with a polarizing beam splitter cube allowed us to image the same structure using two different (orthogonal) polarization directions. For better comparison, we selected cells with vertically oriented apical dendrites for imaging.
The dopamine sensor dLight1.2 contains a circularly permuted GFP inserted between transmembrane domain 5 and 6 of the D1 dopamine receptor, orienting the GFP barrel parallel to the plasma membrane. For optical measurements, we added 10 µM dopamine to the extracellular solution to maximize the brightness of dLight1.2. Fluorescence of dLight1.2 was strongly dependent on the direction of laser polarization (Fig. 1(a)-(c)). Dendrites orthogonal to the direction of laser polarization were fluorescent while parallel ones were not. In contrast, polarization direction had no effect on the red fluorescence signal of (freely diffusible) tdimer2. When we compared dLight1.2 fluorescence intensity in response to dopamine application, we observed a strong dopamine-dependent fluorescence increase for dendrite-normal polarization (background-subtracted Δf/f0 = 123%), but no signal for dendrite-parallel polarization (Fig. 1(d)). Thus, using linear polarized excitation, it is not possible to relate the relative changes in dLight1.2 fluorescence to dopamine concentration changes.
Next, we tested sDarken, a genetically encoded sensor for serotonin we recently developed. It is based on the architecture of dLight1.2, but decreases its fluorescence in response to serotonin. Thus, we could visualize bright images of sDarken-expressing CA1 pyramidal cells in the absence of the agonist serotonin (Fig. 2(a)). Again, we observed no membrane-localized green fluorescence when the polarization of the laser was parallel to the dendritic membrane (Fig. 2(b)). Diffuse green fluorescence in the cytoplasm likely results from sensor molecules that were not trafficked to the surface as well as immature (green fluorescent) tdimer2.
To change the polarization of excitation, we used two different methods: We either coupled two separate lasers tuned to the same wavelength at orthogonal polarization, activating one or the other (Fig. 2), or we rotated the polarization of a single laser by 90° using a half-wave plate (Fig. 1, Fig. 3). The results were identical, confirming that the direction of polarization determined which parts of the neuronal membrane were visible.
Interestingly, not all membrane-tethered GFP constructs are sensitive to laser polarization. The glutamate sensor iGluSnFR, which is anchored to the membrane via a single transmembrane segment, did not change its fluorescence intensity when excited at different polarization directions (Fig. 3(a)), and neither did GFP tethered to the membrane via glycosylphosphatidylinositol (GPI-GFP, Fig. 3(b)). We conclude that single-sided anchors leave sufficient degrees of freedom to randomize the angle of the fluorophore with respect to the membrane plane. GRABNE1h is a GPCR-based, high affinity norepinephrine sensor with cpGFP attached to transmembrane segments on both ends, but in contrast to dLight1.2 and sDarken, it retains a large part of the third intracellular loop as a flexible linker. When expressed in CA1 pyramidal cells, GRABNE1h did show a strong fluorescence increase in response to noradrenaline application, but no polarization dependence (Fig. 3(c)).
For quantitative measurements of neurotransmitter concentrations in vivo, the orientation-sensitivity of some of the GPCR-based sensors is a problem. Coupling two lasers tuned to the same wavelength at orthogonal polarization is possible, but not very economical. We designed a compact device that splits the beam of a single laser into two sub-beams of equal power (50/50) but orthogonal polarization, and recombines the sub-beams collinearly (Fig. 4(a)-(c)). The device uses two polarizing beam-splitter cubes with 980 nm antireflection coating. To minimize group velocity dispersion (GVD), we used cubes made of n-BK7, rather than the denser N-SF1 (Flint) glass. For the same reason, we used silver coated prisms to direct the reflected beam through air instead of using total internal reflection in right angle prisms. The device is mounted in circular clamps so it can be rotated around the beam to 45° (for 50/50 splitting) or used horizontally (no beam splitting). At 45° (the ‘on’ position), it works for horizontally and vertically polarized input beams, creating a stream of output of pulses with alternating polarization (45°, 135°). Both horizontal (0°) and vertical (90°) orientation of the device result in an unchanged output beam (‘off’ positions).
We inserted a beam splitting cube in a rotating mount to analyze the average power after the X-Pol device at different polarization directions (Fig. 4(d)). With the X-Pol device in the ‘off’ position (horizontal), the output beam was still horizontally polarized. When the device was tilted around the optical axis, it started splitting the beam into orthogonal components, reaching equal power at all polarization directions at a device tilt angle of ∼55°. In this position, a power loss of ∼12% occurred at the device (at 930 nm), mostly due to the mixed polarization of the reflected beam exiting the first beam splitter cube. The horizontally polarized component is not reflected at the second beam splitting cube; it crosses the cube and is lost. Therefore, slightly more power has to be directed into the reflected path to achieve a balanced output, which in our prototype was reached at 55° tilt angle. For safety reasons, lateral exit of the ‘lost’ beam has to be prevented. In our prototype, it is absorbed by a strip of matte black aluminum. With a side-mounted Si photodiode, it could be utilized to measure the power of the imaging beam.
Although we achieved a flat distribution of laser power across all polarization directions, two-photon excitation of (perfectly aligned) fluorophores would still be expected to show some polarization dependence (Fig. 4(e)). Two-photon excitation depends on the pulse power squared, and molecules oriented at 45° to the two orthogonally polarized pulse trains are excited at twice the frequency, but with half the power per pulse. Note that pulses cannot interact; they have a temporal offset of ∼70 ps (20 mm in air).
We tested the X-Pol device first on the bright serotonin sensor sDarken, which was completely invisible on dendrites aligned with the polarization direction of the laser, but perfectly visible on membranes orthogonal to the polarization direction (Fig. 5(a)). Inserting the device in the beam path at 55° tilt angle achieved the same results than combining two separate lasers with orthogonal polarization: Using the same average laser power, green fluorescence appeared bright along the entire membrane, independent of dendritic orientation (Fig. 5(b)). The intensity of red fluorescence (cytoplasmic tdimer2) is near identical in images 5a and 5b. The less efficient excitation when pulses with half the energy are used at twice the frequency was apparently compensated by the larger fraction of fluorophores available for excitation. Note that the green/red ratio is higher in thinner dendrites as it reflects the surface-to-volume ratio of the compartment.
We tested the X-Pol device also in label-free imaging of thin sections from human bone. Oriented collagen fibers generate a strong second harmonic signal when oriented parallel to the polarization direction of the excitation laser, revealing details of fiber organization below the diffraction limit of the microscope [17]. We used polarization sensitive second harmonic generation (P-SHG) to investigate the lamellar organization of osteons (Fig. 6(a)). In standard configuration, the two-photon microscope did not generate green emission from the sectors of the osteon where its lamellae were orthogonal to the polarization, resulting in two dark sectors. The tilt of the dark sectors in our images reflects a 15° tilt of the galvanometric x-mirror’s axis of rotation (XY Galvo Block, Cambridge Technology), resulting in a 75° polarization angle in the images.
Inserting the X-Pol device into the beam path (Fig. 6(b)) and rotating it to 55° with respect to the laser polarization (Fig. 6(c)) resulted in evenly distributed signals from all sectors of the osteon, removing the artifact produced by linear polarization. To compensate for light loss inside the polarization device, laser power was increased by ∼20% (Fig. 6(c)). We did not see dark sectors between the cardinal polarization directions generated by the X-Pol device (as we predicted from theory, see Fig. 4(e)), most likely because the orientation of collagen fibrils is not parallel to the imaging plane, but at oblique angles to it [18]. In addition, a high-NA objective produces polarization components that are not aligned with the optical axis, reducing the anisotropy in the images [19].
Osteons provide a convenient test sample as their lamellar collagen organization is known to be rotationally symmetric. The strong polarization dependence of SHG images is even more problematic in tissues that have not the same degree of crystalline organization (e.g. skin and tissue), as the absence of signal in such images can be mistakenly interpreted as absence of collagen. Mamillary tumors, for example, are characterized by an increased collagen density, and two-photon SHG imaging has been successfully used to visualize such malignancies [20,21]. As we show, orthogonal polarization excitation makes SHG images easier to interpret, which could be helpful for the investigation and diagnosis of cancer. Analogous polarization-dependent artifacts appear in other kinds of microscopy techniques, e.g. coherent anti-Stokes Raman scattering (CARS) microscopy [22]. Orthogonal polarization could be used to create a hardware implementation of software-based methods designed to mitigate this artifact.
4. Discussion
We report unexpectedly strong effects of polarized excitation in two-photon microscopy of a particular type of GFP-labeled membrane proteins, namely GPCR-based neurotransmitter sensors. No polarization dependence was observed for the soluble red fluorescent protein tdimer2. Similarly, flexible, single-ended tethering, as in GPI-anchored GFP and iGluSnFR, did not result in measurable anisotropy. If the fluorophore can rotate in a cone, its dipole orientation relative to the direction of polarization is randomized. To obtain polarization-sensitive probes with single-ended tether, the linker has to form a continuous α-helix to prevent angular movements [23].
In GPCR-based neurotransmitter sensors, the GFP barrel is inserted into the third intracellular loop and thus tethered at both ends (Fig. 7). This architecture orients the GFP barrel more or less parallel to the cell membrane, depending on the length of the linkers [24]. Based on available structural information [23,25–27], we propose that in highly anisotropic sensors, the dipole moment of cpGFP is oriented normal to the membrane plane (Fig. 7). Interestingly, we could not detect significant polarization dependence of the norepinephrine sensor GRABNE1h. It has been pointed out that in these kind of GRAB sensors, large parts of the third intracellular loop are retained and provide a flexible link to the GFP barrel [24]. Thus, subtle differences in sensor design can have profound effects on the sensitivity to polarization. The effects of adding single or multiple lipid anchors to GFP were previously explored [27,28] and the resulting linear dichroism was most clearly detected using two-photon exaction. In the development of sensors intended for in vivo imaging, typically performed with two-photon microscopes using polarized excitation, isotropic fluorescence should indeed be a selection criterion.
For quantitative neurotransmitter measurements deep in in brain tissue, we propose a simple solution, to alternate pulses of orthogonal polarization. We show that the X-Pol device significantly reduces direction bias in neurons. For specific applications, fluorescence-detected linear dichroism could also be exploited, e.g. to investigate lipid domains or physical interactions between subunits of G proteins [28,29]. However, this works best in spherical cells and requires an active device to rapidly switch laser polarization. For the class of GPCR-based neurotransmitter sensors, the goal is to precisely measure extracellular neurotransmitter concentrations on the entire surface of the neuron. As we demonstrate, this goal is put in jeopardy if linear polarized excitation is used.
In most confocal microscopes, lasers are coupled by conventional single mode optical fibers that do not preserve the polarization state of the light propagating along the fiber. As the sample is excited at arbitrary polarization that also fluctuates over time, we would not expect systematic polarization artifacts in confocal images. Confocal polarizing microscopes have been built for special purposes [28,30,31], but are not in general use for neuroscience applications.
Our prototype X-Pol device was assembled from off-the-shelf components and is therefore inexpensive to reproduce. We would like to point out, however, that precise alignment of the optical components inside the device is important to ensure complete overlap of the two point-spread functions generated by the sub-beams. If the mirror prisms are not well adjusted, loss of resolution or even double images may occur. Aligning the device to the center of the beam is less critical as long as beam clipping is avoided. The added dispersion of 20 mm N-BK7 glass (GVD = 70 fs2/mm) was readily compensated by pre-chirping (MaiTai DeepSee, SpectraPhysics). Flint glass (N-SF1), which is typically used for beam splitting cubes, induces more GVD (247 fs2/mm). To further reduce GVD, broadband thin plate beam splitters could be arranged at Brewster’s angle in a chevron design, again mounted with axial tilt to achieve 50/50 splitting [32]. Instead of rotating the X-Pol device, the polarization of the incoming beam could be tilted with a half-wave plate. However, adding an achromatic half-wave plate roughly doubles the cost of the device without obvious benefits. It is also possible to generate orthogonal polarization by a birefringent crystal, e.g. yttrium vanadate (YVO4), but this creates a lateral displacement of the two output beams.
The X-Pol device doubles the pulse repetition rate while cutting the energy of individual pulses in half. This effect should require an increase in average power (by 20.5) to achieve the same image intensity. On the other hand, more fluorophores can be excited when using orthogonal pulse trains, increasing the effective concentration of fluorescent proteins. Examining the red fluorescence of tdimer2, we found that these two effects were in balance, and no increase in average power was required when imaging with the X-Pol device at 55°. Green fluorescence was independent of dendritic orientation with the X-Pol device, but less bright than in optimally oriented dendrites excited with linear polarized light. This was expected, as the weaker excitation by the split pulses was not compensated by recruitment of additional green fluorophores. Increased repetition rates have been reported to reduce bleaching in biological preparations [33–35] but require higher average power, which can affect cellular physiology by local heating [36]. Some of the beneficial effects shown by Ji et al. [33], using a more complex pulse splitter design, may have been due to orthogonally polarized output pulses and in consequence, recruitment of additional fluorophores.
An intuitive solution to overcome linear polarization artifacts is to use circularly-polarized light by employing an achromatic quarter-wave plate. This solution would indeed work for one-photon illumination, however it is not effective for two-photon excitation. For efficient two-photon absorption, the sum of angular momenta of the absorbed photons is required to be zero [37,38].
In addition to optical measurements of neurotransmitter concentration with GPCR-based sensors, excitation with orthogonal polarization could also be advantageous for two-photon excitation of channelrhodopsin as well as opsin-derived sensors of membrane voltage [39]. Indeed, polarization-dependence of genetically encoded voltage sensors has recently been reported [40]. In membranes, cytoskeletal labeling and in fixed tissue, where the rotation of fluorophores is slowed down or locked, conventional linearly polarized two-photon microscopes continuously excite the same subset of fluorophores, leading to anisotropic photobleaching [41]. Orthogonally polarized pulse trains double the number of excitable fluorophores, increasing brightness (compared to excitation pulses of the same energy and repetition frequency polarized in one direction) or reducing bleaching when imaging with adapted laser power. Based on our experience, we predict that two-photon microscopy will follow the lead of confocal microscopy and use mixed polarization as default imaging mode.
Funding
H2020 European Research Council (714762, 951515); Deutsche Forschungsgemeinschaft (178316478, 278170285, 315380903, 394482524, 408367170, 458878282, 122679504).
Acknowledgments
The authors would like to thank Iris Ohmert, Jan Schröder and Kathrin Sauter for excellent technical assistance. pAAV-hSyn-dLight1.2 was a gift from Lin Tian (Addgene plasmid # 111068), pAAV-hSyn-GRAB_DA1 h (Addgene plasmid # 113050) and pAAV-hSyn-GRAB_NE1 h (Addgene plasmid # 123309) were gifts from Yulong Li, iGluSnFR was a gift from Loren Looger (Addgene plasmid # 41732), GPI-GFP was a gift from Martin Heine, tdimer2 was a gift from Roger Y. Tsien.
Disclosures
MP (P), TGO (P).
Data availability
Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.
References
1. S. T. Hess, E. D. Sheets, A. Wagenknecht-Wiesner, and A. A. Heikal, “Quantitative analysis of the fluorescence properties of intrinsically fluorescent proteins in living cells,” Biophys. J. 85(4), 2566–2580 (2003). [CrossRef]
2. A. H. A. Clayton, Q. S. Hanley, D. J. Arndt-Jovin, V. Subramaniam, and T. M. Jovin, “Dynamic fluorescence anisotropy imaging microscopy in the frequency domain (rFLIM),” Biophys. J. 83(3), 1631–1649 (2002). [CrossRef]
3. G. de Vito, P. Ricci, L. Turrini, V. Gavryusev, C. Müllenbroich, N. Tiso, F. Vanzi, L. Silvestri, and F. S. Pavone, “Effects of excitation light polarization on fluorescence emission in two-photon light-sheet microscopy,” Biomed. Opt. Express 11(8), 4651 (2020). [CrossRef]
4. T. Patriarchi, J. R. Cho, K. Merten, M. W. Howe, A. Marley, W.-H. Xiong, R. W. Folk, G. J. Broussard, R. Liang, M. J. Jang, H. Zhong, D. Dombeck, M. von Zastrow, A. Nimmerjahn, V. Gradinaru, J. T. Williams, and L. Tian, “Ultrafast neuronal imaging of dopamine dynamics with designed genetically encoded sensors,” Science 360(6396), eaat4422 (2018). [CrossRef]
5. F. Sun, J. Zeng, M. Jing, J. Zhou, J. Feng, S. F. Owen, Y. Luo, F. Li, H. Wang, T. Yamaguchi, Z. Yong, Y. Gao, W. Peng, L. Wang, S. Zhang, J. Du, D. Lin, M. Xu, A. C. Kreitzer, G. Cui, and Y. Li, “A genetically encoded fluorescent sensor enables rapid and specific detection of dopamine in flies, fish, and mice,” Cell 174(2), 481–496.e19 (2018). [CrossRef]
6. M. Jing, Y. Li, J. Zeng, P. Huang, M. Skirzewski, O. Kljakic, W. Peng, T. Qian, K. Tan, J. Zou, S. Trinh, R. Wu, S. Zhang, S. Pan, S. A. Hires, M. Xu, H. Li, L. M. Saksida, V. F. Prado, T. J. Bussey, M. A. M. Prado, L. Chen, H. Cheng, and Y. Li, “An optimized acetylcholine sensor for monitoring in vivo cholinergic activity,” Nat. Methods 17(11), 1139–1146 (2020). [CrossRef]
7. J. Feng, C. Zhang, J. E. Lischinsky, M. Jing, J. Zhou, H. Wang, Y. Zhang, A. Dong, Z. Wu, H. Wu, W. Chen, P. Zhang, J. Zou, S. A. Hires, J. J. Zhu, G. Cui, D. Lin, J. Du, and Y. Li, “A genetically encoded fluorescent sensor for rapid and specific in vivo detection of norepinephrine,” Neuron 102(4), 745–761.e8 (2019). [CrossRef]
8. J. Wan, W. Peng, X. Li, T. Qian, K. Song, J. Zeng, F. Deng, S. Hao, J. Feng, P. Zhang, Y. Zhang, J. Zou, S. Pan, M. Shin, B. J. Venton, J. J. Zhu, M. Jing, M. Xu, and Y. Li, “A genetically encoded sensor for measuring serotonin dynamics,” Nat. Neurosci. 24(5), 746–752 (2021). [CrossRef]
9. J. S. Wiegert, C. E. Gee, and T. G. Oertner, “Single-cell electroporation of neurons,” Cold Spring Harb. Protoc. 2017(2), pdb.prot094904 (2017). [CrossRef]
10. P. Milovanovic, E. A. Zimmermann, C. Riedel, A. vom Scheidt, L. Herzog, M. Krause, D. Djonic, M. Djuric, K. Püschel, M. Amling, R. O. Ritchie, and B. Busse, “Multi-level characterization of human femoral cortices and their underlying osteocyte network reveal trends in quality of young, aged, osteoporotic and antiresorptive-treated bone,” Biomaterials 45, 46–55 (2015). [CrossRef]
11. T. A. Pologruto, B. L. Sabatini, and K. Svoboda, “ScanImage: flexible software for operating laser scanning microscopes,” Biomed Eng Online 2(1), 13 (2003). [CrossRef]
12. T. G. Oertner, “Functional imaging of single synapses in brain slices,” Exp. Phys. 87(6), 733–736 (2002). [CrossRef]
13. C. E. Gee, I. Ohmert, J. S. Wiegert, and T. G. Oertner, “Preparation of slice cultures from rodent hippocampus,” Cold Spring Harb. Protoc. 2017(2), pdb.prot094888 (2017). [CrossRef]
14. R. E. Campbell, O. Tour, A. E. Palmer, P. A. Steinbach, G. S. Baird, D. A. Zacharias, and R. Y. Tsien, “A monomeric red fluorescent protein,” Proc. Natl. Acad. Sci. U. S. A. 99(12), 7877–7882 (2002). [CrossRef]
15. J. S. Marvin, B. Scholl, D. E. Wilson, K. Podgorski, A. Kazemipour, J. A. Müller, S. Schoch, F. J. U. Quiroz, N. Rebola, H. Bao, J. P. Little, A. N. Tkachuk, E. Cai, A. W. Hantman, S. S.-H. Wang, V. J. DePiero, B. G. Borghuis, E. R. Chapman, D. Dietrich, D. A. DiGregorio, D. Fitzpatrick, and L. L. Looger, “Stability, affinity, and chromatic variants of the glutamate sensor iGluSnFR,” Nat. Methods 15(11), 936–939 (2018). [CrossRef]
16. G. Kondoh, X. H. Gao, Y. Nakano, H. Koike, S. Yamada, M. Okabe, and J. Takeda, “Tissue-inherent fate of GPI revealed by GPI-anchored GFP transgenesis,” FEBS Lett. 458(3), 299–303 (1999). [CrossRef]
17. J. C. Mansfield, V. Mandalia, A. Toms, C. Peter Winlove, and S. Brasselet, “Collagen reorganization in cartilage under strain probed by polarization sensitive second harmonic generation microscopy,” J. R. Soc. Interface 16(150), 20180611 (2019). [CrossRef]
18. K. E. Stockhausen, M. Qwamizadeh, E. M. Wölfel, H. Hemmatian, I. A. K. Fiedler, S. Flenner, E. Longo, M. Amling, I. Greving, R. O. Ritchie, F. N. Schmidt, and B. Busse, “Collagen fiber orientation is coupled with specific nano-compositional patterns in dark and bright osteons modulating their biomechanical properties,” ACS Nano 15(1), 455–467 (2021). [CrossRef]
19. D. Axelrod, “Carbocyanine dye orientation in red cell membrane studied by microscopic fluorescence polarization,” Biophys. J. 26(3), 557–573 (1979). [CrossRef]
20. T. Meyer, N. Bergner, C. Bielecki, C. Krafft, D. Akimov, B. F. M. Romeike, R. Reichart, R. Kalff, B. Dietzek, and J. Popp, “Nonlinear microscopy, infrared, and Raman microspectroscopy for brain tumor analysis,” J. Biomed. Opt. 16(2), 021113 (2011). [CrossRef]
21. P. P. Provenzano, D. R. Inman, K. W. Eliceiri, J. G. Knittel, L. Yan, C. T. Rueden, J. G. White, and P. J. Keely, “Collagen density promotes mammary tumor initiation and progression,” BMC Med. 6(1), 11 (2008). [CrossRef]
22. Y. Fu, T. B. Huff, H.-W. Wang, J.-X. Cheng, and H. Wang, “Ex vivo and in vivo imaging of myelin fibers in mouse brain by coherent anti-Stokes Raman scattering microscopy,” Opt. Express 16(24), 19396 (2008). [CrossRef]
23. A. Sugizaki, K. Sato, K. Chiba, K. Saito, M. Kawagishi, Y. Tomabechi, S. B. Mehta, H. Ishii, N. Sakai, M. Shirouzu, T. Tani, and S. Terada, “POLArIS, a versatile probe for molecular orientation, revealed actin filaments associated with microtubule asters in early embryos,” Proc. Natl. Acad. Sci. 118(11), e2019071118 (2021). [CrossRef]
24. L. Ravotto, L. Duffet, X. Zhou, B. Weber, and T. Patriarchi, “A Bright and Colorful Future for G-Protein Coupled Receptor Sensors,” Front. Cell. Neurosci. 14, 67 (2020). [CrossRef]
25. F. I. Rosell and S. G. Boxer, “Polarized absorption spectra of green fluorescent protein single crystals: Transition dipole moment directions,” Biochemistry 42(1), 177–183 (2003). [CrossRef]
26. Y. Chen, X. Song, S. Ye, L. Miao, Y. Zhu, R. G. Zhang, and G. Ji, “Structural insight into enhanced calcium indicator GCaMP3 and GCaMPJ to promote further improvement,” Protein Cell 4(4), 299–309 (2013). [CrossRef]
27. R. D. Roorda, T. M. Hohl, R. Toledo-Crow, and G. Miesenböck, “Video-rate nonlinear microscopy of neuronal membrane dynamics with genetically encoded probes,” J. Neurophysiol. 92(1), 609–621 (2004). [CrossRef]
28. J. Lazar, A. Bondar, S. Timr, and S. J. Firestein, “Two-photon polarization microscopy reveals protein structure and function,” Nat. Methods 8(8), 684–690 (2011). [CrossRef]
29. A. Gasecka, T. J. Han, C. Favard, B. R. Cho, and S. Brasselet, “Quantitative imaging of molecular order in lipid membranes using two-photon fluorescence polarimetry,” Biophys. J. 97(10), 2854–2862 (2009). [CrossRef]
30. I. I. Smalyukh, S. V. Shiyanovskii, and O. D. Lavrentovich, “Three-dimensional imaging of orientational order by fluorescence confocal polarizing microscopy,” Chem. Phys. Lett. 336(1-2), 88–96 (2001). [CrossRef]
31. C. E. Bigelow, D. L. Conover, and T. H. Foster, “Confocal fluorescence spectroscopy and anisotropy imaging system,” Opt. Lett. 28(9), 695–697 (2003). [CrossRef]
32. D. J. Dummer, S. G. Kaplan, L. M. Hanssen, A. S. Pine, and Y. Zong, “High-quality Brewster’s angle polarizer for broadband infrared application,” Appl. Opt. 37(7), 1194 (1998). [CrossRef]
33. N. Ji, J. C. Magee, and E. Betzig, “High-speed, low-photodamage nonlinear imaging using passive pulse splitters,” Nat. Methods 5(2), 197–202 (2008). [CrossRef]
34. F. Niu, R. Wu, D. Wu, D. Gou, L. Feng, L. Chen, Z. Zhang, and A. Wang, “Photobleaching of ultrashort pulses with different repetition rates in two-photon excitation microscopy,” Laser Phys. 29(4), 046001 (2019). [CrossRef]
35. J. Mütze, V. Iyer, J. J. MacKlin, J. Colonell, B. Karsh, Z. Petrášek, P. Schwille, L. L. Looger, L. D. Lavis, and T. D. Harris, “Excitation spectra and brightness optimization of two-photon excited probes,” Biophys. J. 102(4), 934–944 (2012). [CrossRef]
36. E. Schmidt and M. Oheim, “Infrared excitation induces heating and calcium microdomain hyperactivity in cortical astrocytes,” Biophys. J. 119(11), 2153–2165 (2020). [CrossRef]
37. K. D. Bonin and T. J. McIlrath, “Two-photon electric-dipole selection rules,” J. Opt. Soc. Am. B 1(1), 52 (1984). [CrossRef]
38. I. Micu, C. Brideau, L. Lu, and P. K. Stys, “Effects of laser polarization on responses of the fluorescent Ca2+ indicator X-Rhod-1 in neurons and myelin,” Neurophotonics 4(2), 025002 (2017). [CrossRef]
39. J. M. Kralj, A. D. Douglass, D. R. Hochbaum, D. Maclaurin, and A. E. Cohen, “Optical recording of action potentials in mammalian neurons using a microbial rhodopsin,” Nat. Methods 9(1), 90–95 (2012). [CrossRef]
40. W. Bloxham, D. Brinks, S. Kheifets, and A. E. Cohen, “Linearly polarized excitation enhances signals from fluorescent voltage indicators,” bioRxiv 2021.07.21.453006 (2021).
41. K. Sharma, D. K. Fong, and A. M. Craig, “Postsynaptic protein mobility in dendritic spines: long-term regulation by synaptic NMDA receptor activation,” Mol. Cell. Neurosci. 31(4), 702–712 (2006). [CrossRef]