Understanding the relationship between visual-angle and eye-angle for reliable determination of the field-of-view in ultra-wide field fundus photography

Open Access Open Access

Abstract

Visual-angle has been used as the conventional unit to determine the field-of-view (FOV) in traditional fundus photography. Recently emerging usage of eye-angle as the unit in wide field fundus photography creates confusion about FOV interpretation in instrumentation design and clinical application. This study aims to systematically derive the relationship between the visual-angle θv and eye-angle θe, and thus to enable reliable determination of the FOV in wide field fundus photography. FOV conversion ratio θev, angular conversion ratio Δθe/Δθv, retinal conversion ratio Δd/Δθv, retinal distance and area are quantitatively evaluated. Systematic analysis indicates that reliable conversion between the θv and θe requires determined nodal point and spherical radius of the eye; and the conversion ratio is not linear from the central field to peripheral region. Based on the eye model with average parameters, both angular conversion (Δθe/Δθv) and retinal conversion (Δd/Δθv) ratios are observed to have a 1.51-fold difference at the central field and far peripheral region. A conversion table, including θev, Δθe/Δθv, Δd/Δθv, retinal area and percentage ratio, is created for reliable assessment of imaging systems with variable FOV.

© 2021 Optical Society of America under the terms of the OSA Open Access Publishing Agreement

1. Introduction

Fundus photography is important for eye disease screening, diagnosis and treatment assessment. In traditional fundus cameras, visual-angle θv is used for quantitative evaluation of the field-of-view (FOV). For visual perception, the visual-angle is the angle of a target subtends at the eye. Under emmetropic condition, the normal eye can image distant objects precisely in focus into the retina [1]. At the nodal point of the eye, the angle subtended by the target image at the retina is identical to the visual-angle θv (Fig. 1(a)). In other word, the incident light from each direction corresponds to a focused point at the retina. For fundus photography, the purpose of retinal imager is to collect light from the back of the eye. According to the principle of reversibility of light, the light coming from each retinal point corresponds to a specific direction of the light leaving the eye (Fig. 1(b)). The light angle range detectable by the fundus imager defines the FOV. Therefore, the FOV and also angular resolution of a fundus imager can be readily calibrated in the unit of visual-angle θv, by imaging a target placed at a certain distance [2].

 figure: Fig. 1.

Fig. 1. (a) Illustration of visual-angle and eye-angle. (b) Illustration of FOV.

Download Full Size | PDF

The FOV of traditional fundus cameras is typically limited at 30° to 60° visual-angle [3]. In principle, eye diseases may affect any part of the retina. The retina covers 72% of the inside of the eye globe [4], or roughly 180° visual field. In order to achieve the necessary view field coverage, mydriatic ETDRS 7-field photography for screening diabetic retinopathy (DR) has been developed based on the use of the traditional fundus camera with a 30° visual-angle FOV [5]. Conventional binocular indirect ophthalmoscopy (BIO), which is a time-consuming procedure that is painful for the patient and stressful for the ophthalmologist, is still the gold standard for reliable screening of retinopathy of prematurity (ROP) screening [68]. In order to foster digital fundus photography for eye disease detection and treatment management, there are active efforts to develop wide field imaging systems [918]. Instead of using traditional visual-angle to specify the FOV, some of these wide field systems such as the Optos (Optos, Marlborough, MA), Retcam (Clarity Medical Systems, Pleasanton, CA) and Clarus 500 (Carl Zeiss Meditec, Inc., Dublin, CA) systems use eye-angle θe to specify the FOV. As shown in Fig. 1(a), the eye-angle θe is the angle subtended by the imaged retinal region at the spherical center of the eye.

Given the factor of that not all researchers, ophthalmologists and photographers are aware of the difference between visual-angle and eye-angle, the mixed usages of these two terms have produced confusions about the visual field interpretation. In principle, the visual-angle θv and eye-angle θe can be converted to each other. By considering the refractive difference between the air and the eye which is assumed with 11 mm spherical radius, a conversion factor ∼1.5 was derived to convert the visual and eye angles as θe=1.5θv in a white paper [19]. The empirical factor ∼1.5 has been adopted in recent publications [1316,20,21]. However, without considering detailed properties such as optical powers of the cornea and crystalline lens, the simplified model in the white paper [19] is not sufficient to reliably correlate the visual-angle and eye-angle values over the visual field. In other words, the empirical factor ∼1.5 is reasonable only for a certain angle range nearby the eye axis, and can produce overestimated eye-angle number when the visual-angle increases to the peripheral retina. In this article, we report a systematic analysis of the relationship between visual-angle and eye-angle from the optical axis to the ora serrata, i.e. the far peripheral region of the retina.

2. Materials and methods

Figure 2(a) illustrates the eye model used for following analysis. The parameters in Fig. 2(a) have been popularly reported in previous publications [1,22,23], although slight difference may exist among different studies. Because every eye is intrinsically unique, these parameters represent average values for the purpose of generalization [1]. As shown in Fig. 2(a), the cornea is the mostly outer layer of the eye. The cornea is a spherical section with an anterior radius of curvature of 7.8 mm, posterior radius of curvature of 6.7 mm, thickness of 0.55 mm, and refractive index of 1.3771. The crystalline lens is a biconvex lens with 10 mm and −6 mm radii of curvature for the anterior and posterior surfaces. The refractive indices of the aqueous, crystalline lens, and vitreous humors are 1.3374, 1.42 and 1.336, respectively. The distance between the posterior surface of the cornea to the anterior surface to the crystalline lens is 3.1 mm, and the thickness of the crystalline lens is 4.0 mm. The radius of posterior eye is 11 mm.

 figure: Fig. 2.

Fig. 2. (a) Schematic diagram of the eye model. (b) Ray tracing through the eye model. (c) Illustration of the visual-angle and eye-angle.

Download Full Size | PDF

The nodal point is the location in the eye where light entering or leaving the eye and passing through is undeviated. This allows the same visual-angle to be used to determine the retinal region corresponding to the target in external space (Fig. 1). The nodal point of the eye is known to be nearby the posterior surface of the crystalline lens, i.e., anterior vitreous [2426] (Fig. 2(a)). Using the parameters in Fig. 2(a), a ZEMAX simulation confirms the lens posterior as the nodal point in the eye (Fig. 2(b)). Figure 2(c) is used to explain the relationship between visual-angle θv and eye-angle θe. In order to simplify the discussion, we only use the space above the optical axis to illustrate the half visual-angle θv′ and half eye-angle θe′. The same relationship can be applied to the bottom half of the space. In other words, the FOV can be simply doubled compared to the half space values in Fig. 2(c), i.e., θv=2θv′ and θe=2θe′.

As shown in Fig. 2(c), the half eye-angle θe$^{\prime}$ can be estimated as follows:

$${\theta _e}^{\prime} = {\theta _v}^{\prime} + {\theta _r}^{\prime}$$
where ${\theta _v}^{\prime}$ is the half visual-angle, and
$${\theta _r}^{\prime} = {\sin ^{ - 1}}(\frac{{NO}}{{OR}}\sin {\theta _v}^{\prime})$$

Substitute NO=5.6 mm and OR=11 mm into Eq. (2)

$${\theta _r}^{\prime} = {\sin ^{ - 1}}(0.51\sin {\theta _v}^{\prime})$$

Substitute Eq. (3) into Eq. (1)

$${\theta _e}^{\prime} = {\theta _v}^{\prime} + {\sin ^{ - 1}}(0.51\sin {\theta _v}^{\prime})$$

If the FOV is known in the unit of eye-angle, corresponding retinal distance can be estimated as a spherical arc (Fig. 3).

$$d = 2\pi r{\theta _e}^{\prime}/360\; = \pi r{\theta _e}^{\prime}/180\; $$
where r is the eyeball radius.

 figure: Fig. 3.

Fig. 3. Illustration of retinal area estimation as a spherical cap.

Download Full Size | PDF

Substitute Eq. (4) into 5,

$$d = \pi r({\theta _v}^{\prime} + {\sin ^{ - 1}}(0.51\sin {\theta _v}^{\prime})/180$$

Corresponding retinal area A can be estimated as a spherical cap (Fig. 3).

$$A = 2\pi r\textrm{h}$$
and h is the height of the spherical cap corresponding to the visual field. The h can be calculated as
$$h = r(1 - \cos {\theta _e}^{\prime})$$

Substitute Eq. (8) into Eq. (7)

$$A = 2\pi {r^2}({1 - \cos {\theta_e}^{\prime}} )$$

3. Results

According to Eq. (4), Fig. 4(a) show the relationship between visual-angle θv and eye-angle θe. Moreover, Fig. 4(b) illustrate the FOV conversion ratio θev and angular conversion ratio Δθe/Δθv from the central to peripheral regions. As shown in Fig. 4(a), the relationship between visual-angle and eye-angle FOVs is not linear. Figure 4(b) confirm the FOV conversion ratio θev is ∼1.5 for FOV values within 75°, but the factor gradually decreases to 1.34 for large angle FOV. The 180° visual-angle FOV, which is roughly the total visual field, corresponds to 241° eye-angle FOV (Fig. 4(a)).

 figure: Fig. 4.

Fig. 4. (a) The relationship between visual-angle θv and eye-angle θe. (b) Conversion ratio ${\theta _e}/{\theta _v}$. (c) ${\varDelta\theta _e}/{\varDelta\theta _v}$ at different visual field.

Download Full Size | PDF

In order to further evaluate the angular conversion factor Δθe/Δθv, the first order derivation of the Eq. (4) is also implemented. As shown in Fig. 4(b), the angular conversion factor Δθe/Δθv is ∼1.51 at the central field, i.e., nearby the optical axis of the eye. The conversion factor can be reduced up to 1.0 at the far periphery, i.e. the ora serrata.

Equation (5) reveals that the imaged retinal distance d is linearly proportional to the eye-angle value. Assume r=11 mm, the retinal conversion ratio Δd/Δθe can be estimated at 192 µm per eye-angle degree, which is constant over the whole visual field. In contrary, Eq. (6) reveals that the imaged retinal distance d is not linearly proportional to the visual-angle value. According to Eq. (6), Fig. 5(a) show the relationship between the retinal distance d and visual-angle θv. At the center of the visual field, by considering the angular conversion factor Δθe/Δθv 1.51 (Fig. 4), 290 µm per visual-angle degree can be estimated. The 290 µm per visual-angle degree is reasonably consistent to the popularly known 288 µm per visual-angle degree in previous publications [4]. However, the angular conversion factor is gradually reduced, up to 192 µm/degree at the far peripheral visual field (Fig. 5(b)).

 figure: Fig. 5.

Fig. 5. (a) Relationship between retinal distance d and visual-angle θv. (b)$\; \varDelta d/{\varDelta\theta _v}$ at different visual field.

Download Full Size | PDF

Figure 6 shows the relationships of visual-angle FOV (Fig. 6(a)) and eye-angle FOV (Fig. 6(b)) to retinal area and percentage ratio. As shown in Fig. 6(a), the 180° visual-angle, i.e., 241° eye-angle, corresponds to 1,148 mm2, which is reasonably close to previously reported retinal area 1,094 mm2 [4] and 1,134 mm2 [27].

 figure: Fig. 6.

Fig. 6. (a) Relationship between retinal area and visual-angle FOV. (b) Relationship between retinal area and eye-angle FOV.

Download Full Size | PDF

For easy reference, Table 1 shows the representative values of visual-angle, eye-angle, conversion factor and retinal area, corresponding to variable FOVs. 30° and 45° visual-angle FOVs are mostly popular for traditional fundus cameras. As shown in Table 1, the 30° visual-angle corresponds to a 45.2° eye-angle; and the 45° visual-angle corresponds to a 67.5° eye-angle. A 145° visual-angle corresponds to a 203° eye-angle FOV.

Tables Icon

Table 1. Conversion ratios of visual-angle, eye-angle, retinal dimension corresponding to visual field

4. Discussion

In summary, the relationship between visual-angle θv and eye-angle θe has been derived based on the eye model. As shown in Fig. 2(c) and Eqs. (14), the conversion between visual-angle θv and eye-angle θe dependent on the distance between the nodal point N to spherical center O and the radius r of the eye. With the average values, i.e., 5.6 mm from N to O and 11 mm radius, the relationship between visual-angle θv and eye-angle θe has quantitatively evaluated. The relationship between visual-angle and eye-angle is not linear (Fig. 4). The FOV conversion factor θve is changed from 1.51 at the central field, i.e., nearby the optical axis, to 1.34 at the far peripheral field, i.e. ora serratta. Similarly, the angular conversion factor Δθv/Δθe is changed from 1.51 to 1.0 at the far peripheral field.

The empirical factor FOV conversion factor ∼1.5 has been used to convert visual-angle θv into eye-angle θe in previous publications [1316,18,19]. As shown in Table 1 and Fig. 4, the FOV conversion factor 1.5 is reasonable for visual-angle θv<80°. Therefore, for traditional fundus cameras with typical FOV of 30° or 45° visual-angle, the conversion θe=1.5θv is appropriate. The 30° and 45° visual-angles correspond to 45.2° and 67.5° eye-angles, respectively. The determined FOV in the unit of eye-angle can be readily used to estimate corresponding retinal distance and area. Individual snapshot 30° and 45° visual-angle FOVs correspond to 5.1% and 11.2% of the whole retinal area, respectively. For the wide field imaging systems with 60–80° visual-angle FOV [15,16,21], it is also acceptable to use the empirical factor ∼1.5 to estimate eye-angle FOV.

However, for ultra-wide field imaging systems [13,14], direct use of the empirical conversion factor 1.5 might produce overestimated FOV in the unit of eye-angle, and thus overestimated retinal area. For example, the 180° visual-angle FOV corresponds to 241.3° in Table 1, with conversion factor 1.34 derived symmetrically. The 241° eye-angle FOV corresponds to the whole retinal area of 1,148 mm2, which is reasonably close to previously reported total retinal area of 1,134 mm2 [27] or 1,094 mm2 [4]. However, if the empirical factor 1.5 is applied, the 180° visual-angle corresponds to 270°. Using the Eq. (7), the 270° eye-angle corresponds to a retinal area of 1298 mm2, which is ∼15% overestimated compared to the reported 1,134 mm2 [27] or 1,094 mm2 [4]. Moreover, 288 µm per visual-angle has been well established to estimate the retinal dimension [4]. Using the conversion factor 1.5, the Eq. (6) can perfectly match to 288 µm per visual-angle at the central field. However, for peripheral retina, As the conversion factor is gradually reduced up to 192 µm per degree at the far peripheral field (Fig. 5). This is consistent to the effect of the conversion factors on the eye-angle (Fig. 4) and retinal area (Fig. 6) estimations.

We are aware of that the retinal area using a spherical cap in this article is a simplified approximation, with the FOV centered along the visual axis. For non-central fixation images, accurate estimation of the retinal area will need to consider the nonlinearity of the conversion factors from the visual-angle to eye-angle at both lateral and vertical dimensions. Moreover, the spherical model in this article is only applicable to the emmetropic condition. For myopic or hyperopic eyes, precise conversion between the visual-angle and eye-angle will need to consider the elliptical shape of the eye and extra optics required for ametropic compensation.

5. Conclusion

Reliable conversion between the visual-angle θv and eye-angle θe requires determined nodal point and spherical radius of the eye. Because each eye can be unique, it is difficult to have accurate information of the eye dimension and nodal point. For the consideration of generalization, traditional visual-angle can be a simple solution to specify the FOV of the fundus imager. Using the model eye with average parameters, FOV conversion ratio θev, angular conversion ratio Δθe/Δθv, and retinal conversion ratio (Δd/Δθv) can be quantitatively evaluated. These conversion ratios are not linear over the visual field. The FOV conversion ratio θev can be changed from 1.51 in narrow field systems, to 1.34 in ultra-wide field imagers. Both angular conversion (Δθe/Δθv) and retinal conversion (Δd/Δθv) ratio can have a 1.51-fold difference at the central and far peripheral regions. The empirical factor θev ≈1.5 is appropriate for imaging systems with visual-angle FOV < 80°. However, its use in ultra-wide field imagers may provide overestimated eye-angle FOV, and then retinal area.

Funding

Richard and Loan Hill Endowment; Research to Prevent Blindness; National Eye Institute (P30 EY001792, R01 EY030101, R01EY023522, R01EY029673, R01EY030842, R44 EY028786).

Disclosures

The authors declare that there are no conflicts of interest related to this article.

Data availability

All relative data are presented in this article. No other data were generated or analyzed in the presented research.

References

1. P. Artal, “Optics of the eye and its impact in vision: a tutorial,” Adv. Opt. Photon. 6(3), 340–367 (2014). [CrossRef]  

2. “Ophthalmic instruments-Fundus cameras-International standard (ISO 10940: 2009).”

3. K. V. Chalam, V. S. Brar, and R. Keshavamurthy, “Evaluation of modified portable digital camera for screening of diabetic retinopathy,” Ophthalmic. Res. 42(1), 60–62 (2009). [CrossRef]  

4. H. Kolb, “Facts and figures concerning the human retina,” https://www.ncbi.nlm.nih.gov/books/NBK11556/ (2007, last access on June 3, 2021).

5. D. A. Salz and A. J. Witkin, “Imaging in diabetic retinopathy,” Middle East Afr. J. Ophthalmol. 22(2), 145–150 (2015). [CrossRef]  

6. C. Dhaliwal, E. Wright, C. Graham, N. McIntosh, and B. W. Fleck, “Wide-field digital retinal imaging versus binocular indirect ophthalmoscopy for retinopathy of prematurity screening: a two-observer prospective, randomised comparison,” Br. J. Ophthalmol. 93(3), 355–359 (2009). [CrossRef]  

7. W. L. Jones and ScienceDirect, Peripheral Ocular Fundus (Elsevier Butterworth-Heinemann St. Louis, 2007).

8. J. P. Greenberg, L. Sun, and R. M. Lieberman, “Advances in retinal imaging,” Advances in Ophthalmology and Optometry (2018).

9. A. Manivannan, J. Plskova, A. Farrow, S. McKay, P. F. Sharp, and J. V. Forrester, “Ultra-wide-field fluorescein angiography of the ocular fundus,” Am. J. Ophthalmol. 140(3), 525–527 (2005). [CrossRef]  

10. T. Hirano, A. Imai, H. Kasamatsu, S. Kakihara, Y. Toriyama, and T. Murata, “Assessment of diabetic retinopathy using two ultra-wide-field fundus imaging systems, the Clarus and Optos systems,” BMC Ophthalmol. 18(1), 332 (2018). [CrossRef]  

11. A. Olvera-Barrios, T. F. Heeren, K. Balaskas, R. Chambers, L. Bolter, A. Tufail, C. Egan, and J. Anderson, “Comparison of true-colour wide-field confocal scanner imaging with standard fundus photography for diabetic retinopathy screening,” Br J Ophthalmol 104, 1579–1584 (2020). [CrossRef]  

12. T. H. Fung, I. H. Yusuf, L. M. Smith, J. Brett, L. Weston, and C. K. Patel, “Outpatient ultra wide-field intravenous fundus fluorescein angiography in infants using the Optos P200MA scanning laser ophthalmoscope,” Br. J. Ophthalmol. 98(3), 302–304 (2014). [CrossRef]  

13. D. Toslak, T. Son, M. K. Erol, H. Kim, T. H. Kim, R. V. P. Chan, and X. Yao, “Portable ultra-widefield fundus camera for multispectral imaging of the retina and choroid,” Biomed. Opt. Express 11(11), 6281–6292 (2020). [CrossRef]  

14. D. Toslak, F. Chau, M. K. Erol, C. Liu, R. V. P. Chan, T. Son, and X. Yao, “Trans-pars-planar illumination enables a 200 degrees ultra-wide field pediatric fundus camera for easy examination of the retina,” Biomed. Opt. Express 11(1), 68–76 (2020). [CrossRef]  

15. D. Toslak, C. Liu, M. N. Alam, and X. Yao, “Near-infrared light-guided miniaturized indirect ophthalmoscopy for nonmydriatic wide-field fundus photography,” Opt. Lett. 43(11), 2551–2554 (2018). [CrossRef]  

16. D. Toslak, A. Ayata, C. Liu, M. K. Erol, and X. Yao, “Wide-field smartphone fundus video camera based on miniaturized indirect ophthalmoscopy,” Retina 38(2), 438–441 (2018). [CrossRef]  

17. T. A. Nakagawa and R. Skrinska, “Improved documentation of retinal hemorrhages using a wide-field digital ophthalmic camera in patients who experienced abusive head trauma,” Arch. Pediatr. Adolesc. Med. 155(10), 1149–1152 (2001). [CrossRef]  

18. N. Choudhry, J. S. Duker, K. B. Freund, S. Kiss, G. Querques, R. Rosen, D. Sarraf, E. H. Souied, P. E. Stanga, G. Staurenghi, and S. R. Sadda, “Classification and guidelines for widefield imaging: recommendations from the international widefield imaging study group,” Ophthalmol. Retina 3(10), 843–849 (2019). [CrossRef]  

19. Alastair Atkinson and C. Mazo, “Imaged area of the retina “ https://www.freelists.org/archives/optimal/02-2017/pdf91WmMGLh6Q.pdf (last access on June 9, 2021.).

20. D. Toslak, D. Thapa, Y. Chen, M. K. Erol, R. V. Paul Chan, and X. Yao, “Trans-palpebral illumination: an approach for wide-angle fundus photography without the need for pupil dilation,” Opt. Lett. 41(12), 2688–2691 (2016). [CrossRef]  

21. B. Wang, D. Toslak, M. N. Alam, R. V. P. Chan, and X. Yao, “Contact-free trans-pars-planar illumination enables snapshot fundus camera for nonmydriatic wide field photography,” Sci. Rep. 8(1), 8768 (2018). [CrossRef]  

22. H. L. Liou and N. A. Brennan, “Anatomically accurate, finite model eye for optical modeling,” J. Opt. Soc. Am. A Opt. Image Sci. Vis. 14(8), 1684–1695 (1997). [CrossRef]  

23. Q. Wu, Y. Tang, X. Chen, C. Ma, F. Yao, and L. Liu, “Method for evaluating ophthalmic lens based on eye-lens-object optical system,” Opt Express 27(26), 37274–37285 (2019). [CrossRef]  

24. W. F. Harris, “Nodes and nodal points and lines in eyes and other optical systems,” Ophthalmic. Physiol. Opt. 30(1), 24–42 (2009). [CrossRef]  

25. M. Kaschke, K.-H. Donnerhacke, and M. S. Rill, “Optics of the human eye,” in Optical Devices in Ophthalmology and Optometry (Wiley, 2014), pp. 15–48.

26. P. K. Verkicharla, M. Suheimat, J. M. Pope, F. Sepehrband, A. Mathur, K. L. Schmid, and D. A. Atchison, “Validation of a partial coherence interferometry method for estimating retinal shape,” Biomed. Opt. Express 6(9), 3235–3247 (2015). [CrossRef]  

27. E. Taylor and A. Jennings, “Calculation of total retinal area,” Br. J. Ophthalmol. 55(4), 262–265 (1971). [CrossRef]  

Data availability

All relative data are presented in this article. No other data were generated or analyzed in the presented research.

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (6)

Fig. 1.
Fig. 1. (a) Illustration of visual-angle and eye-angle. (b) Illustration of FOV.
Fig. 2.
Fig. 2. (a) Schematic diagram of the eye model. (b) Ray tracing through the eye model. (c) Illustration of the visual-angle and eye-angle.
Fig. 3.
Fig. 3. Illustration of retinal area estimation as a spherical cap.
Fig. 4.
Fig. 4. (a) The relationship between visual-angle θv and eye-angle θe. (b) Conversion ratio ${\theta _e}/{\theta _v}$. (c) ${\varDelta\theta _e}/{\varDelta\theta _v}$ at different visual field.
Fig. 5.
Fig. 5. (a) Relationship between retinal distance d and visual-angle θv. (b)$\; \varDelta d/{\varDelta\theta _v}$ at different visual field.
Fig. 6.
Fig. 6. (a) Relationship between retinal area and visual-angle FOV. (b) Relationship between retinal area and eye-angle FOV.

Tables (1)

Tables Icon

Table 1. Conversion ratios of visual-angle, eye-angle, retinal dimension corresponding to visual field

Equations (9)

Equations on this page are rendered with MathJax. Learn more.

θe=θv+θr
θr=sin1(NOORsinθv)
θr=sin1(0.51sinθv)
θe=θv+sin1(0.51sinθv)
d=2πrθe/360=πrθe/180
d=πr(θv+sin1(0.51sinθv)/180
A=2πrh
h=r(1cosθe)
A=2πr2(1cosθe)