Abstract
We show that various setups for optical microscopy which are commonly used in biomedical laboratories behave like efficient microscope condensers that are responsible for observed subwavelength resolution. We present a series of experiments and simulations that reveal how inclined illumination from such unexpected condensers occurs when the sample is perpendicularly illuminated by a microscope’s built-in white-light source. In addition, we demonstrate an inexpensive add-on optical module that serves as an efficient and lightweight microscope condenser. Using such add-on optical module in combination with a low-numerical-aperture objective lens and Fourier plane imaging microscopy technique, we demonstrate detection of photonic crystals with a period nearly eight times smaller than the Rayleigh resolution limit.
© 2015 Optical Society of America
Full Article | PDF ArticleMore Like This
Darshan B. Desai and Luis Grave de Peralta
Appl. Opt. 54(12) 3580-3587 (2015)
Sanchari Sen, Luis Molina, Dongyu Cao, Darshan B. Desai, Ayrton A. Bernussi, and Luis Grave de Peralta
Biomed. Opt. Express 6(3) 658-667 (2015)
Sanchari Sen, Ishtiaque Ahmed, Batool Aljubran, Ayrton A. Bernussi, and Luis Grave de Peralta
Appl. Opt. 55(23) 6421-6427 (2016)