Abstract
We present the application to time-domain diffuse optics of a device composed of 8x256 CMOS SPAD array with 256 7- bit time-to-digital converters. Thanks to its structure and despite the limitation on the maximum repetition rate of the laser (2 MHz), it has been demonstrated to be suitable for fast acquisitions (10 ms) provided that a high photon count-rate is used and pile-up distortion is corrected. We demonstrate that high penetration depth (>30 mm) and good linearity in absorption coefficient retrieval can be achieved. Finally, we were able to clearly record the heart beat in a resting state forehead measurement.
© 2023 SPIE
PDF ArticleMore Like This
Vamshi Damagatla, Nadia G. Boetti, Laura Di Sieno, Diego Pugliese, Davide Janner, Alberto Dalla Mora, and Antonio Pifferi
1262818 European Conference on Biomedical Optics (ECBO) 2023
Davide Contini, Alberto Dalla Mora, Simon Arridge, Fabrizio Martelli, Alberto Tosi, Gianluca Boso, Andrea Farina, Turgut Durduran, Edoardo Martinenghi, Alessandro Torricelli, and Antonio Pifferi
95380A European Conference on Biomedical Optics (ECBO) 2015
Laura Di Sieno, Alberto Dalla Mora, Anurag Behera, Alberto Gola, and Fabio Acerbi
ES1B.4 European Conference on Biomedical Optics (ECBO) 2021