Dual optical frequency combs with ultra-low relative phase jitters from 550 nm to 1020 nm for precision spectroscopy

Open Access Open Access

Abstract

Here, ultra-low relative phase jitters over a wide optical spectrum were achieved for dual Ti:Sapphire optical frequency combs. The two optical frequency combs were independently phase-locked to a Sr optical lattice clock laser delivered through a commercial optical fiber network. We confirmed that the relative phase jitters between the two combs integrated from 8.3 mHz to 200 kHz were below 1 rad, corresponding to a relative linewidth of below 8.3 mHz, over the entire wavelength of the optical frequency combs ranging from 550 nm to 1020 nm. Our work paves the way for ultrahigh-precision dual-comb spectroscopy covering a wide optical spectral range with a simple setup, and provides an absolute optical frequency reference with great stability over a wide range of wavelengths.

© 2022 Optica Publishing Group under the terms of the Optica Open Access Publishing Agreement

1. Introduction

Ultra-precision spectroscopy plays a key role in modern science and technology, such as tests of fundamental physics, establishment of an optical frequency standard, and precision timing control for GPS and satellites [19]. A laser with a narrow linewidth and low phase noise is essential for ultra-precision spectroscopy. Preparing an accurate optical reference over a broad wavelength range is necessary to meet the wavelength requirements of deep ultraviolet (UV) to far infrared (IR) applications. The most advanced technique to reduce the phase noise of a laser is to lock the laser to a high-finesse cavity. The state-of-the-art laser has a sub-10 mHz linewidth when locked to a silicon Fabry-Pérot cavity at 124 K [10]. However, high-finesse cavities are available only at a few selected wavelengths. Moreover, the setup for such an ultrastable laser requires a considerable amount of time and resources, including a temperature stabilized high-Q cavity in a shielded environment. Therefore, it would be ideal to distribute the phase coherence of an ultrastable optical reference over various locations and optical frequencies.

Recent progress in optical fiber networks has enabled the distribution of the stability of an ultrastable laser to remote locations with unprecedented precision. In addition, by stabilizing an optical frequency comb (OFC) to the received reference laser, all longitudinal modes of the OFC, over several hundreds of THz, can acquire the same precision as the reference laser. In previous research, we have shown that the stability of the optical reference can be reproduced with no degradation by the Ti:Sapphire OFC under challenging circumstances 1) in the presence of a 2.9 km-long fiber network and 2) without the local phase-purification cavity [11]. A longitudinal mode of the OFC with an optical frequency nearing that of the optical reference was used to examine the phase noise and stability of the OFC in the previous experiment. However, the performance of other longitudinal modes, especially those with wavelengths far from that of the optical reference, remains unclear. Every optical frequency in a stabilized OFC is described as $f_n = f_\mathrm {CEO} + n \times f_\mathrm {rep}$ ($f_\mathrm {CEO}$, $f_\mathrm {rep}$, and $n$ are the carrier-envelope offset (CEO) frequency, the repetition frequency, and the mode number, respectively). In principle, the phase noises of all longitudinal modes of an OFC are tightly related to each other by this equation. In practice, however, fluctuations of various physical origins can degrade the phase coherence of the modes that are away from the reference optical frequency. To validate the application of the OFC in ultra-precision spectroscopy over a wide wavelength range, the phase noise has to be characterized at various wavelengths. In a previous research on Ti:Sapphire OFC, a sub-Hz residual linewidth was confirmed at the wavelength of 250 nm away from the optical reference [12]. People have also demonstrated a fiber OFC with sub-Hz linewidths over the wavelength range of 1200 nm to 1720 nm [13].

Here, we studied the residual phase noise of two OFCs by subjecting them to optical heterodyne detection, within the 550 – 1020 nm wavelength range. Two identical Ti:Sapphire OFCs were prepared and stabilized independently to a remotely located Sr optical lattice clock laser through an optical fiber network, as introduced in a previous study [11]. Heterodyne beatnotes were detected at various wavelengths after overlapping the two pulse trains precisely in the time domain. The phase noise on the detected heterodyne beatnotes contains only the residual phase noise of the OFCs because the common noise of the optical reference and the fiber network cancel out. The residual linewidth of the OFCs over a wide range of wavelengths was analyzed based on the detected phase noise.

2. Experimental setup

The overall experimental setup consists of two Ti:Sapphire OFCs that are phase-locked to a remotely placed ultrastable laser for Sr optical lattice clock (optical standard) as described in Ref. [11]. To measure the heterodyne beat signal of the two OFCs, the timing of optical pulses from the two OFCs are precisely synchronized. The residual linewidth of the OFC with respect to the optical standard is evaluated by analyzing the phase noise of the heterodyne beat signal.

2.1 Stabilization of the OFCs

The setup of this experiment comprises two areas (Fig. 1). We constructed two identical home-made Ti:Sapphire OFCs with output powers and repetition frequencies ($f_\textrm {rep}$) of approximately 500 mW and 117 MHz, respectively. After broadening the wavelength range using photonic crystal fibers, the $f_{\mathrm {CEO}}$s of the OFCs were stabilized by a self-referencing technique with an $f$$2f$ interferometer. The optical standard, an ultranarrow-linewidth laser at 698 nm for a Sr optical lattice clock [14], was located at a laboratory (remote lab) about 1 km away from the location of the two OFCs (main lab). The optical standard was delivered to the main lab through a 2.9 km telecom-wavelength optical fiber network (UT-NET, University of Tokyo), phase noise of which was actively canceled [11]. One of the frequency modes of the OFCs was phase-locked to the delivered optical standard with a frequency offset ($f_\mathrm {off}$). All RF signals and measurement apparatus were referenced by a GPS-disciplined Rb standard. A detailed description of the OFC and its performance after the stabilization are given in Appendix 5.1 and Ref. [11].

 figure: Fig. 1.

Fig. 1. Overall experimental setup. The optical standard at the remote lab is delivered to the main lab through a telecom-wavelength fiber network (UT-NET) with fiber noise cancellation. Two identical OFCs, with self-referenced $f_{\mathrm {CEO}}$s, are phase-locked to the optical standard. The beat signal of the two OFCs is measured at various optical wavelengths to evaluate the residual phase noise of the OFCs. DM: dichroic mirror, BS: beam splitter, G: grating, PD: photo detector, PCF: photonic crystal fiber.

Download Full Size | PDF

In the previous experiment, the out-of-loop measurement at the wavelength near the optical standard detected the phase noise of 0.54 rad integrated from 8.3 mHz to 1.95 MHz with respect to the optical standard. However, this phase noise included high-frequency phase noise from the repeater laser of the optical standard. The high-frequency phase noise cannot be transfered to the OFC due to the intrinsic narrow linewidth of the OFC. To correctly evaluate the phase noise of the OFC itself, we constructed two identical OFCs and evaluated the beat signal at various optical wavelengths.

2.2 Heterodyne beatnote measurement of the two OFCs

The optical paths of the two OFCs were carefully overlapped using a beam splitter so that the two light waves would interfere with each other. The full optical spectrum of the OFCs, 550 nm to 1100 nm, was used in the experiment. To reduce the phase noise due to different optical paths, we minimized the lengths of the out-of-loop path of the two OFCs. Once the paths were combined, a grating was employed to separate the wavelengths of the OFCs. The phase noise at selected wavelength ranges was measured by placing a slit between the grating and a photo detector, as shown in Fig. 1. By rotating the grating, we changed the wavelength range to $549$ nm, $650$ nm, $750$ nm, $850$ nm, $948$ nm, and $1020$ nm. The post-slit wavelength was measured using a spectrometer (Thorlabs, CCS200/M).

The pulse width of the spectrally resolved OFCs was approximately 100 fs. For the two OFCs to interfere, a prerequisite is to synchronize the arrival times of the two OFC pulses within the pulse width. In other words, the two OFCs should have the same $f_\mathrm {rep}$, and their repetition phases should match. The frequency of the heterodyne beat ($f_\mathrm {beat}$) then appears at $f_\mathrm {CEO,1} - f_\mathrm {CEO,2}$. To synchronize the timing of the pulses, we first chose the $f_\mathrm {CEO}$s and $f_\mathrm {off}$s, and match the $f_\mathrm {rep}$s of the two OFCs by properly chose the longitudinal mode locked to the optical reference. We set $f_\mathrm {CEO,1} = 10.3$ MHz, $f_\mathrm {off,1} = -11.1$ MHz, $f_\mathrm {CEO,2} = -11.1$ MHz, and $f_\mathrm {off,2} = 10.3$ MHz, which resulted in $f_\mathrm {rep}=117$ MHz and $f_\mathrm {beat}=21.4$ MHz.

Matching the repetition phases is more challenging because the path-length of the two OFCs should be matched within a 30 $\mu$m - spatial spread of the pulses. To achieve this, we intentionally changed the $f_\mathrm {rep}$ of one OFC by a small amount ($< 10$ mHz), thereby changing the relative position between the two femtosecond pulses as a function of time. It was done by changing $f_\mathrm {beat}$ between one of the OFCs and the optical standard by a few MHz. This allowed us to change the $f_\mathrm {rep}$ in a controlled way and the OFC was always locked throughout the process. A heterodyne beat at 21.4 MHz appears when the two pulses overlap. At this moment, we immediately set $f_\mathrm {rep}$ of the OFC back to its original value.

3. Results and discussions

The heterodyne beat signal at 650 nm, measured by a spectrum analyzer (Anritsu, MS2720T), is shown in Fig. 2 (red lines). The signal was amplified by 50 dB using a home-made amplifier. The center frequency of the beat was 21.4 MHz as expected. We observed a central coherent peak, the width of which is limited by the instrumental resolution. The phase noise at low frequencies matched very well to the in-loop phase noise of $f_\mathrm {off}$. However, at high frequencies, the heterodyne beat between the two OFCs had a significantly smaller phase noise. This indicates that the high frequency noise of $f_\mathrm {off}$ is from the repeater laser of the optical standard and is rejected during the active stabilization process of the OFCs.

 figure: Fig. 2.

Fig. 2. Heterodyne beat signal between the two OFCs at 650 nm. For reference, the in-loop spectra of $f_\mathrm {CEO}$ and $f_\mathrm {off}$ are also shown. (a) Resolution bandwidth of 300 Hz, (b) resolution bandwidth of 1 Hz.

Download Full Size | PDF

The phase noise of the heterodyne beat was also measured after subtracting the carrier frequency to further examine the low frequency components of the phase noise. We compared the phase of the heterodyne beat to a reference RF signal generator using a home-made digital phase-frequency discriminator (DPFD). The output of the DPFD was measured using an oscilloscope (Pico Technology, PicoScope5244B) for 120 s with a sampling rate of 3.9 MHz. The two OFCs were phase-locked during the 120 s measurement without any phase jumps. However, the output power fluctuation of the OFCs caused the phase signal of the resulting beat to be discontinuous, at wavelength of 550 nm, 850 nm, 950 nm, and 1020 nm. The discontinuities were especially significant for 550 nm and 1020 nm, corresponding to the edges of the OFC optical spectra. This occurred when the input signal voltage was insufficient relative to the threshold voltage of the DPFD. In order to concentrate only on the phase noise, we selected time windows where there were no discontinuous phase signals for each wavelength.

The power spectral density (PSD) of the phase noise obtained from the raw data had a spectral range of 8.3 mHz to 1.95 MHz. The 8.3 mHz resolution was set from the 120 s measurement time of the phase noise. We confirmed that there were negligible amount of phase noise at frequencies higher than 20 kHz due to the intrinsic narrow linewidth of our OFCs. Therefore, the PSD components upto 20 kHz were used for the following analysis.

The measured phase noise had a linear drift on the order of 0.003 rad/s, with magnitude and sign varying over time. The drift is due to the different optical paths between the OFCs. The drifts, as they are not essential for our discussion, are excluded in the following analysis. Detailed discussion about the linear drift is given in Appendix 5.2.

The PSD of the phase noise of the beat between the two OFCs is shown in Fig. 3(a). The spectral shapes of the PSDs at this frequency regime followed the spectral shapes of the locked beatnote between the OFC and the optical standard. The peak at 10 kHz is one of the feedback resonance of the phase-locked loop. All PSDs display identical shapes over the entire wavelength range of over an octave of the OFCs, indicating that no additional phase noise was added to the OFCs. Below 1 Hz, the PSDs show the typical $1/f^2$ dependence of white frequency modulation noise. Figure 3(b) shows the accumulated phase noise of each OFC calculated from the PSDs. The accumulated phase is smaller than 1 rad over the whole wavelength range. Therefore, the residual linewidth of the OFC is smaller than the resolution of our measurement, namely 8.3 mHz over the whole wavelength range. The relative linewidth demonstrated here is two orders of magnitudes narrower than the previous works [12,13].

 figure: Fig. 3.

Fig. 3. (a) PSD and (b) integrated phase jitters of the phase noise of the heterodyne beat between the two OFCs at each wavelength.

Download Full Size | PDF

When the $f_\textrm {CEO}$ and $f_\textrm {rep}$ of an OFC are stabilized, the phase accumulation should increase linearly as a function of the optical frequency of the OFC [13]. As shown in Fig. 4, the accumulated phase jitters at 1 Hz follow a linear function over most of the optical spectrum of the OFC. The accumulated phase jitter at 298 THz ($= 1020$ nm) is more pronounced than expected because of the optical power fluctuation. The timing jitter of the OFC pulses is estimated to be approximated 150 as from the accumulated phase jitter.

 figure: Fig. 4.

Fig. 4. Accumulated phase jitter at 1 Hz for various wavelengths. The black dashed line indicates the linear guideline. The y intercept of this line is zero. The frequency of the optical standard is represented by the gray vertical dashed line.

Download Full Size | PDF

The two tightly phase-locked OFCs with absolute frequency accuracy are excellent tools for the precision dual-comb spectroscopy [15]. The optical bandwidth and the frequency accuracy are the two features that describe the performance of the dual comb spectroscopy. The optical bandwidth of the dual-comb spectroscopy is determined by the ratio between the relative linewidth of the two OFCs and $f_\textrm {rep}$. Therefore, it is important that the OFCs maintain the narrow relative linewidth over a wide optical wavelength range to achieve a large optical bandwidth. The frequency resolution and accuracy are assured when the two OFCs have absolute frequency modes precisely described by $f_n = f_\mathrm {CEO} + n \times f_\mathrm {rep}$ with narrow linewidths over the whole spectral range. Our result with < 1 rad phase jitters over 120 s for an octave-spanning wavelength range would allow dual comb spectroscopy with an unprecedented precision.

Figure 5 presents an example of a dual-comb spectroscopy signal using our OFCs. We selected wavelengths in the 655 to 665 nm range (optical frequency width of 7.2 THz, approximately 60,000 frequency modes) using a band-pass filter for this test. We set the $\Delta f_\textrm {rep} = 26.45$ Hz in this measurement. Figure 5(a) is the raw data taken by a photo detector with 150 MHz bandwidth and a 16-bit digitizer with a 100 MHz sampling frequency. The interference signals of the two OFCs appear every $1/26.45 \sim 37.8$ ms. By zooming into Fig. 5(b), the beat inside the pulse can be observed. The RF spectrum of the signal is obtained by performing a fast Fourier transform (FFT) of the data after filtering out the noise during periods of no interference (Fig. 5(c)). To conserve the periodicity of the pulse arrays during the FFT, we cut out the time length of our data to $1/(26.45 \textrm { Hz}) \times 26 \sim 983$ ms. The width of the RF spectrum is approximately 1 MHz, which is comparable to the expected value of $26.45 \times 60,000 \sim 1.6$ MHz. Figure 5(d) shows the zoomed in spectrum where one can clearly resolve each mode of the RF comb separated by 26.45 Hz with the linewidth of 1 Hz limited by our measurement time of 1 s. This confirms that our OFCs maintain their coherence over 1 s. The signal-to-noise(S/N) can be estimated from the spectral density. The expected maximum S/N per 1-second-long measurement was deduced to be 14 from the peak value and the RMS of the noise in Fig. 5(d).

 figure: Fig. 5.

Fig. 5. Demonstration of the dual-comb spectroscopy using our OFCs. Owing to the long coherence time of the two OFCs, RF spectra with a 1 Hz linewidth could be obtained. (a)(b): The raw interference data. (c)(d) The RF spectrum obtained by Fourier transform. Every RF mode is separated by 26.45 Hz and the linedwidth of 1 Hz is set by the measurement time.

Download Full Size | PDF

4. Conclusion

We constructed two Ti:Sapphire OFCs that were tightly phase-locked to each other over a wide wavelength range, 550 nm to 1020 nm. The two OFCs were independently phase-locked to an optical lattice clock at 698 nm via an optical fiber network. The phase noises at different wavelengths were measured by taking heterodyne beatnotes of the two OFCs. The resulting phase jitters displayed a linear increase as a function of the optical frequencies, which is in agreement with the theoretical prediction. Phase jitters of less than 1 rad integrated from 8.3 mHz to 20 kHz indicate that the two OFCs are tightly phase-locked to each other over the entire wavelength range of 550 nm to 1020 nm. The phase jitter measured here corresponds to a relative linewidth of < 8.3 mHz and a timing jitter of 150 as. Our result shows that these OFCs transfer the coherence of a remote optical clock laser with minimal phase noise to arbitraty wavelengths in the range stated above. Therefore, the referenced OFCs with the current setup can be utilized for state-of-the-art precision spectroscopy over a wide wavelength range. Also, the ultrastable OFC demonstrated here is an excellent tool for dual-comb spectroscopy covering a wide optical bandwidth with a Hz-level resolution and a pump-probe measurement with sub-fs timing resolution. It also provides an absolute optical frequency reference with great stability over a wide range of wavelengths.

5. Appendices

5.1. Properties of the OFCs

Home-made Ti:Sapphire OFCs were used in this experiment. The repetition rate was 117 MHz and the pulse duration was about 40 fs. The spectra of the OFCs spanned from 750 nm to 850 nm (Fig. 6(a)) and had an output power of 500 mW. Commercial PCFs (FemtoWHITE 800, NKT Photonics) were used to broaden the spectral range to 500 nm to 1100 nm (Fig. 6(b)).

 figure: Fig. 6.

Fig. 6. (a) The spectrum of the OFC before the PCF. The spectrum spanned from 750 nm to 850 nm and the center wavelength was about 770 nm. (b) The spectrum of the OFC after the PCF. The spectrum was broadened to 500 nm to over 1100 nm range.

Download Full Size | PDF

5.2. Linear drift of the phase noise

In the experiment, we observed a linear drift of the phase noise between the two OFCs as shown in Fig. 7(a). This linear drift came from the thermal fluctuation of the long distance between the OFC and the stabilization point where we took the beatnote between the OFC and the optical standard. The distances were 2 m for one OFC and 1 m for the other. Even though we controlled the temperature of the laboratory and covered our optical table to minimize temperature changes, it is reasonable to expect about 0.1 $^{\cdot }$C change in an hour in winter when the experiment was conducted. With this temperature fluctuation, the path length would change about 0.3 nm/s, which results in 0.003 rad/s linear drift of the phase. We observed linear drifts about 0.001 rad/s to 0.003 rad/s, both positive and negative, depending on measurements which agree to the properties of the temperature fluctuation. This temperature fluctuation was not an inherent property of the OFC and it hindered proper analysis of the phase noise of the OFC (Fig. 7(b)). Therefore, we excluded the linear drift in our analysis.

 figure: Fig. 7.

Fig. 7. (a) The linear drift of the phase noise between the two OFCs at 650 nm. The noise had a positive linear drift in this case. This linear drift was caused by the path length change due to the temperature fluctuation in the laboratory. (b) The effect of the linear drift on the PSD analysis. The PSD showed a strong $1/f^2$ dependence at low frequency range when we included the linear drift of the phase noise.

Download Full Size | PDF

5.3. Relative intensity noise of the OFC

Figure 8 is the expected relative intensity noise (RIN) of the OFC at 620 nm for 1-second-long measurement. The RIN was about −87 dB/Hz for most of the frequency regime. Some spikes were from the circuit noise of the home-made photodetector. This RIN was induced by the spectrum broadening of the PCF.

 figure: Fig. 8.

Fig. 8. RIN of the OFC at 620 nm. It was about −87 dB/Hz over the most of the frequency regime.

Download Full Size | PDF

Funding

Japan Society for the Promotion of Science (JP17H06205, JP20K05357); Ministry of Education, Culture, Sports, Science and Technology (JPMXS0118067246); Precursory Research for Embryonic Science and Technology (JPMJPR190B); National Research Foundation of Korea (2020R1A4A1018015, 2021R1C1C1009450); Korea University (K2009521).

Acknowledgments

The authors thank H. Katori for providing access to the clock laser and other equipment in his laboratory. The authors appreciate the technical support provided by T. Takano, N. Ohmae, T. Akatsuka and I. Ushijima. The authors thank M. Kuwata-Gonokami for presenting the conceptual idea in Advanced Photon Science Alliance (APSA) project that led to the present work, and providing basic equipment at the OFC site.

Disclosures

The authors declare no conflicts of interest.

Data availability

Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.

References

1. R. Altmann, L. Dreissen, E. Salumbides, W. Ubachs, and K. Eikema, “Deep-ultraviolet frequency metrology of H2 for tests of molecular quantum theory,” Phys. Rev. Lett. 120(4), 043204 (2018). [CrossRef]  

2. J. Biesheuvel, J.-P. Karr, L. Hilico, K. S. E. Eikema, W. Ubachs, and J. C. J. Koelemeij, “Probing QED and fundamental constants through laser spectroscopy of vibrational transitions in HD+,” Nat. Commun. 7(1), 10385 (2016). [CrossRef]  

3. J. K. Webb, J. A. King, M. T. Murphy, V. V. Flambaum, R. F. Carswell, and M. B. Bainbridge, “Indications of a spatial variation of the fine structure constant,” Phys. Rev. Lett. 107(19), 191101 (2011). [CrossRef]  

4. N. Huntemann, B. Lipphardt, C. Tamm, V. Gerginov, S. Weyers, and E. Peik, “Improved limit on a temporal variation of mp/me from comparisons of Yb+ and Cs atomic clocks,” Phys. Rev. Lett. 113(21), 210802 (2014). [CrossRef]  

5. R. Godun, P. Nisbet-Jones, J. Jones, S. King, L. Johnson, H. Margolis, K. Szymaniec, S. Lea, K. Bongs, and P. Gill, “Frequency ratio of two optical clock transitions in 171Yb+ and constraints on the time variation of fundamental constants,” Phys. Rev. Lett. 113(21), 210801 (2014). [CrossRef]  

6. C. W. Chou, D. B. Hume, T. Rosenband, and D. J. Wineland, “Optical clocks and relativity,” Science 329(5999), 1630–1633 (2010). [CrossRef]  

7. C. Lisdat, G. Grosche, N. Quintin, C. Shi, S. Raupach, C. Grebing, D. Nicolodi, F. Stefani, A. Al-Masoudi, S. Dörscher, S. Häfner, J.-L. Robyr, N. Chiodo, S. Bilicki, E. Bookjans, A. Koczwara, S. Koke, A. Kuhl, F. Wiotte, F. Meynadier, E. Camisard, M. Abgrall, M. Lours, T. Legero, H. Schnatz, U. Sterr, H. Denker, C. Chardonnet, Y. Le Coq, G. Santarelli, A. Amy-Klein, R. Le Targat, J. Lodewyck, O. Lopez, and P.-E. Pottie, “A clock network for geodesy and fundamental science,” Nat. Commun. 7(1), 12443 (2016). [CrossRef]  

8. A. Derevianko, “Atomic clocks and dark-matter signatures,” J. Phys.: Conf. Ser. 723, 012043 (2016). [CrossRef]  

9. B. M. Roberts, G. Blewitt, C. Dailey, M. Murphy, M. Pospelov, A. Rollings, J. Sherman, W. Williams, and A. Derevianko, “Search for domain wall dark matter with atomic clocks on board global positioning system satellites,” Nat. Commun. 8(1), 1195 (2017). [CrossRef]  

10. D. G. Matei, T. Legero, S. Häfner, C. Grebing, R. Weyrich, W. Zhang, L. Sonderhouse, J. M. Robinson, J. Ye, F. Riehle, and U. Sterr, “1.5 μm lasers with sub-10 mHz linewidth,” Phys. Rev. Lett. 118(26), 263202 (2017). [CrossRef]  

11. E. Chae, K. Sugiyama, K. Yoshioka, K. Nakashima, and T. Ikeda, “Direct phase-locking of a Ti:Sapphire optical frequency comb to a remote optical frequency standard,” Opt. Express 27(11), 15649–15661 (2019). [CrossRef]  

12. A. Bartels, C. W. Oates, L. Hollberg, and S. A. Diddams, “Stabilization of femtosecond laser frequency combs with subhertz residual linewidths,” Opt. Lett. 29(10), 1081–1083 (2004). [CrossRef]  

13. W. C. Swann, J. J. McFerran, I. Coddington, N. R. Newbury, I. Hartl, M. E. Fermann, P. S. Westbrook, J. W. Nicholson, K. S. Feder, C. Langrock, and M. M. Fejer, “Fiber-laser frequency combs with subhertz relative linewidths,” Opt. Lett. 31(20), 3046–3048 (2006). [CrossRef]  

14. T. Takano, M. Takamoto, I. Ushijima, N. Ohmae, T. Akatsuka, A. Yamaguchi, Y. Kuroishi, H. Munekane, B. Miyahara, and H. Katori, “Geopotential measurements with synchronously linked optical lattice clocks,” Nat. Photonics 10(10), 662–666 (2016). [CrossRef]  

15. I. Coddington, N. Newbury, and W. Swann, “Dual-comb spectroscopy,” Optica 3(4), 414–426 (2016). [CrossRef]  

Data availability

Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (8)

Fig. 1.
Fig. 1. Overall experimental setup. The optical standard at the remote lab is delivered to the main lab through a telecom-wavelength fiber network (UT-NET) with fiber noise cancellation. Two identical OFCs, with self-referenced $f_{\mathrm {CEO}}$s, are phase-locked to the optical standard. The beat signal of the two OFCs is measured at various optical wavelengths to evaluate the residual phase noise of the OFCs. DM: dichroic mirror, BS: beam splitter, G: grating, PD: photo detector, PCF: photonic crystal fiber.
Fig. 2.
Fig. 2. Heterodyne beat signal between the two OFCs at 650 nm. For reference, the in-loop spectra of $f_\mathrm {CEO}$ and $f_\mathrm {off}$ are also shown. (a) Resolution bandwidth of 300 Hz, (b) resolution bandwidth of 1 Hz.
Fig. 3.
Fig. 3. (a) PSD and (b) integrated phase jitters of the phase noise of the heterodyne beat between the two OFCs at each wavelength.
Fig. 4.
Fig. 4. Accumulated phase jitter at 1 Hz for various wavelengths. The black dashed line indicates the linear guideline. The y intercept of this line is zero. The frequency of the optical standard is represented by the gray vertical dashed line.
Fig. 5.
Fig. 5. Demonstration of the dual-comb spectroscopy using our OFCs. Owing to the long coherence time of the two OFCs, RF spectra with a 1 Hz linewidth could be obtained. (a)(b): The raw interference data. (c)(d) The RF spectrum obtained by Fourier transform. Every RF mode is separated by 26.45 Hz and the linedwidth of 1 Hz is set by the measurement time.
Fig. 6.
Fig. 6. (a) The spectrum of the OFC before the PCF. The spectrum spanned from 750 nm to 850 nm and the center wavelength was about 770 nm. (b) The spectrum of the OFC after the PCF. The spectrum was broadened to 500 nm to over 1100 nm range.
Fig. 7.
Fig. 7. (a) The linear drift of the phase noise between the two OFCs at 650 nm. The noise had a positive linear drift in this case. This linear drift was caused by the path length change due to the temperature fluctuation in the laboratory. (b) The effect of the linear drift on the PSD analysis. The PSD showed a strong $1/f^2$ dependence at low frequency range when we included the linear drift of the phase noise.
Fig. 8.
Fig. 8. RIN of the OFC at 620 nm. It was about −87 dB/Hz over the most of the frequency regime.