Abstract
Accurate determination of scan positions is essential for achieving high-quality reconstructions in ptychographic imaging. This study presents and demonstrates a method for determining the rotation angle of the scan pattern relative to the detector pixel array using diffraction data. The method is based on the Fourier–Mellin transform and cross-correlation calculation. It can correct rotation errors up to 60 deg. High-quality reconstructions were obtained for visible light and electron microscopy datasets, and intricate structures of samples can be revealed. We believe that this refinement method for rotary position errors can be valuable for improving the performance of ptychographic four-dimensional scanning transmission electron microscopy.
© 2024 Optica Publishing Group
Full Article | PDF ArticleMore Like This
Jiantai Dou, Zihao Pei, Youyou Hu, and Zhongming Yang
Opt. Express 31(25) 42464-42477 (2023)
Lu Rong, Chao Tang, Dayong Wang, Bing Li, Fangrui Tan, Yunxin Wang, and Xiaoyu Shi
Opt. Express 27(2) 938-950 (2019)
Fucai Zhang, Isaac Peterson, Joan Vila-Comamala, Ana Diaz, Felisa Berenguer, Richard Bean, Bo Chen, Andreas Menzel, Ian K. Robinson, and John M. Rodenburg
Opt. Express 21(11) 13592-13606 (2013)