Abstract
Precise mode matching is needed to maximize performance in coupled cavity interferometers such as Advanced Laser Interferometer Gravitational-Wave Observatory (LIGO). In this paper, we present a new mode matching sensing scheme, to the best of our knowledge, that uses a single radio-frequency higher-order-mode sideband and single-element photodiodes. It is first-order insensitive to misalignment and can serve as an error signal in a closed loop control system for a set of mode matching actuators. We also discuss how it may be implemented in Advanced LIGO. The proposed mode matching error signal has been successfully demonstrated on a tabletop experiment, where the error signal increased the mode matching of a beam to a cavity to 99.9%.
© 2020 Optical Society of America
Full Article | PDF ArticleMore Like This
P. Fulda, D. Voss, C. Mueller, L. F. Ortega, G. Ciani, G. Mueller, and D. B. Tanner
Appl. Opt. 56(13) 3879-3888 (2017)
Aidan F. Brooks, Benjamin Abbott, Muzammil A. Arain, Giacomo Ciani, Ayodele Cole, Greg Grabeel, Eric Gustafson, Chris Guido, Matthew Heintze, Alastair Heptonstall, Mindy Jacobson, Won Kim, Eleanor King, Alexander Lynch, Stephen O’Connor, David Ottaway, Ken Mailand, Guido Mueller, Jesper Munch, Virginio Sannibale, Zhenhua Shao, Michael Smith, Peter Veitch, Thomas Vo, Cheryl Vorvick, and Phil Willems
Appl. Opt. 55(29) 8256-8265 (2016)
S. Dwyer, L. Barsotti, S. S. Y. Chua, M. Evans, M. Factourovich, D. Gustafson, T. Isogai, K. Kawabe, A. Khalaidovski, P. K. Lam, M. Landry, N. Mavalvala, D. E. McClelland, G. D. Meadors, C. M. Mow-Lowry, R. Schnabel, R. M. S. Schofield, N. Smith-Lefebvre, M. Stefszky, C. Vorvick, and D. Sigg