Abstract
The optical parameters of a thin-film coating determined from the spectral reflectance and transmittance measurements at various incidence angles, including the normal incidence and the Brewster's angle, are compared in this paper. The high-accuracy measurements were carried out through visible–near-infrared spectral regions by using our purpose-built instruments. The optical parameters obtained from the reflectance and the transmittance data are consistent over the angles of incidence and agree within 0.2%. The effect of important systematic factors in the oblique-incidence spectrophotometric measurements is also discussed.
© 2006 Optical Society of America