Abstract
Optode geometry plays an important role in achieving both good spatial resolution and spatial uniformity of detection in diffuse-optical-imaging-based brain activation studies. The quality of reconstructed images for six optode geometries were studied and compared using a laboratory tissue phantom model that contained an embedded object at two separate locations. The number of overlapping measurements per pixel (i.e., the measurement density) and their spatial distributions were quantified for all six geometries and were correlated with the quality of the resulting reconstructed images. The latter were expressed by the area ratio (AR) and contrast-to-noise ratio (CNR) between reconstructed and actual objects. Our results revealed clearly that AR and CNR depended on the measurement density asymptotically, having an optimal point for measurement density beyond which more overlapping measurements would not significantly improve the quality of reconstructed images. Optimization of probe geometry based on our method demonstrated that a practical compromise can be attained between DOI spatial resolution and measurement density.
© 2009 Optical Society of America
Full Article | PDF ArticleMore Like This
Hamid Dehghani, Brian R. White, Benjamin W. Zeff, Andrew Tizzard, and Joseph P. Culver
Appl. Opt. 48(10) D137-D143 (2009)
David A. Boas and Anders M. Dale
Appl. Opt. 44(10) 1957-1968 (2005)
Danny K. Joseph, Theodore J. Huppert, Maria Angela Franceschini, and David A. Boas
Appl. Opt. 45(31) 8142-8151 (2006)