Abstract
Fluorescence has been widely used in biological research and clinical diagnosis. One challenge facing the rapid growth of fluorescence application is the inability to make comparable fluorescence intensity measurements during a long period of clinical study and across laboratories. We propose a method to implement system-independent fluorescence intensity calibration in fiber-optic fluorescence spectrometer systems. This method is based on a National Institute for Standards and Technology traceable standard light source for system spectral response calibration, and a fluorescence reference standard for fluorescence intensity calibration. Human skin in vivo and a fluorescence phantom made of fluorescent microspheres were measured with two different system configurations and at different probe-sample distances. Experimental results showed very good agreement with theory after system-independent fluorescence intensity calibration.
© 2007 Optical Society of America