Expand this Topic clickable element to expand a topic
Skip to content
Optica Publishing Group

Modulation transfer function system for testing telescopic and telemicroscopic low-vision aids

Open Access Open Access

Abstract

As part of a study of the relationship between the optical quality of low-vision aids and the quality of vision provided by these devices we wanted to measure their modulation transfer functions (MTF). We present our apparatus for measuring the MTF of low-power telescopes and telemicroscopes (instruments that are nominally afocal in image space). We can test at spatial frequencies up to 100 cycles/deg in object space. MTF tests may be made across the field of vision, through focus, at 0°, 45°, 90°, and 135° scan azimuths, with white light or passband filters, and with artificial pupils. A computerized data acquisition and analysis system corrects digital data for noise, and computes and plots the MTF curves. MTFs of some commonly prescribed telescopic aids are shown and compared.

© 1985 Optical Society of America

PDF Article
More Like This
Relating low-vision CSFs to functional classifications of vision loss

Gary S. Rubin
FI7 OSA Annual Meeting (FIO) 1985

Design, fabrication, and testing of a 31-in. coudé telescope for atmospheric characterization

JAMES A. DOWUNG, C. VERNE MUFFOLETTO, and RICHARD F. HORTON
WM3 Conference on Lasers and Electro-Optics (CLEO:S&I) 1985