Abstract
Temperature uncertainty improvements of surface phosphor thermometry imaging are reported, with the intensity-ratio method. Analysis of optical filters/post-processing spatial filters with Mg4FGeO6:Mn4+ thermographic phosphors results in a ≈1% temperature uncertainty between 300 and 825 K.
© 2022 The Author(s)
PDF Article | Presentation VideoMore Like This
T. Kissel, J. Brübach, and A. Dreizler
LWB5 Laser Applications to Chemical, Security and Environmental Analysis (LACSEA) 2008
J. Brübach, A. Patt, and A. Dreizler
TuE6 Laser Applications to Chemical, Security and Environmental Analysis (LACSEA) 2006
Kentaro Hayashi, Hiroshi Isoya, Toshihisa Ueda, Takeshi Yokomori, and Taro Hirasawa
FW3A.27 Frontiers in Optics (FiO) 2012