Abstract
Bulky, expensive Nd:YAG lasers are used in conventional photoacoustic tomography (PAT) systems, making them difficult to translate into clinics. Moreover, real-time imaging is not feasible when a single-element ultrasound transducer is used with these low-pulse-repetition-rate lasers (10–100 Hz). Low-cost pulsed laser diodes (PLDs) can be used instead for photoacoustic imaging due to their high-pulse-repetition rates and compact size. Together with acoustic-reflector-based multiple single-element ultrasound transducers, a portable desktop PAT system was developed. This second-generation PLD-based PAT achieved 0.5 s cross-sectional imaging time with high spatial resolution of and an imaging depth of 3 cm. The performance of this system was characterized using phantom and in vivo studies. Dynamic in vivo imaging was also demonstrated by monitoring the fast uptake and clearance of indocyanine green in small animal (rat) brain vasculature.
© 2018 Optical Society of America
Full Article | PDF ArticleMore Like This
Paul Kumar Upputuri and Manojit Pramanik
Biomed. Opt. Express 6(10) 4118-4129 (2015)
Praveenbalaji Rajendran and Manojit Pramanik
Opt. Lett. 46(18) 4510-4513 (2021)
Xianjin Dai, Hao Yang, and Huabei Jiang
Opt. Lett. 42(7) 1456-1459 (2017)