Abstract
Ultrasound detection is performed by measuring laser reflection off a surface plasmon resonance cavity which is integrated at a single-mode fiber end facet. It shows a total noise (or laser RIN) equivalent pressure of 9.7 KPa (or 5.2 KPa) over 0–20 MHz, a 6-dB angular detection range as large as 70° near 10 MHz, a detection bandwidth larger than 125 MHz, and a stable performance for over 20 min without feedback in ambient conditions. Its small form factor, fiber-optic integration, almost omnidirectional responsivity and large bandwidth are favorable for in vivo applications and high resolution imaging.
© 2018 Optical Society of America
Full Article | PDF ArticleMore Like This
Yujie Chen, Yan Wang, Tunan Lv, Junning Zhang, and Hongbin Yu
Opt. Lett. 48(4) 868-871 (2023)
Liuyang Yang, Chenhao Dai, Anqi Wang, Geng Chen, Dongchen Xu, Yanpeng Li, Zhijun Yan, and Qizhen Sun
Opt. Lett. 47(15) 3700-3703 (2022)
Álvaro González-Vila, Andreas Ioannou, Médéric Loyez, Marc Debliquy, Driss Lahem, and Christophe Caucheteur
Opt. Lett. 43(10) 2308-2311 (2018)