Abstract
We report the design and characterization of a 6 mm outer diameter pull-back circumferential scanning visible optical coherence tomography probe. The probe’s large visible bandwidth (500–695 nm) allowed for inverse spectroscopic analysis and an axial resolution of in tissue. We verify spectral imaging capabilities by measuring microsphere backscattering spectra and demonstrate in vivo spatial nanoscale characterization of tissue.
© 2018 Optical Society of America
Full Article | PDF ArticleMore Like This