Abstract
One of the major constraints facing laser speckle imaging for blood-flow measurement is reliable measurement of the correlation time () of the back-scattered light and, hence, the blood’s speed in blood vessels. In this Letter, we present a new model expression for integrated speckle contrast, which accounts not only for temporal integration but spatial integration, too, due to the finite size of the pixel of the CCD camera; as a result, we find that a correction factor should be introduced to the measured speckle contrast to properly determine ; otherwise, the measured blood’s speed is overestimated. Experimental results support our theoretical model.
© 2014 Optical Society of America
Full Article | PDF ArticleMore Like This
Caitlin Regan, Julio C. Ramirez-San-Juan, and Bernard Choi
Opt. Lett. 39(17) 5006-5009 (2014)
Peng Miao, Zhen Chao, Yiguang Zhang, Nan Li, and Nitish V. Thakor
Opt. Lett. 39(13) 3907-3910 (2014)
P. Zakharov, A.C. Völker, M.T. Wyss, F. Haiss, N. Calcinaghi, C. Zunzunegui, A. Buck, F. Scheffold, and B. Weber