Abstract
We developed handheld photoacoustic microscopy (PAM) to detect melanoma and determine tumor depth in nude mice in vivo. Compared to our previous PAM system for melanoma imaging, a new light delivery mechanism is introduced to improve light penetration. We show that melanomas with 4.1 and 3.7 mm thicknesses can be successfully detected in phantom and in in vivo experiments, respectively. With its deep melanoma imaging ability and handheld design, this system can be tested for clinical melanoma diagnosis, prognosis, and surgical planning for patients at the bedside.
© 2014 Optical Society of America
Full Article | PDF ArticleMore Like This
Congcong Wang, Lili Guo, Guan Wang, Tong Ye, Bo Wang, Jiaying Xiao, and Xinli Liu
Appl. Opt. 60(13) 3772-3778 (2021)
Yating Wang, Dong Xu, Sihua Yang, and Da Xing
Biomed. Opt. Express 7(2) 279-286 (2016)
Jenny Hult, Aboma Merdasa, Agnes Pekar-Lukacs, Magne Tordengren Stridh, Azin Khodaverdi, John Albinsson, Bodil Gesslein, Ulf Dahlstrand, Linn Engqvist, Yousef Hamid, Douglas Larsson Albèr, Bertil Persson, Tobias Erlöv, Rafi Sheikh, Magnus Cinthio, and Malin Malmsjö