The relative spectral responses of erythemally weighted broadband radiometers determined at three different laboratories are compared, and the systems are described. The results of measurements of four different broadband radiometers are discussed. Although the common dynamic range of the measured relative spectral responses is approximately , the differences in the relative spectral response functions are lower than 20%. These differences are related mostly to measurement uncertainties and differences in the spectral response facilities.