Abstract

We describe the application of wide-field frequency domain Fluorescence Lifetime Imaging Microscopy (FLIM) to imaging in microfluidic devices. FLIM is performed using low cost, intensity modulated Light Emitting Diodes (LEDs) for illumination. The use of lifetime imaging for quantitative analysis within such devices is demonstrated by mapping the molecular diffusion of iodide ions across a microchannel.

©2006 Optical Society of America

1. Introduction

Microfluidic devices represent an excellent platform to implement a large number of chemical processing and sensing applications [17]. Their benefits derive from the fact that timescales for mixing, heating and flow manipulation are much reduced compared to macroscopic flow systems. Furthermore the small size means that little reagent consumption is required and integration with other technologies is readily achieved. A key factor in the usefulness of microfluidics is the availability of matching online diagnostic systems that permit as much physical and chemical information to be sampled as possible. Such analytical techniques require species selectivity, sensitivity, and a high spatial and temporal resolution to be useful for online process monitoring, sensing applications, and product characterisation inside microfluidic channels.

Optical techniques are very powerful in this context and in the past techniques have focused on fluorescence microscopy [8], light microscopy [9], confocal laser scanning microscopy [10], and optical coherence tomography [11]. Of these, fluorescence based techniques are among the most versatile and certainly they are the most widely employed analytical tools in this context. Fluorescence imaging offers information on spatial and temporal scales of concentration, and is easy to implement. There are drawbacks however, relating to signal quantification, which is affected by the presence of local quencher species and also by signal reabsorption. Furthermore there is a requirement for the analyte to have clearly distinguishable absorption and emission spectra. Often this requires elaborate labelling techniques for one to obtain specificity and/or sensitivity - something that is not always a feasible option. In multicomponent mixtures, spectral overlap may cause ambiguity preventing a discrimination of the different components. In such situations the fluorescence lifetime of a sample may give additional or superior information.

The fluorescence lifetime represents the average time a fluorophore will remain in an excited state upon absorption of light before emitting a fluorescence photon. It is a sensitive indicator of a molecule’s interaction with its immediate environment: The presence of local quencher species, pH mediated conformation changes, molecular association and aggregation, diffusional mobilities, etc. all affect the fluorescence lifetime and its measurement thus provides information beyond that of purely intensity based diagnostic techniques.

Furthermore, components whose spectral signatures are too similar to permit their discrimination can often be differentiated according to their lifetime differences. There is thus a strong case to be made to combine the benefits of fluorescence lifetime imaging with microfluidic device applications. To date, however, there have been very few reported applications of FLIM (fluorescence lifetime imaging microscopy) to microfluidic devices [12,13]. Benninger et al. recently reported on the use of two photon excited time resolved Fluorescence Anisotropy and lifetime imaging in microfluidic devices to demonstrate its use for the study of molecular mixing and rotational diffusion effects in microfluidic channels. The system employed an Ar ion pumped Ti:Sapphire laser and an ultrafast image intensification device to achieve time resolution and the usefulness of lifetime resolved imaging was clearly demonstrated. Magennis et al. used a similar system comprising of a gated intensified CCD camera and an ultrafast pulsed laser to investigate diffusional mixing.

In this paper we report the application of a novel, Frequency Domain Lifetime Imaging system to microfluidic imaging employing inexpensive incoherent LED sources for excitation, whose intensity is modulated at high frequency (40 MHz). Using a gain modulated wide field detection system in a scheme analogous to homodyne detection, the fluorescence lifetime of the sample can be deduced from the phase shift (and demodulation) of the signal with respect to the excitation waveform [14,15]. The fluorescence lifetime deduced is a single exponential approximation to the true lifetime decay, which for many biological systems may be multi-exponential in nature. Multi-exponential lifetimes can be deduced by taking measurements at several different modulation frequencies and fitting a multi-exponential model to the data, or in special cases by use of techniques such as AB-plot analysis [16]. The system has great potential for miniaturisation and integration with microfluidic technology in a “lab on chip” approach and permits accurate lifetime differences at 100 ps and below to be clearly resolved. We demonstrate quantitative use of the system by imaging the lifetime changes of Rhodamine 6G fluorescent dye (which has a single exponential decay) in a dual inlet microreactor as iodide ions diffuse across two parallel aqueous streams.

2. Material and methods

2.1 Microfluidic device fabrication

A microchannel stamp as detailed in Fig. 1, was fabricated on cleaned wafers using standard photolithographic techniques [17,18,19,20]. This stamp was then used to create a microchannel structure in Poly Dimethylsiloxane (PDMS) using soft lithography techniques (Fig. 2). This fabrication method is extremely flexible with respect to channel design whilst also allowing repeated use of the same stamp. Additionally PDMS was noted as a suitable substrate for the channel fabrication as no background fluorescence was observed during the studies described at the wavelengths used.

 

Fig. 1. Design of the dual-inlet stamp used to create PDMS microchannel structure.

Download Full Size | PPT Slide | PDF

The process involved use of a technical drawing software package to create a high resolution acetate mask of the microchannel design (printed by Circuit Graphics, Chelmsford). Thin films of approximately 250µm Microchem SU8-2100 (Microchem) were coated over the glass wafers using a spin coater (Karl Suss Delta 10) and were pre-baked at 65°C for 12 minutes and 95°C for 60 minutes. The acetate mask was used to project an ultraviolet (UV) light source (340nm) from a mask aligner (Karl Suss MJB3) onto the substrates coated with the film of photoresist. The exposed films of photoresist were then post baked at 65°C and 95°C for 1 minute and 21 minutes respectively, before revealing the photoresist pattern using a developer solution (EC Solvent, Microchem). PDMS (Sylgard 184, Dow Corning) was then poured over the SU-8 master and allowed to cure. Once hardened, the PDMS was peeled off of the master, inlets and outlet attached and sealed onto a clean glass wafer using the natural adhesivesness of the PDMS to create a seal. Figure 2 illustrates the steps involved in fabricating the photoresist pattern.

 

Fig. 2. Microfabrication of dual inlet channel using photolithography and soft lithography.

Download Full Size | PPT Slide | PDF

2.2 FLIM system

The FLIM system (LIFA, Lambert Instruments, Leutingewolde, The Netherlands) [21] was attached to an Olympus IX70 inverted microscope (Olympus UK Ltd, Southall, UK). The excitation light is provided by a high power LED (3W) which can be modulated at frequencies up to 40 MHz. Images are detected by an ICCD whose gain is modulated at the same frequency. To obtain a lifetime image a series of intensity images are acquired at different detector gain phase offsets (typically 8–12 steps). The demodulation and phase shift of the fluorescence light are a function of the lifetime and can be determined over the entire field of view by fitting a sine function to every pixel along the set of images [22]. The absolute phase of the system is determined by taking a reference measurement of a standard with a known lifetime.

For Rhodamine 6G measurements we used an LED centred at 473 nm. The excitation light was filtered by a 470–490 nm bandpass filter and the emitted fluorescence light was filtered by a 500 nm cut-off dichroic mirror and a 515 nm longpass filter. A 20×, NA=0.70 objective was used for the imaging. The setup allows lifetime images to be acquired in approximately 1 second.

2.3 FLIM data analysis

All data analysis was performed using software written by the authors in IDL 6.1 (Interactive Data Language, Research Systems, Inc., Boulder, CO). Lifetimes can be determined from either the modulation depth m or the phase shift ϕ, referred to as modulation (τm) and phase lifetime (τϕ), respectively. The lifetimes returned are single exponential approximations of the true fluorescence lifetimes. In the case of a single exponential decay (as for Rhodamine 6G) the phase and modulation lifetimes are identical. For a multi-exponential decay the phase lifetime will be shorter than the modulation lifetime. With a modulation frequency f the modulation lifetime is given by

τm=12πf1m21

and the phase lifetime by

τϕ=tanϕ2πf

To account for the phase shift and demodulation introduced by the electronics and optics it is usually necessary to perform measurements relative to a reference sample with a known fluorescence lifetime, τref, using the following equations:

τϕ=12πftan((ϕϕref)+tan1(2πfτref))
τm=12πf(mref2m2(1+(2πf)2τref2)1)12

where ϕref, mref and τref refer to the phase shift, demodulation and lifetime of the reference respectively.

The optimal frequency in terms of lifetime selectivity can be derived from Equations 1 and 2 by differentiation. This results in optimal frequencies foptϕ=1/(2πτ) for the phase lifetime and fopt,m=2(2πτ) for the modulation lifetime.

The modulation bandwidths of the LEDs used here is limited to around 40 MHz. This is far from ideal for the shortest lifetimes we observe here (0.5 ns) especially for determinations of τm. Demodulations may to levels of a percent or so, making τm measurements susceptible to noise and baseline drifts. Determination of phase lifetimes, however, remains relatively robust at theses frequencies and as a consequence all data presented here is from phase lifetimes.

2.4 Rhodamine 6G solutions and flow set-up

Iodide quenching of Rhodamine 6G has been used as a convenient method of producing mono-exponential decay standards with variable lifetimes [23,24]. This quenching process follows a Stern-Vollmer model, in which the observed lifetime τ is given by

1τ=dq[Q]+1τ0

where τ 0 is the lifetime of Rhodamine 6G in the absence of quencher, [Q] is the concentration of the quencher, and kq is the bimolecular reaction rate constant for the reaction of the excited state with the quencher Q. Previously reported values for Rhodamine 6G are approximately τ 0=4.1 ns.

Rhodamine 6G was first prepared as a stock solution in methanol. This was added to a 0.2 M solution of potassium chloride (KCl) to a dye concentration of 1 µM. A solution of 0.05 M KI and 0.15 M KCl was prepared to be used as the quenching agent. Solutions containing 0.1 M KI and 0.1 M KCl, 0.03 M KI and 0.17 M KCl, 0.005 M KI and 0.195 M KCl were also prepared in order to calibrate the bimolecular reaction rate constant.

Flow in the microchannel device was controlled by a syringe pump (Harvard apparatus, PHD 2000), and was approximately 10µl/min.

In order to examine the flow pattern the channel was imaged just after the inlets using a confocal microscope (Olympus FV300, Olympus UK Ltd, Southall, UK) by taking a stack of images from the top to the bottom of the channel (Fig. 3). In order to clearly distinguish the two parallel streams both contained 0.2M KCl solutions but the dye, Rhodamine 6G, was only present in one.

 

Fig. 3. Confocal 3D image of two parallel aqueous streams flowing down a microchannel, just after the inlets. Only the left flow contained fluorescent dye. The signals seen on the right define the channel wall and stem from residual dye attached to the wall from previous experiments. The loss of fluorescence intensity in the upper part of the channel is due to reabsorption and optical distortion near the walls. The wider profile of the upper part of the fluorescent stream is due to small imperfections in the geometry of the microchannel and its inlets.

Download Full Size | PPT Slide | PDF

3. Results

We used the Rhodamine 6G solution without quencher as the reference standard to calibrate the FLIM system and assumed a lifetime of 4.08 ns [24].

In order to determine the bimolecular reaction rate constant the lifetimes of solutions with different iodide concentrations were imaged in a black well plate. Figure 4 shows a plot of measured lifetime times KI concentration against lifetime. Theory predicts a linear relationship with gradient -kqτ 0 and intercept τ 0. From the graph we determine the bimolecular reaction rate constant to be 6.64×109 dm3 mole-1 s-1.

 

Fig. 4. Calibration plot to determine the Stern-Vollmer quenching parameters for Rhodamine 6G solutions with KI. Each solution contains a different concentration of KI, but the same ionic strength through addition of KCl. From the gradient and intercept the bimolecular reaction rate constant, kq, is found to be 6.64×109 dm3 mole-1 s-1.

Download Full Size | PPT Slide | PDF

Lifetime images taken at increasing distances along the length of the channel are presented in Fig. 5. Profiles of the lifetime and calculated iodide concentration perpendicular to the flow direction for increasing distances down the channel are shown in Fig. 6.

 

Fig. 5. Lifetime images taken at increasing distances along the length of the channel. The increasing degree of mixing with increasing distance is clearly seen. [Media 1]

Download Full Size | PPT Slide | PDF

 

Fig. 6. (a) Plot of lifetime profiles perpendicular to the flow direction for 5 different distances down the channel. Initially one sees a sharper gradient, which gradually becomes less distinct. (b) Concentration profiles derived from the lifetime data according to Equation 3.

Download Full Size | PPT Slide | PDF

4. Discussion

The 3D images from the confocal microscope show a very clear and sharp interface between the two flows just after the inlets which is in agreement with what one would expect from a highly laminar flow (Fig. 3). This confirms that the microchannel device is very well suited for studies of molecular mixing by diffusion across the flow boundary. The confocal images point to the problems associated with fluorescence intensity imaging: variations in the images occur due to absorption of the excitation light, reabsorption of the fluorescence light, and optical distortions when imaging deep into the channel. The optical distortions are the result of either spherical or chromatic aberrations caused by focussing using a thicker effective coverslip (required to seal the channel) than the objective was designed for, or a combination of both. All these issues complicate quantitative evaluation of intensity images.

The lifetime images initially show a sharp concentration gradient between the two streams (Fig. 5). As the distance down the channel is increased, the diffusion gradient becomes less well defined, as one would expect from diffusional mixing (Fig. 6). Of particular note is the point where all the lines intersect - at a distance of approximately 290 microns across the channel and at a concentration of 0.021 M KI. The concentration of 0.021 M KI represents the final uniform concentration that the channel will eventually reach. The distance of 290 microns across the channel represents the interface between the two streams, and from its position we can infer the relative flowrates of each stream. By comparing the relative flow rate in each stream multiplied by its initial KI concentration ((0.0×290500)+(0.05×500290500)=0.021) with the final ‘mixed’ concentration (0.021) we see very good agreement.

In Fig. 6 the concentration of KI does not reach the maximum of 0.05 M. We believe that this is because of the lack of z (depth) resolution in the images. This means that out of focus light, from a slower moving part of the flow is collected. Because the flow is slower, there has been more time for diffusion and so the KI concentration is reduced. Because the KI concentration is lower, the signal is not only from a higher lifetime but it is also of a brighter intensity, which weights the lifetime signal more in favour of the longer lifetime.

The advantage of lifetime data is clear when considering quantitative interpretation of the results, in our case determination of the concentration of iodide by means of Eqn. 3, owing to the direct relationship between lifetime and concentration. For high resolution 3D mapping of the concentration higher z-resolution would be advantageous. This can be achieved by wide-field optical sectioning techniques, either by means of deconvolution techniques [25] or by the use of structured illumination [26]. Present work is focussed on implementing wide-field optical sectioning on our FLIM set up.

For our system the minimum detectable lifetime difference between a small region of 100 pixels and a large region was evaluated from simulations and found to be around 100 ps when imaging with a SNR (signal-to-noise ratio) of 50 and taking measurements at 12 phase steps [27]. The simulations were performed by generating intensity images corresponding to assumed lifetime distributions and then adding normally distributed noise with a SNR of 50 to the data for every pixel and every phase step. This level of minimum detectable lifetime difference corresponds well with the level of noise seen in the channel profiles in Fig. 6. In the KI concentration region most sensitive to lifetime changes such a change in lifetime would be equivalent to a concentration change of approximately 0.001 M. This illustrates the potential sensitivity of lifetime measurements.

The FLIM system conveniently covers the entire visible spectrum with the use of inexpensive, easily interchangable LEDs. This is in contrast to previous implementations of FLIM systems which have utilised expensive and complex pulsed laser systems or high-frequency optical modulators. The use of low cost LEDs opens the gate for widespread use of FLIM as an analytical tool, and provides the potential for integration in “lab-on-a-chip” type devices. A further development that would again reduces the potential cost of FLIM technology is the use of unintensified rapid-modulation gain camera systems [28].

5. Conclusion

To summarise, we have demonstrated the usefulness of frequency domain FLIM for the study of flows in microchannels by imaging changes in lifetime of Rhodamine 6G due to the diffusion of iodide ions. Changes in fluorescence lifetime profiles were seen as a function of distance along the channel. The lifetime images provide excellent contrast between the two parallel flows and permit detailed studies of molecular mixing to be performed. We found that contrast is obtained in situations where intensity based measurements lack specificity and sensitivity and thus the work shown provides an alternative and complementary diagnostic for the study of such systems. The results were obtained with a high power diode as illumination source, the intensity of which was modulated at up to 40 MHz. The work shows that sensitive lifetime information on a microscopic scale is obtainable with such devices and the technology is cost effective and simple to implement.

Acknowledgments

Clemens Kaminski gratefully acknowledges support from the Leverhulme Trust, UK.

Jonathan Frank and Clemens Kaminski gratefully acknowledge support for this work by EPSRC via a PLATFORM grant (GR/R98679/01). Jonathan Frank acknowledges support from the U.S. Department of Energy, Office of Basic Energy Sciences, Division of Chemical Sciences, Geosciences, and Biosciences.

This work was supported by grants received from the Royal Society, Syngenta Ltd, the EPSRC (Engineering and Physical Sciences Research Council, UK), and HEFCE (The Higher Education Funding Council for England, UK).

References

1. D.J. Beebe, G.A. Mensing, and G.M. Walker, “Physics and Applications of Microfluidics in Biology,” Annu. Rev. Biomed. Eng. 4, 261–286 (2002). [CrossRef]   [PubMed]  

2. S.C. Jakeway, A.J. de Mello, and E.L. Russell, “Miniaturized total analysis systems for biological analysis,” Fresenius J. Anal. Chem. 366, 525–539 (2000). [CrossRef]  

3. T. Chovan and A. Guttman, “Microfabricated devices in biotechnology and biochemical processing,” Trends Biotechnol. 20, 116–122 (2002). [CrossRef]   [PubMed]  

4. A.J. Tüdős, G.A.J. Besselink, and R.B.M. Schasfoort, “Trends in miniaturized total analysis systems for point-of-care testing in clinical chemistry,” Lab. Chip 1, 83–96 (2001). [CrossRef]  

5. E. Verpoorte, “Microfluidic chips for clinical and forensic analysis,” Electrophoresis 23, 677–712 (2002). [CrossRef]   [PubMed]  

6. Y. Huang, E.L. Mather, J.L. Bell, and M. Madou, “MEMS-based sample preparation for molecular diagnostics,” Anal. Bioanal. Chem. 372, 49–65 (2002). [CrossRef]   [PubMed]  

7. T. Vo-Dinh and B. Cullum, “Biosensors and biochips: advances in biological and medical diagnostics,” Fresenius J. Anal. Chem. 366, 540–551 (2000). [CrossRef]  

8. Y. Sato, G. Irisawa, M. Ishizuka, K. Hishida, and M. Maeda, “Visualization of convective mixing in microchannel by fluorescence imaging,” Meas. Sci. Technol. 14, 114–122 (2003). [CrossRef]  

9. R.H. Liu, M.A. Stremler, K.V. Sharp, M.G. Olsen, J.G. Santiago, R.J. Adrian, H. Aref, and D.J. Beebe, “Passive mixing in a three-dimensional serpentine microchannel,” J. Microelectromech. Syst. 9, 190–197 (2000). [CrossRef]  

10. R.F. Ismagilov, A.D. Stroock, P.J.A. Kenis, G. Whitesides, and H.A. Stone, “Experimental and theoretical scaling laws for transverse diffusive broadening in two-phase laminar flows in microchannels,” Appl. Phys. Lett. 76, 2376–2378 (2000). [CrossRef]  

11. C. Xi, D.L. Marks, D.S. Parikh, L. Raskin, and S.A Boppart, “Structural and functional imaging of 3D microfluidic mixers using optical coherence tomography,” P. Natl. Acad. Sci. USA 101, 7516–7521 (2004). [CrossRef]  

12. R.K.P. Benninger, O. Hofmann, J. McGinty, J. Requejo-Isidro, I. Munro, M.A.A. Neil, A.J. deMello, and P.M.W. French, “Time-resolved fluorescence imaging of solvent interactions in microfluidic devices,” Opt. Express 13, 6275–6285 (2005). [CrossRef]   [PubMed]  

13. S.W. Magennis, E.M. Graham, and A.C. Jones, “Quantitative Spatial Mapping of Mixing in Microfluidic Systems,” Angewandte Chemie International Edition 44, 6512–6516 (2005). [CrossRef]  

14. J. R. Lakowicz and K. W. Berndt, “Lifetime-Selective Fluorescence Imaging Using An Rf Phase- Sensitive Camera,” Rev. Sci. Instrum. 62, 1727–1734 (1991). [CrossRef]  

15. G. Marriott, R.M. Clegg, D.J. Arndt-Jovin, and T.M. Jovin, “Time resolved imaging microscopy. Phosphorescence and delayed fluorescence imaging,” Biophys. J. 60, 1374–1387 (1991). [CrossRef]   [PubMed]  

16. Q. S. Hanley and A. H. A. Clayton, “AB-plot assisted determination of fluorophore mixtures in a fluorescence lifetime microscope using spectra or quenchers,” J. Microscopy 218, 62–67 (2005). [CrossRef]  

17. D. A. Jeong, G. Markle, F. Owen, A. Pease, and R. Von Bünau Grenville, “The future of optical lithography,” Solid State Technol. 37, 39–47 (1994).

18. M. D. Levenson, “Extending optical lithography to the gigabit era,” Solid State Technol. 38, 57–66 (1995).

19. L. Geppert, “Semiconductor lithography for the next millennium,” IEEE Spectrum 33, 33–38 (1996). [CrossRef]  

20. S. Okazaki, “Resolution limits of optical lithography,” J. Vac. Sci. Technol. B 9, 2829–2833 (1991). [CrossRef]  

21. LK van Geest and KWJ Stoop, “FLIM on a wide field fluorescence microscope,” Lett. Peptide Sci. 10, 501–510 (2003). [CrossRef]  

22. Robert M. Clegg, Thomas M. Jovin Theodorus, and W J Gadella Jr, “Fluorescence lifetime imaging microscopy: Pixel-by-pixel analysis of phase-modulation data,” Bioimaging 2, 139–159 (1994). [CrossRef]  

23. J.M. Harris and F.E. Lytle, “Measurement of subnanosecond fluorescence decays by sampled single-photon detection,” Rev. Sci. Instrum. 48, 1470–1476 (1977). [CrossRef]  

24. Q. S. Hanley, V. Subramaniam, and D. J. Arndt-Jovin, “Fluorescence lifetime imaging: multi-point calibration, minimum resolvable differences, and artifact suppression,” Cytometry 43, 248–260 (2001). [CrossRef]   [PubMed]  

25. A. Squire and P.I.H. Bastiaens, “Three dimensional image restoration in fluorescence lifetime imaging microscopy,” J. Mircoscopy 193, 36–49 (1999). [CrossRef]  

26. M.J. Cole, J. Siegel, S.E.D. Webb, R. Jones, K. Dowling, P.M.W. French, M.J. Lever, L.O.D. Sucharov, M.A.A. Neil, R. Juskaitis, and T. Wilson, “Whole-field optically sectioned fluorescence lifetime imaging,” Opt. Lett. 25, 1361–1363 (2000). [CrossRef]  

27. A. Elder, J. Frank, J. Swartling, X. Dai, and C.F. Kaminski, “Calibration of a wide-field frequency-domain fluorescence lifetime microscopy system using light emitting diodes as light sources,” J. Microscopy2006 (accepted for publication). [CrossRef]  

28. A. C. Mitchell, J. E. Wall, and J. G. Murray, “Direct modulation of the effective sensitivity of a CCD detector: a new approach to time-resolved fluorescence imaging,” J. Microsc. Oxf. 206, 225–232 (2002). [CrossRef]  

References

  • View by:
  • |
  • |
  • |

  1. D.J. Beebe, G.A. Mensing, and G.M. Walker, “Physics and Applications of Microfluidics in Biology,” Annu. Rev. Biomed. Eng. 4, 261–286 (2002).
    [Crossref] [PubMed]
  2. S.C. Jakeway, A.J. de Mello, and E.L. Russell, “Miniaturized total analysis systems for biological analysis,” Fresenius J. Anal. Chem. 366, 525–539 (2000).
    [Crossref]
  3. T. Chovan and A. Guttman, “Microfabricated devices in biotechnology and biochemical processing,” Trends Biotechnol. 20, 116–122 (2002).
    [Crossref] [PubMed]
  4. A.J. Tüdős, G.A.J. Besselink, and R.B.M. Schasfoort, “Trends in miniaturized total analysis systems for point-of-care testing in clinical chemistry,” Lab. Chip 1, 83–96 (2001).
    [Crossref]
  5. E. Verpoorte, “Microfluidic chips for clinical and forensic analysis,” Electrophoresis 23, 677–712 (2002).
    [Crossref] [PubMed]
  6. Y. Huang, E.L. Mather, J.L. Bell, and M. Madou, “MEMS-based sample preparation for molecular diagnostics,” Anal. Bioanal. Chem. 372, 49–65 (2002).
    [Crossref] [PubMed]
  7. T. Vo-Dinh and B. Cullum, “Biosensors and biochips: advances in biological and medical diagnostics,” Fresenius J. Anal. Chem. 366, 540–551 (2000).
    [Crossref]
  8. Y. Sato, G. Irisawa, M. Ishizuka, K. Hishida, and M. Maeda, “Visualization of convective mixing in microchannel by fluorescence imaging,” Meas. Sci. Technol. 14, 114–122 (2003).
    [Crossref]
  9. R.H. Liu, M.A. Stremler, K.V. Sharp, M.G. Olsen, J.G. Santiago, R.J. Adrian, H. Aref, and D.J. Beebe, “Passive mixing in a three-dimensional serpentine microchannel,” J. Microelectromech. Syst. 9, 190–197 (2000).
    [Crossref]
  10. R.F. Ismagilov, A.D. Stroock, P.J.A. Kenis, G. Whitesides, and H.A. Stone, “Experimental and theoretical scaling laws for transverse diffusive broadening in two-phase laminar flows in microchannels,” Appl. Phys. Lett. 76, 2376–2378 (2000).
    [Crossref]
  11. C. Xi, D.L. Marks, D.S. Parikh, L. Raskin, and S.A Boppart, “Structural and functional imaging of 3D microfluidic mixers using optical coherence tomography,” P. Natl. Acad. Sci. USA 101, 7516–7521 (2004).
    [Crossref]
  12. R.K.P. Benninger, O. Hofmann, J. McGinty, J. Requejo-Isidro, I. Munro, M.A.A. Neil, A.J. deMello, and P.M.W. French, “Time-resolved fluorescence imaging of solvent interactions in microfluidic devices,” Opt. Express 13, 6275–6285 (2005).
    [Crossref] [PubMed]
  13. S.W. Magennis, E.M. Graham, and A.C. Jones, “Quantitative Spatial Mapping of Mixing in Microfluidic Systems,” Angewandte Chemie International Edition 44, 6512–6516 (2005).
    [Crossref]
  14. J. R. Lakowicz and K. W. Berndt, “Lifetime-Selective Fluorescence Imaging Using An Rf Phase- Sensitive Camera,” Rev. Sci. Instrum. 62, 1727–1734 (1991).
    [Crossref]
  15. G. Marriott, R.M. Clegg, D.J. Arndt-Jovin, and T.M. Jovin, “Time resolved imaging microscopy. Phosphorescence and delayed fluorescence imaging,” Biophys. J. 60, 1374–1387 (1991).
    [Crossref] [PubMed]
  16. Q. S. Hanley and A. H. A. Clayton, “AB-plot assisted determination of fluorophore mixtures in a fluorescence lifetime microscope using spectra or quenchers,” J. Microscopy 218, 62–67 (2005).
    [Crossref]
  17. D. A. Jeong, G. Markle, F. Owen, A. Pease, and R. Von Bünau Grenville, “The future of optical lithography,” Solid State Technol. 37, 39–47 (1994).
  18. M. D. Levenson, “Extending optical lithography to the gigabit era,” Solid State Technol. 38, 57–66 (1995).
  19. L. Geppert, “Semiconductor lithography for the next millennium,” IEEE Spectrum 33, 33–38 (1996).
    [Crossref]
  20. S. Okazaki, “Resolution limits of optical lithography,” J. Vac. Sci. Technol. B 9, 2829–2833 (1991).
    [Crossref]
  21. LK van Geest and KWJ Stoop, “FLIM on a wide field fluorescence microscope,” Lett. Peptide Sci. 10, 501–510 (2003).
    [Crossref]
  22. Robert M. Clegg, Thomas M. Jovin Theodorus, and W J Gadella, “Fluorescence lifetime imaging microscopy: Pixel-by-pixel analysis of phase-modulation data,” Bioimaging 2, 139–159 (1994).
    [Crossref]
  23. J.M. Harris and F.E. Lytle, “Measurement of subnanosecond fluorescence decays by sampled single-photon detection,” Rev. Sci. Instrum. 48, 1470–1476 (1977).
    [Crossref]
  24. Q. S. Hanley, V. Subramaniam, and D. J. Arndt-Jovin, “Fluorescence lifetime imaging: multi-point calibration, minimum resolvable differences, and artifact suppression,” Cytometry 43, 248–260 (2001).
    [Crossref] [PubMed]
  25. A. Squire and P.I.H. Bastiaens, “Three dimensional image restoration in fluorescence lifetime imaging microscopy,” J. Mircoscopy 193, 36–49 (1999).
    [Crossref]
  26. M.J. Cole, J. Siegel, S.E.D. Webb, R. Jones, K. Dowling, P.M.W. French, M.J. Lever, L.O.D. Sucharov, M.A.A. Neil, R. Juskaitis, and T. Wilson, “Whole-field optically sectioned fluorescence lifetime imaging,” Opt. Lett. 25, 1361–1363 (2000).
    [Crossref]
  27. A. Elder, J. Frank, J. Swartling, X. Dai, and C.F. Kaminski, “Calibration of a wide-field frequency-domain fluorescence lifetime microscopy system using light emitting diodes as light sources,” J. Microscopy2006 (accepted for publication).
    [Crossref]
  28. A. C. Mitchell, J. E. Wall, and J. G. Murray, “Direct modulation of the effective sensitivity of a CCD detector: a new approach to time-resolved fluorescence imaging,” J. Microsc. Oxf. 206, 225–232 (2002).
    [Crossref]

2005 (3)

R.K.P. Benninger, O. Hofmann, J. McGinty, J. Requejo-Isidro, I. Munro, M.A.A. Neil, A.J. deMello, and P.M.W. French, “Time-resolved fluorescence imaging of solvent interactions in microfluidic devices,” Opt. Express 13, 6275–6285 (2005).
[Crossref] [PubMed]

S.W. Magennis, E.M. Graham, and A.C. Jones, “Quantitative Spatial Mapping of Mixing in Microfluidic Systems,” Angewandte Chemie International Edition 44, 6512–6516 (2005).
[Crossref]

Q. S. Hanley and A. H. A. Clayton, “AB-plot assisted determination of fluorophore mixtures in a fluorescence lifetime microscope using spectra or quenchers,” J. Microscopy 218, 62–67 (2005).
[Crossref]

2004 (1)

C. Xi, D.L. Marks, D.S. Parikh, L. Raskin, and S.A Boppart, “Structural and functional imaging of 3D microfluidic mixers using optical coherence tomography,” P. Natl. Acad. Sci. USA 101, 7516–7521 (2004).
[Crossref]

2003 (2)

Y. Sato, G. Irisawa, M. Ishizuka, K. Hishida, and M. Maeda, “Visualization of convective mixing in microchannel by fluorescence imaging,” Meas. Sci. Technol. 14, 114–122 (2003).
[Crossref]

LK van Geest and KWJ Stoop, “FLIM on a wide field fluorescence microscope,” Lett. Peptide Sci. 10, 501–510 (2003).
[Crossref]

2002 (5)

A. C. Mitchell, J. E. Wall, and J. G. Murray, “Direct modulation of the effective sensitivity of a CCD detector: a new approach to time-resolved fluorescence imaging,” J. Microsc. Oxf. 206, 225–232 (2002).
[Crossref]

E. Verpoorte, “Microfluidic chips for clinical and forensic analysis,” Electrophoresis 23, 677–712 (2002).
[Crossref] [PubMed]

Y. Huang, E.L. Mather, J.L. Bell, and M. Madou, “MEMS-based sample preparation for molecular diagnostics,” Anal. Bioanal. Chem. 372, 49–65 (2002).
[Crossref] [PubMed]

D.J. Beebe, G.A. Mensing, and G.M. Walker, “Physics and Applications of Microfluidics in Biology,” Annu. Rev. Biomed. Eng. 4, 261–286 (2002).
[Crossref] [PubMed]

T. Chovan and A. Guttman, “Microfabricated devices in biotechnology and biochemical processing,” Trends Biotechnol. 20, 116–122 (2002).
[Crossref] [PubMed]

2001 (2)

A.J. Tüdős, G.A.J. Besselink, and R.B.M. Schasfoort, “Trends in miniaturized total analysis systems for point-of-care testing in clinical chemistry,” Lab. Chip 1, 83–96 (2001).
[Crossref]

Q. S. Hanley, V. Subramaniam, and D. J. Arndt-Jovin, “Fluorescence lifetime imaging: multi-point calibration, minimum resolvable differences, and artifact suppression,” Cytometry 43, 248–260 (2001).
[Crossref] [PubMed]

2000 (5)

M.J. Cole, J. Siegel, S.E.D. Webb, R. Jones, K. Dowling, P.M.W. French, M.J. Lever, L.O.D. Sucharov, M.A.A. Neil, R. Juskaitis, and T. Wilson, “Whole-field optically sectioned fluorescence lifetime imaging,” Opt. Lett. 25, 1361–1363 (2000).
[Crossref]

S.C. Jakeway, A.J. de Mello, and E.L. Russell, “Miniaturized total analysis systems for biological analysis,” Fresenius J. Anal. Chem. 366, 525–539 (2000).
[Crossref]

T. Vo-Dinh and B. Cullum, “Biosensors and biochips: advances in biological and medical diagnostics,” Fresenius J. Anal. Chem. 366, 540–551 (2000).
[Crossref]

R.H. Liu, M.A. Stremler, K.V. Sharp, M.G. Olsen, J.G. Santiago, R.J. Adrian, H. Aref, and D.J. Beebe, “Passive mixing in a three-dimensional serpentine microchannel,” J. Microelectromech. Syst. 9, 190–197 (2000).
[Crossref]

R.F. Ismagilov, A.D. Stroock, P.J.A. Kenis, G. Whitesides, and H.A. Stone, “Experimental and theoretical scaling laws for transverse diffusive broadening in two-phase laminar flows in microchannels,” Appl. Phys. Lett. 76, 2376–2378 (2000).
[Crossref]

1999 (1)

A. Squire and P.I.H. Bastiaens, “Three dimensional image restoration in fluorescence lifetime imaging microscopy,” J. Mircoscopy 193, 36–49 (1999).
[Crossref]

1996 (1)

L. Geppert, “Semiconductor lithography for the next millennium,” IEEE Spectrum 33, 33–38 (1996).
[Crossref]

1995 (1)

M. D. Levenson, “Extending optical lithography to the gigabit era,” Solid State Technol. 38, 57–66 (1995).

1994 (2)

D. A. Jeong, G. Markle, F. Owen, A. Pease, and R. Von Bünau Grenville, “The future of optical lithography,” Solid State Technol. 37, 39–47 (1994).

Robert M. Clegg, Thomas M. Jovin Theodorus, and W J Gadella, “Fluorescence lifetime imaging microscopy: Pixel-by-pixel analysis of phase-modulation data,” Bioimaging 2, 139–159 (1994).
[Crossref]

1991 (3)

S. Okazaki, “Resolution limits of optical lithography,” J. Vac. Sci. Technol. B 9, 2829–2833 (1991).
[Crossref]

J. R. Lakowicz and K. W. Berndt, “Lifetime-Selective Fluorescence Imaging Using An Rf Phase- Sensitive Camera,” Rev. Sci. Instrum. 62, 1727–1734 (1991).
[Crossref]

G. Marriott, R.M. Clegg, D.J. Arndt-Jovin, and T.M. Jovin, “Time resolved imaging microscopy. Phosphorescence and delayed fluorescence imaging,” Biophys. J. 60, 1374–1387 (1991).
[Crossref] [PubMed]

1977 (1)

J.M. Harris and F.E. Lytle, “Measurement of subnanosecond fluorescence decays by sampled single-photon detection,” Rev. Sci. Instrum. 48, 1470–1476 (1977).
[Crossref]

Adrian, R.J.

R.H. Liu, M.A. Stremler, K.V. Sharp, M.G. Olsen, J.G. Santiago, R.J. Adrian, H. Aref, and D.J. Beebe, “Passive mixing in a three-dimensional serpentine microchannel,” J. Microelectromech. Syst. 9, 190–197 (2000).
[Crossref]

Aref, H.

R.H. Liu, M.A. Stremler, K.V. Sharp, M.G. Olsen, J.G. Santiago, R.J. Adrian, H. Aref, and D.J. Beebe, “Passive mixing in a three-dimensional serpentine microchannel,” J. Microelectromech. Syst. 9, 190–197 (2000).
[Crossref]

Arndt-Jovin, D. J.

Q. S. Hanley, V. Subramaniam, and D. J. Arndt-Jovin, “Fluorescence lifetime imaging: multi-point calibration, minimum resolvable differences, and artifact suppression,” Cytometry 43, 248–260 (2001).
[Crossref] [PubMed]

Arndt-Jovin, D.J.

G. Marriott, R.M. Clegg, D.J. Arndt-Jovin, and T.M. Jovin, “Time resolved imaging microscopy. Phosphorescence and delayed fluorescence imaging,” Biophys. J. 60, 1374–1387 (1991).
[Crossref] [PubMed]

Bastiaens, P.I.H.

A. Squire and P.I.H. Bastiaens, “Three dimensional image restoration in fluorescence lifetime imaging microscopy,” J. Mircoscopy 193, 36–49 (1999).
[Crossref]

Beebe, D.J.

D.J. Beebe, G.A. Mensing, and G.M. Walker, “Physics and Applications of Microfluidics in Biology,” Annu. Rev. Biomed. Eng. 4, 261–286 (2002).
[Crossref] [PubMed]

R.H. Liu, M.A. Stremler, K.V. Sharp, M.G. Olsen, J.G. Santiago, R.J. Adrian, H. Aref, and D.J. Beebe, “Passive mixing in a three-dimensional serpentine microchannel,” J. Microelectromech. Syst. 9, 190–197 (2000).
[Crossref]

Bell, J.L.

Y. Huang, E.L. Mather, J.L. Bell, and M. Madou, “MEMS-based sample preparation for molecular diagnostics,” Anal. Bioanal. Chem. 372, 49–65 (2002).
[Crossref] [PubMed]

Benninger, R.K.P.

Berndt, K. W.

J. R. Lakowicz and K. W. Berndt, “Lifetime-Selective Fluorescence Imaging Using An Rf Phase- Sensitive Camera,” Rev. Sci. Instrum. 62, 1727–1734 (1991).
[Crossref]

Besselink, G.A.J.

A.J. Tüdős, G.A.J. Besselink, and R.B.M. Schasfoort, “Trends in miniaturized total analysis systems for point-of-care testing in clinical chemistry,” Lab. Chip 1, 83–96 (2001).
[Crossref]

Boppart, S.A

C. Xi, D.L. Marks, D.S. Parikh, L. Raskin, and S.A Boppart, “Structural and functional imaging of 3D microfluidic mixers using optical coherence tomography,” P. Natl. Acad. Sci. USA 101, 7516–7521 (2004).
[Crossref]

Chovan, T.

T. Chovan and A. Guttman, “Microfabricated devices in biotechnology and biochemical processing,” Trends Biotechnol. 20, 116–122 (2002).
[Crossref] [PubMed]

Clayton, A. H. A.

Q. S. Hanley and A. H. A. Clayton, “AB-plot assisted determination of fluorophore mixtures in a fluorescence lifetime microscope using spectra or quenchers,” J. Microscopy 218, 62–67 (2005).
[Crossref]

Clegg, R.M.

G. Marriott, R.M. Clegg, D.J. Arndt-Jovin, and T.M. Jovin, “Time resolved imaging microscopy. Phosphorescence and delayed fluorescence imaging,” Biophys. J. 60, 1374–1387 (1991).
[Crossref] [PubMed]

Clegg, Robert M.

Robert M. Clegg, Thomas M. Jovin Theodorus, and W J Gadella, “Fluorescence lifetime imaging microscopy: Pixel-by-pixel analysis of phase-modulation data,” Bioimaging 2, 139–159 (1994).
[Crossref]

Cole, M.J.

Cullum, B.

T. Vo-Dinh and B. Cullum, “Biosensors and biochips: advances in biological and medical diagnostics,” Fresenius J. Anal. Chem. 366, 540–551 (2000).
[Crossref]

Dai, X.

A. Elder, J. Frank, J. Swartling, X. Dai, and C.F. Kaminski, “Calibration of a wide-field frequency-domain fluorescence lifetime microscopy system using light emitting diodes as light sources,” J. Microscopy2006 (accepted for publication).
[Crossref]

de Mello, A.J.

S.C. Jakeway, A.J. de Mello, and E.L. Russell, “Miniaturized total analysis systems for biological analysis,” Fresenius J. Anal. Chem. 366, 525–539 (2000).
[Crossref]

deMello, A.J.

Dowling, K.

Elder, A.

A. Elder, J. Frank, J. Swartling, X. Dai, and C.F. Kaminski, “Calibration of a wide-field frequency-domain fluorescence lifetime microscopy system using light emitting diodes as light sources,” J. Microscopy2006 (accepted for publication).
[Crossref]

Frank, J.

A. Elder, J. Frank, J. Swartling, X. Dai, and C.F. Kaminski, “Calibration of a wide-field frequency-domain fluorescence lifetime microscopy system using light emitting diodes as light sources,” J. Microscopy2006 (accepted for publication).
[Crossref]

French, P.M.W.

Gadella, W J

Robert M. Clegg, Thomas M. Jovin Theodorus, and W J Gadella, “Fluorescence lifetime imaging microscopy: Pixel-by-pixel analysis of phase-modulation data,” Bioimaging 2, 139–159 (1994).
[Crossref]

Geppert, L.

L. Geppert, “Semiconductor lithography for the next millennium,” IEEE Spectrum 33, 33–38 (1996).
[Crossref]

Graham, E.M.

S.W. Magennis, E.M. Graham, and A.C. Jones, “Quantitative Spatial Mapping of Mixing in Microfluidic Systems,” Angewandte Chemie International Edition 44, 6512–6516 (2005).
[Crossref]

Guttman, A.

T. Chovan and A. Guttman, “Microfabricated devices in biotechnology and biochemical processing,” Trends Biotechnol. 20, 116–122 (2002).
[Crossref] [PubMed]

Hanley, Q. S.

Q. S. Hanley and A. H. A. Clayton, “AB-plot assisted determination of fluorophore mixtures in a fluorescence lifetime microscope using spectra or quenchers,” J. Microscopy 218, 62–67 (2005).
[Crossref]

Q. S. Hanley, V. Subramaniam, and D. J. Arndt-Jovin, “Fluorescence lifetime imaging: multi-point calibration, minimum resolvable differences, and artifact suppression,” Cytometry 43, 248–260 (2001).
[Crossref] [PubMed]

Harris, J.M.

J.M. Harris and F.E. Lytle, “Measurement of subnanosecond fluorescence decays by sampled single-photon detection,” Rev. Sci. Instrum. 48, 1470–1476 (1977).
[Crossref]

Hishida, K.

Y. Sato, G. Irisawa, M. Ishizuka, K. Hishida, and M. Maeda, “Visualization of convective mixing in microchannel by fluorescence imaging,” Meas. Sci. Technol. 14, 114–122 (2003).
[Crossref]

Hofmann, O.

Huang, Y.

Y. Huang, E.L. Mather, J.L. Bell, and M. Madou, “MEMS-based sample preparation for molecular diagnostics,” Anal. Bioanal. Chem. 372, 49–65 (2002).
[Crossref] [PubMed]

Irisawa, G.

Y. Sato, G. Irisawa, M. Ishizuka, K. Hishida, and M. Maeda, “Visualization of convective mixing in microchannel by fluorescence imaging,” Meas. Sci. Technol. 14, 114–122 (2003).
[Crossref]

Ishizuka, M.

Y. Sato, G. Irisawa, M. Ishizuka, K. Hishida, and M. Maeda, “Visualization of convective mixing in microchannel by fluorescence imaging,” Meas. Sci. Technol. 14, 114–122 (2003).
[Crossref]

Ismagilov, R.F.

R.F. Ismagilov, A.D. Stroock, P.J.A. Kenis, G. Whitesides, and H.A. Stone, “Experimental and theoretical scaling laws for transverse diffusive broadening in two-phase laminar flows in microchannels,” Appl. Phys. Lett. 76, 2376–2378 (2000).
[Crossref]

Jakeway, S.C.

S.C. Jakeway, A.J. de Mello, and E.L. Russell, “Miniaturized total analysis systems for biological analysis,” Fresenius J. Anal. Chem. 366, 525–539 (2000).
[Crossref]

Jeong, D. A.

D. A. Jeong, G. Markle, F. Owen, A. Pease, and R. Von Bünau Grenville, “The future of optical lithography,” Solid State Technol. 37, 39–47 (1994).

Jones, A.C.

S.W. Magennis, E.M. Graham, and A.C. Jones, “Quantitative Spatial Mapping of Mixing in Microfluidic Systems,” Angewandte Chemie International Edition 44, 6512–6516 (2005).
[Crossref]

Jones, R.

Jovin, T.M.

G. Marriott, R.M. Clegg, D.J. Arndt-Jovin, and T.M. Jovin, “Time resolved imaging microscopy. Phosphorescence and delayed fluorescence imaging,” Biophys. J. 60, 1374–1387 (1991).
[Crossref] [PubMed]

Jovin Theodorus, Thomas M.

Robert M. Clegg, Thomas M. Jovin Theodorus, and W J Gadella, “Fluorescence lifetime imaging microscopy: Pixel-by-pixel analysis of phase-modulation data,” Bioimaging 2, 139–159 (1994).
[Crossref]

Juskaitis, R.

Kaminski, C.F.

A. Elder, J. Frank, J. Swartling, X. Dai, and C.F. Kaminski, “Calibration of a wide-field frequency-domain fluorescence lifetime microscopy system using light emitting diodes as light sources,” J. Microscopy2006 (accepted for publication).
[Crossref]

Kenis, P.J.A.

R.F. Ismagilov, A.D. Stroock, P.J.A. Kenis, G. Whitesides, and H.A. Stone, “Experimental and theoretical scaling laws for transverse diffusive broadening in two-phase laminar flows in microchannels,” Appl. Phys. Lett. 76, 2376–2378 (2000).
[Crossref]

Lakowicz, J. R.

J. R. Lakowicz and K. W. Berndt, “Lifetime-Selective Fluorescence Imaging Using An Rf Phase- Sensitive Camera,” Rev. Sci. Instrum. 62, 1727–1734 (1991).
[Crossref]

Levenson, M. D.

M. D. Levenson, “Extending optical lithography to the gigabit era,” Solid State Technol. 38, 57–66 (1995).

Lever, M.J.

Liu, R.H.

R.H. Liu, M.A. Stremler, K.V. Sharp, M.G. Olsen, J.G. Santiago, R.J. Adrian, H. Aref, and D.J. Beebe, “Passive mixing in a three-dimensional serpentine microchannel,” J. Microelectromech. Syst. 9, 190–197 (2000).
[Crossref]

Lytle, F.E.

J.M. Harris and F.E. Lytle, “Measurement of subnanosecond fluorescence decays by sampled single-photon detection,” Rev. Sci. Instrum. 48, 1470–1476 (1977).
[Crossref]

Madou, M.

Y. Huang, E.L. Mather, J.L. Bell, and M. Madou, “MEMS-based sample preparation for molecular diagnostics,” Anal. Bioanal. Chem. 372, 49–65 (2002).
[Crossref] [PubMed]

Maeda, M.

Y. Sato, G. Irisawa, M. Ishizuka, K. Hishida, and M. Maeda, “Visualization of convective mixing in microchannel by fluorescence imaging,” Meas. Sci. Technol. 14, 114–122 (2003).
[Crossref]

Magennis, S.W.

S.W. Magennis, E.M. Graham, and A.C. Jones, “Quantitative Spatial Mapping of Mixing in Microfluidic Systems,” Angewandte Chemie International Edition 44, 6512–6516 (2005).
[Crossref]

Markle, G.

D. A. Jeong, G. Markle, F. Owen, A. Pease, and R. Von Bünau Grenville, “The future of optical lithography,” Solid State Technol. 37, 39–47 (1994).

Marks, D.L.

C. Xi, D.L. Marks, D.S. Parikh, L. Raskin, and S.A Boppart, “Structural and functional imaging of 3D microfluidic mixers using optical coherence tomography,” P. Natl. Acad. Sci. USA 101, 7516–7521 (2004).
[Crossref]

Marriott, G.

G. Marriott, R.M. Clegg, D.J. Arndt-Jovin, and T.M. Jovin, “Time resolved imaging microscopy. Phosphorescence and delayed fluorescence imaging,” Biophys. J. 60, 1374–1387 (1991).
[Crossref] [PubMed]

Mather, E.L.

Y. Huang, E.L. Mather, J.L. Bell, and M. Madou, “MEMS-based sample preparation for molecular diagnostics,” Anal. Bioanal. Chem. 372, 49–65 (2002).
[Crossref] [PubMed]

McGinty, J.

Mensing, G.A.

D.J. Beebe, G.A. Mensing, and G.M. Walker, “Physics and Applications of Microfluidics in Biology,” Annu. Rev. Biomed. Eng. 4, 261–286 (2002).
[Crossref] [PubMed]

Mitchell, A. C.

A. C. Mitchell, J. E. Wall, and J. G. Murray, “Direct modulation of the effective sensitivity of a CCD detector: a new approach to time-resolved fluorescence imaging,” J. Microsc. Oxf. 206, 225–232 (2002).
[Crossref]

Munro, I.

Murray, J. G.

A. C. Mitchell, J. E. Wall, and J. G. Murray, “Direct modulation of the effective sensitivity of a CCD detector: a new approach to time-resolved fluorescence imaging,” J. Microsc. Oxf. 206, 225–232 (2002).
[Crossref]

Neil, M.A.A.

Okazaki, S.

S. Okazaki, “Resolution limits of optical lithography,” J. Vac. Sci. Technol. B 9, 2829–2833 (1991).
[Crossref]

Olsen, M.G.

R.H. Liu, M.A. Stremler, K.V. Sharp, M.G. Olsen, J.G. Santiago, R.J. Adrian, H. Aref, and D.J. Beebe, “Passive mixing in a three-dimensional serpentine microchannel,” J. Microelectromech. Syst. 9, 190–197 (2000).
[Crossref]

Owen, F.

D. A. Jeong, G. Markle, F. Owen, A. Pease, and R. Von Bünau Grenville, “The future of optical lithography,” Solid State Technol. 37, 39–47 (1994).

Parikh, D.S.

C. Xi, D.L. Marks, D.S. Parikh, L. Raskin, and S.A Boppart, “Structural and functional imaging of 3D microfluidic mixers using optical coherence tomography,” P. Natl. Acad. Sci. USA 101, 7516–7521 (2004).
[Crossref]

Pease, A.

D. A. Jeong, G. Markle, F. Owen, A. Pease, and R. Von Bünau Grenville, “The future of optical lithography,” Solid State Technol. 37, 39–47 (1994).

Raskin, L.

C. Xi, D.L. Marks, D.S. Parikh, L. Raskin, and S.A Boppart, “Structural and functional imaging of 3D microfluidic mixers using optical coherence tomography,” P. Natl. Acad. Sci. USA 101, 7516–7521 (2004).
[Crossref]

Requejo-Isidro, J.

Russell, E.L.

S.C. Jakeway, A.J. de Mello, and E.L. Russell, “Miniaturized total analysis systems for biological analysis,” Fresenius J. Anal. Chem. 366, 525–539 (2000).
[Crossref]

Santiago, J.G.

R.H. Liu, M.A. Stremler, K.V. Sharp, M.G. Olsen, J.G. Santiago, R.J. Adrian, H. Aref, and D.J. Beebe, “Passive mixing in a three-dimensional serpentine microchannel,” J. Microelectromech. Syst. 9, 190–197 (2000).
[Crossref]

Sato, Y.

Y. Sato, G. Irisawa, M. Ishizuka, K. Hishida, and M. Maeda, “Visualization of convective mixing in microchannel by fluorescence imaging,” Meas. Sci. Technol. 14, 114–122 (2003).
[Crossref]

Schasfoort, R.B.M.

A.J. Tüdős, G.A.J. Besselink, and R.B.M. Schasfoort, “Trends in miniaturized total analysis systems for point-of-care testing in clinical chemistry,” Lab. Chip 1, 83–96 (2001).
[Crossref]

Sharp, K.V.

R.H. Liu, M.A. Stremler, K.V. Sharp, M.G. Olsen, J.G. Santiago, R.J. Adrian, H. Aref, and D.J. Beebe, “Passive mixing in a three-dimensional serpentine microchannel,” J. Microelectromech. Syst. 9, 190–197 (2000).
[Crossref]

Siegel, J.

Squire, A.

A. Squire and P.I.H. Bastiaens, “Three dimensional image restoration in fluorescence lifetime imaging microscopy,” J. Mircoscopy 193, 36–49 (1999).
[Crossref]

Stone, H.A.

R.F. Ismagilov, A.D. Stroock, P.J.A. Kenis, G. Whitesides, and H.A. Stone, “Experimental and theoretical scaling laws for transverse diffusive broadening in two-phase laminar flows in microchannels,” Appl. Phys. Lett. 76, 2376–2378 (2000).
[Crossref]

Stoop, KWJ

LK van Geest and KWJ Stoop, “FLIM on a wide field fluorescence microscope,” Lett. Peptide Sci. 10, 501–510 (2003).
[Crossref]

Stremler, M.A.

R.H. Liu, M.A. Stremler, K.V. Sharp, M.G. Olsen, J.G. Santiago, R.J. Adrian, H. Aref, and D.J. Beebe, “Passive mixing in a three-dimensional serpentine microchannel,” J. Microelectromech. Syst. 9, 190–197 (2000).
[Crossref]

Stroock, A.D.

R.F. Ismagilov, A.D. Stroock, P.J.A. Kenis, G. Whitesides, and H.A. Stone, “Experimental and theoretical scaling laws for transverse diffusive broadening in two-phase laminar flows in microchannels,” Appl. Phys. Lett. 76, 2376–2378 (2000).
[Crossref]

Subramaniam, V.

Q. S. Hanley, V. Subramaniam, and D. J. Arndt-Jovin, “Fluorescence lifetime imaging: multi-point calibration, minimum resolvable differences, and artifact suppression,” Cytometry 43, 248–260 (2001).
[Crossref] [PubMed]

Sucharov, L.O.D.

Swartling, J.

A. Elder, J. Frank, J. Swartling, X. Dai, and C.F. Kaminski, “Calibration of a wide-field frequency-domain fluorescence lifetime microscopy system using light emitting diodes as light sources,” J. Microscopy2006 (accepted for publication).
[Crossref]

Tüdos, A.J.

A.J. Tüdős, G.A.J. Besselink, and R.B.M. Schasfoort, “Trends in miniaturized total analysis systems for point-of-care testing in clinical chemistry,” Lab. Chip 1, 83–96 (2001).
[Crossref]

van Geest, LK

LK van Geest and KWJ Stoop, “FLIM on a wide field fluorescence microscope,” Lett. Peptide Sci. 10, 501–510 (2003).
[Crossref]

Verpoorte, E.

E. Verpoorte, “Microfluidic chips for clinical and forensic analysis,” Electrophoresis 23, 677–712 (2002).
[Crossref] [PubMed]

Vo-Dinh, T.

T. Vo-Dinh and B. Cullum, “Biosensors and biochips: advances in biological and medical diagnostics,” Fresenius J. Anal. Chem. 366, 540–551 (2000).
[Crossref]

Von Bünau Grenville, R.

D. A. Jeong, G. Markle, F. Owen, A. Pease, and R. Von Bünau Grenville, “The future of optical lithography,” Solid State Technol. 37, 39–47 (1994).

Walker, G.M.

D.J. Beebe, G.A. Mensing, and G.M. Walker, “Physics and Applications of Microfluidics in Biology,” Annu. Rev. Biomed. Eng. 4, 261–286 (2002).
[Crossref] [PubMed]

Wall, J. E.

A. C. Mitchell, J. E. Wall, and J. G. Murray, “Direct modulation of the effective sensitivity of a CCD detector: a new approach to time-resolved fluorescence imaging,” J. Microsc. Oxf. 206, 225–232 (2002).
[Crossref]

Webb, S.E.D.

Whitesides, G.

R.F. Ismagilov, A.D. Stroock, P.J.A. Kenis, G. Whitesides, and H.A. Stone, “Experimental and theoretical scaling laws for transverse diffusive broadening in two-phase laminar flows in microchannels,” Appl. Phys. Lett. 76, 2376–2378 (2000).
[Crossref]

Wilson, T.

Xi, C.

C. Xi, D.L. Marks, D.S. Parikh, L. Raskin, and S.A Boppart, “Structural and functional imaging of 3D microfluidic mixers using optical coherence tomography,” P. Natl. Acad. Sci. USA 101, 7516–7521 (2004).
[Crossref]

Anal. Bioanal. Chem. (1)

Y. Huang, E.L. Mather, J.L. Bell, and M. Madou, “MEMS-based sample preparation for molecular diagnostics,” Anal. Bioanal. Chem. 372, 49–65 (2002).
[Crossref] [PubMed]

Angewandte Chemie International Edition (1)

S.W. Magennis, E.M. Graham, and A.C. Jones, “Quantitative Spatial Mapping of Mixing in Microfluidic Systems,” Angewandte Chemie International Edition 44, 6512–6516 (2005).
[Crossref]

Annu. Rev. Biomed. Eng. (1)

D.J. Beebe, G.A. Mensing, and G.M. Walker, “Physics and Applications of Microfluidics in Biology,” Annu. Rev. Biomed. Eng. 4, 261–286 (2002).
[Crossref] [PubMed]

Appl. Phys. Lett. (1)

R.F. Ismagilov, A.D. Stroock, P.J.A. Kenis, G. Whitesides, and H.A. Stone, “Experimental and theoretical scaling laws for transverse diffusive broadening in two-phase laminar flows in microchannels,” Appl. Phys. Lett. 76, 2376–2378 (2000).
[Crossref]

Bioimaging (1)

Robert M. Clegg, Thomas M. Jovin Theodorus, and W J Gadella, “Fluorescence lifetime imaging microscopy: Pixel-by-pixel analysis of phase-modulation data,” Bioimaging 2, 139–159 (1994).
[Crossref]

Biophys. J. (1)

G. Marriott, R.M. Clegg, D.J. Arndt-Jovin, and T.M. Jovin, “Time resolved imaging microscopy. Phosphorescence and delayed fluorescence imaging,” Biophys. J. 60, 1374–1387 (1991).
[Crossref] [PubMed]

Cytometry (1)

Q. S. Hanley, V. Subramaniam, and D. J. Arndt-Jovin, “Fluorescence lifetime imaging: multi-point calibration, minimum resolvable differences, and artifact suppression,” Cytometry 43, 248–260 (2001).
[Crossref] [PubMed]

Electrophoresis (1)

E. Verpoorte, “Microfluidic chips for clinical and forensic analysis,” Electrophoresis 23, 677–712 (2002).
[Crossref] [PubMed]

Fresenius J. Anal. Chem. (2)

S.C. Jakeway, A.J. de Mello, and E.L. Russell, “Miniaturized total analysis systems for biological analysis,” Fresenius J. Anal. Chem. 366, 525–539 (2000).
[Crossref]

T. Vo-Dinh and B. Cullum, “Biosensors and biochips: advances in biological and medical diagnostics,” Fresenius J. Anal. Chem. 366, 540–551 (2000).
[Crossref]

IEEE Spectrum (1)

L. Geppert, “Semiconductor lithography for the next millennium,” IEEE Spectrum 33, 33–38 (1996).
[Crossref]

J. Microelectromech. Syst. (1)

R.H. Liu, M.A. Stremler, K.V. Sharp, M.G. Olsen, J.G. Santiago, R.J. Adrian, H. Aref, and D.J. Beebe, “Passive mixing in a three-dimensional serpentine microchannel,” J. Microelectromech. Syst. 9, 190–197 (2000).
[Crossref]

J. Microsc. Oxf. (1)

A. C. Mitchell, J. E. Wall, and J. G. Murray, “Direct modulation of the effective sensitivity of a CCD detector: a new approach to time-resolved fluorescence imaging,” J. Microsc. Oxf. 206, 225–232 (2002).
[Crossref]

J. Microscopy (1)

Q. S. Hanley and A. H. A. Clayton, “AB-plot assisted determination of fluorophore mixtures in a fluorescence lifetime microscope using spectra or quenchers,” J. Microscopy 218, 62–67 (2005).
[Crossref]

J. Mircoscopy (1)

A. Squire and P.I.H. Bastiaens, “Three dimensional image restoration in fluorescence lifetime imaging microscopy,” J. Mircoscopy 193, 36–49 (1999).
[Crossref]

J. Vac. Sci. Technol. B (1)

S. Okazaki, “Resolution limits of optical lithography,” J. Vac. Sci. Technol. B 9, 2829–2833 (1991).
[Crossref]

Lab. Chip (1)

A.J. Tüdős, G.A.J. Besselink, and R.B.M. Schasfoort, “Trends in miniaturized total analysis systems for point-of-care testing in clinical chemistry,” Lab. Chip 1, 83–96 (2001).
[Crossref]

Lett. Peptide Sci. (1)

LK van Geest and KWJ Stoop, “FLIM on a wide field fluorescence microscope,” Lett. Peptide Sci. 10, 501–510 (2003).
[Crossref]

Meas. Sci. Technol. (1)

Y. Sato, G. Irisawa, M. Ishizuka, K. Hishida, and M. Maeda, “Visualization of convective mixing in microchannel by fluorescence imaging,” Meas. Sci. Technol. 14, 114–122 (2003).
[Crossref]

Opt. Express (1)

Opt. Lett. (1)

P. Natl. Acad. Sci. USA (1)

C. Xi, D.L. Marks, D.S. Parikh, L. Raskin, and S.A Boppart, “Structural and functional imaging of 3D microfluidic mixers using optical coherence tomography,” P. Natl. Acad. Sci. USA 101, 7516–7521 (2004).
[Crossref]

Rev. Sci. Instrum. (2)

J. R. Lakowicz and K. W. Berndt, “Lifetime-Selective Fluorescence Imaging Using An Rf Phase- Sensitive Camera,” Rev. Sci. Instrum. 62, 1727–1734 (1991).
[Crossref]

J.M. Harris and F.E. Lytle, “Measurement of subnanosecond fluorescence decays by sampled single-photon detection,” Rev. Sci. Instrum. 48, 1470–1476 (1977).
[Crossref]

Solid State Technol. (2)

D. A. Jeong, G. Markle, F. Owen, A. Pease, and R. Von Bünau Grenville, “The future of optical lithography,” Solid State Technol. 37, 39–47 (1994).

M. D. Levenson, “Extending optical lithography to the gigabit era,” Solid State Technol. 38, 57–66 (1995).

Trends Biotechnol. (1)

T. Chovan and A. Guttman, “Microfabricated devices in biotechnology and biochemical processing,” Trends Biotechnol. 20, 116–122 (2002).
[Crossref] [PubMed]

Other (1)

A. Elder, J. Frank, J. Swartling, X. Dai, and C.F. Kaminski, “Calibration of a wide-field frequency-domain fluorescence lifetime microscopy system using light emitting diodes as light sources,” J. Microscopy2006 (accepted for publication).
[Crossref]

Supplementary Material (1)

» Media 1: AVI (4783 KB)     

Cited By

OSA participates in Crossref's Cited-By Linking service. Citing articles from OSA journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (6)

Fig. 1.
Fig. 1. Design of the dual-inlet stamp used to create PDMS microchannel structure.
Fig. 2.
Fig. 2. Microfabrication of dual inlet channel using photolithography and soft lithography.
Fig. 3.
Fig. 3. Confocal 3D image of two parallel aqueous streams flowing down a microchannel, just after the inlets. Only the left flow contained fluorescent dye. The signals seen on the right define the channel wall and stem from residual dye attached to the wall from previous experiments. The loss of fluorescence intensity in the upper part of the channel is due to reabsorption and optical distortion near the walls. The wider profile of the upper part of the fluorescent stream is due to small imperfections in the geometry of the microchannel and its inlets.
Fig. 4.
Fig. 4. Calibration plot to determine the Stern-Vollmer quenching parameters for Rhodamine 6G solutions with KI. Each solution contains a different concentration of KI, but the same ionic strength through addition of KCl. From the gradient and intercept the bimolecular reaction rate constant, kq, is found to be 6.64×109 dm3 mole-1 s-1.
Fig. 5.
Fig. 5. Lifetime images taken at increasing distances along the length of the channel. The increasing degree of mixing with increasing distance is clearly seen. [Media 1]
Fig. 6.
Fig. 6. (a) Plot of lifetime profiles perpendicular to the flow direction for 5 different distances down the channel. Initially one sees a sharper gradient, which gradually becomes less distinct. (b) Concentration profiles derived from the lifetime data according to Equation 3.

Equations (5)

Equations on this page are rendered with MathJax. Learn more.

τ m = 1 2 π f 1 m 2 1
τ ϕ = tan ϕ 2 π f
τ ϕ = 1 2 π f tan ( ( ϕ ϕ ref ) + tan 1 ( 2 π f τ ref ) )
τ m = 1 2 π f ( m ref 2 m 2 ( 1 + ( 2 π f ) 2 τ ref 2 ) 1 ) 1 2
1 τ = d q [ Q ] + 1 τ 0