Abstract
We replicated results of the theoretical accuracy of the Howland crossed-cylinder aberroscope for measuring the monochromatic aberration of the human eye [J. Opt. Soc. Am. A 15, 24771998]. Two simple modifications to the previous mathematical analysis techniques, namely, the use of a new method for estimating the size of the unaberrated retinal grid element length and the correction of the aberroscope grid spacing size for magnification differences introduced during predistortion, almost completely eliminate the errors introduced into 4th-order wave-front-aberration coefficients for a centered predistorted aberroscope grid over a large range of back vertex distances. The use of the crossed cylinders in the aberroscope introduces only small amounts of wave-front aberration to the measurements of the schematic eye. Wave-front aberrations above the 4th order in a Taylor polynomial expression can introduce errors that vary as a function of pupil area. Results produced by orthogonal polynomial analysis confirm the least-squares analysis results. We believe that the Howland crossed-cylinder aberroscope can be used in its objective form to make accurate measurements of at least the 4th-order components of the wave-front aberration of the human eye. We note that care must be exercised both in the use of the equipment and in the analysis of the results to ensure that accuracy is maintained.
© 1999 Optical Society of America
Full Article | PDF ArticleMore Like This
George Smith, Raymond A. Applegate, and David A. Atchison
J. Opt. Soc. Am. A 15(9) 2477-2487 (1998)
G. Walsh, W. N. Charman, and H. C. Howland
J. Opt. Soc. Am. A 1(9) 987-992 (1984)
N. López-Gil, H. C. Howland, B. Howland, N. Charman, and R. Applegate
J. Opt. Soc. Am. A 15(9) 2563-2571 (1998)