Abstract
Optical coherence tomography is one of the standard imaging modalities at present, widely used in the medical and biological fields to obtain three-dimensional (3D) images with high spatial resolution. However, the depth up to which the 3D images can be directly obtained is limited to within 3 mm. Therefore, the suitability of many kinds of catheters and needles has been considered for minimally invasive imaging. We have examined the utility of a short multimode fiber (SMMF) using graded index optical fibers for minimal invasive imaging of deeper areas, up to 6–8 mm. The diameter and length of the SMMF are 125 μm and 6–8 mm, respectively. In the core of the SMMF, scattering and multirefraction occur due to small variations in the refractive index to generate deformations and depolarization of images. In order to investigate the depolarization characteristics, the images reflected at the facet of the SMMF were measured by changing the angle of the polarizer, using an LED as the light source. The reflection image almost corresponds to that obtained with combined linearly polarized modes with the ratio of equal to . Comparing the measured results with simulations in the simple model, the depolarization ratio was estimated at 0.7 in the core. The degrees of polarization were measured to be 0.15 around the center and increased to 0.90 at the periphery.
© 2018 Optical Society of America