Land surface temperature algorithms for the moderate resolution imaging spectroradiometer satellite instrument will require the spectral bidirectional reflectance distribution function (BRDF) of natural surfaces in the thermal infrared. We designed the spectral infrared bidirectional reflectance and emissivity instrument to provide such measurements by the use of a Fourier transform infrared spectrometer. A problem we encountered is the unavoidable surface heating caused by the source irradiance. For our system, the effects of the heating can cause a 30% error in the measured BRDF. The error caused by heating is corrected by temporally curve fitting the radiance signal. This curve-fitting technique isolates the radiance caused by reflected irradiance. With this correction, other factors dominate the BRDF error. It is now ~5% and can be improved further. The method is illustrated with measurements of soil BRDF.