Abstract
Frequency-domain fluorometry relies on the measurement of the phase and amplitudes of the Fourier components of the time-dependent fluorescence signal. Experimental results that show that a conventional photomultiplier is subject to intensity-dependent phase shifts are presented. The measurements indicate that this is a problem well below the maximum linear current of the photomultiplier response. These results have important implications in frequency-domain fluorescence anisotropy experiments, in which the parallel and the perpendicular components of the emission intensity are inherently different from one another: a phase shift can be introduced by the photomultiplier.
© 1993 Optical Society of America
Full Article | PDF ArticleMore Like This
Jian-Bang Liu and Chang-Sheng Liu
Appl. Opt. 30(13) 1688-1694 (1991)
Dmitri Lanevski, Koit Mauring, Eric R. Tkaczyk, and Raivo Jaaniso
Appl. Opt. 57(27) 8053-8059 (2018)
Michael S. Patterson and Brian W. Pogue
Appl. Opt. 33(10) 1963-1974 (1994)