Abstract
By simple modification of the pattern of fluorescence excitation light in an epi-illumination inverted microscope, one can achieve conditions that produce total internal reflection fluorescence (TIRF) by evanescent wave excitation. Though traditionally requiring a collimated beam traversing through a special prism, TIRF also can be achieved by epi-illumination through the periphery of a 1.4 numerical aperture objective. An opaque disk of appropriate size is placed in the illumination path external to the microscope so as to cast a sharp, real-image shadow at the objective's back focal plane. This shadow allows a hollow cone of epi-illumination rays traveling at only super-critical angles to reach the glass/water interface at the sample plane. Three kinds of TIRF illumination patterns can be produced by variations of this scheme: (1) a small spot of illumination of 1.5 μm radius by use of a laser light source, (2) a large region of illumination by use of a laser-illuminated diffusing screen located upbeam from the opaque disk, and (3) a large region of illumination by use of a conventional mercury arc.
© 1989 Optical Society of America
Full Article | PDF ArticleMore Like This
Thomas P. Burghardt, Andrew D. Hipp, and Katalin Ajtai
Appl. Opt. 48(32) 6120-6131 (2009)
Abdollah Hassanzadeh, Shabbo Saedi, Mohammadbagher Mohammadnezhad, and Salah Raza Saeed
Appl. Opt. 57(30) 9129-9139 (2018)
Henning Ortkrass, Gerd Wiebusch, Jochen Linnenbrügger, Jasmin Schürstedt, Karolina Szafranska, Peter McCourt, and Thomas Huser
Opt. Express 31(24) 40210-40220 (2023)