Abstract
A Quartz-Enhanced Photoacoustic Spectroscopy based gas sensor was optimized to improve its compact spectrophone design and performance. The impact of a 2-tube microresonator geometry on the detected signal and signal to noise ratio (SNR) was investigated. Experimental studies demonstrate that the l=4.4mm long tubes (ID=0.5mm) result in a significantly enhanced SNR.
© 2009 Optical Society of America