Abstract

We developed a novel scanning system that relies on gated detection of late photons at short source-detector separation, enabling the recording of absorption changes in deep tissue compartments. The tissue was scanned by a galvanometer scanner from a distance of more than 10 cm, with a fixed separation of the illumination and the detection spot of a few mm. The light source was a supercontinuum laser with an acousto-optic tunable filter that was used to rapidly switch between two wavelength bands centered at 760 nm and 860 nm. A fast-gated single-photon avalanche diode was employed to eliminate the intense early part of the diffusely remitted signal and to detect photons with long times of flight with improved signal-to-noise ratio. A second detection channel contained a non-gated detector. The gated and non-gated time-of-flight distributions of photons were recorded by imaging time-correlated single photon counting synchronized with the movement of the scanner. A tissue area with dimensions of several cm was scanned with 32×32 pixels within a frame time of 1 s. Sensitivity and spatial resolution of the system were characterized by phantom measurements. In-vivo tests included functional brain activation by various tasks and demonstrated the feasibility of non-contact imaging of hemodynamic changes in the cerebral cortex.

© 2015 SPIE

PDF Article

References

You do not have subscription access to this journal. Citation lists with outbound citation links are available to subscribers only. You may subscribe either as an OSA member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access OSA Member Subscription