Abstract
Based on the measurement of intrinsic fluorescence, a set of bio-aerosol including virus aerosols detection instrument is developed, with which a method of calibration is proposed using tryptophan as the target. The experimental results show a good linear relationship between the fluorescence voltage of the instrument and the concentration of the tryptophan aerosol. An excellent correlation (R2≥0.99) with the sensitivity of 4000PPL is obtained. The research demonstrates the reliability of the bio-aerosol detection by measuring the content of tryptophan. Further more the feasibility of prejudgment to the species of bio-aerosol particles with the multi-channel fluorescence detection technology is discussed.
© 2011 Optical Society of America
PDF ArticleMore Like This
Zhang Pei, Zhao Yongkai, Xiao Yanfen, Cai Shuyao, and Huang Huijie
83111R Asia Communications and Photonics Conference and Exhibition (ACP) 2011
Sarjit Bains, Darrick Niccum, and Richard Remiarz
PTuE2 Photonic Applications Systems Technologies Conference (CLEO:A&T) 2007
J. F. Pinto, M. Seaver, and J. D. Eversole
CTuL49 Conference on Lasers and Electro-Optics (CLEO:S&I) 1996